T1 - A collaborative trial for the evaluation of blood cholesterol measurement in clinical laboratories in Italy
AU - Malavasi, B.
AU - Catapano, A.
AU - Galli, G.
AU - Franzini, C.
PY - 1992
Y1 - 1992
N2 - A collaborative trial for the evaluation of blood cholesterol measurement in Italy was carried out, with the use of two lyophilized controls, whose target values, respectively 4.42 and 6.21mmol/l were established by means of 'definitive' methodology (isotope dilution/mass spectrometry). Results from 480 participants showed a somewhat broad dispersion (CV 6.1% and 6.3% respectively), and a definite bias (-0.25mmol/l and -0.61mmol/l respectively) with respect to the target values. The different analytical systems were characterized by different combinations of inaccuracy and imprecision; however, the bias observed for the higher concentration sample was a constant finding. The behaviour of the control materials, in comparison with that exhibited by patients' sera, was assessed in a manual enzymatic procedure and in the Kodak Ektachem 700 and Technicon Chem 1 systems. The peculiar properly of one control material to behave differently from patients' sera in some analytical systems, i.e. the lack of commutability, was found to be partially responsible for the observed bias in the three methods studied. The importance of testing for commutability of the control materials to be used for the control of accuracy is stressed.
AB - A collaborative trial for the evaluation of blood cholesterol measurement in Italy was carried out, with the use of two lyophilized controls, whose target values, respectively 4.42 and 6.21mmol/l were established by means of 'definitive' methodology (isotope dilution/mass spectrometry). Results from 480 participants showed a somewhat broad dispersion (CV 6.1% and 6.3% respectively), and a definite bias (-0.25mmol/l and -0.61mmol/l respectively) with respect to the target values. The different analytical systems were characterized by different combinations of inaccuracy and imprecision; however, the bias observed for the higher concentration sample was a constant finding. The behaviour of the control materials, in comparison with that exhibited by patients' sera, was assessed in a manual enzymatic procedure and in the Kodak Ektachem 700 and Technicon Chem 1 systems. The peculiar properly of one control material to behave differently from patients' sera in some analytical systems, i.e. the lack of commutability, was found to be partially responsible for the observed bias in the three methods studied. The importance of testing for commutability of the control materials to be used for the control of accuracy is stressed.
UR - http://www.scopus.com/inward/record.url?scp=0026559707&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0026559707&partnerID=8YFLogxK
M3 - Article
C2 - 1599980
AN - SCOPUS:0026559707
VL - 30
SP - 157
EP - 161
JO - European Journal of Clinical Chemistry and Clinical Biochemistry
JF - European Journal of Clinical Chemistry and Clinical Biochemistry
SN - 0939-4974
IS - 3
ER -