A double-layer tracheal tube cuff designed to prevent leakage: A bench-top study

Alberto Zanella, Massimo Cressoni, Myra Epp, Mario Stylianou, Theodor Kolobow

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: We designed a new endotracheal tube (ETT) cuff that does not form the folds that cause leakage of colonized subglottic secretions upon inflation within the trachea: a standard high-volume low-pressure cuff was draped with a second, highly elastic cuff made of a low-protein guayule natural latex rubber with 0.5 ml gel between the cuffs. We compared this prototype ETT cuff with four commercially available ETTs for efficacy in the prevention of fluid leakage across the cuff. Design: In vitro study. Measurements and results: We compared fluid leakage in our prototype cuff with that in four commercially available ETTs. Three cylindrical glass tubes 16, 20, and 22 mm in diameter were used as model tracheas, and five different intracuff pressures (20, 25, 30, 40, and 50 cmH2O) were tested. Each test was repeated three times with new ETTs. The guayule latex ETT cuff showed an average fluid leakage of 0.0007 ± 0.002 ml/min which was significantly lower than that in any of the other ETTs (Microcuff 0.07 ± 0.09, Mallinckrodt/Hi-Lo Evac 5 ± 5, Euromedical 7 ± 4, Sheridan/CF 41 ± 69). Conclusions: Our prototype ETT cuff significantly reduced fluid leakage in this bench-top study. In vivo testing and evaluation is to follow.

Original languageEnglish
Pages (from-to)1145-1149
Number of pages5
JournalIntensive Care Medicine
Volume34
Issue number6
DOIs
Publication statusPublished - Jun 2008

Keywords

  • Aspiration
  • Cuff
  • Leakage, fluid
  • Tube, endotracheal
  • Ventilator-associated pneumonia

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Fingerprint

Dive into the research topics of 'A double-layer tracheal tube cuff designed to prevent leakage: A bench-top study'. Together they form a unique fingerprint.

Cite this