A National project for in vivo dosimetry procedures in radiotherapy: First results

Angelo Piermattei, Francesca Greco, Luigi Azario, Andrea Porcelli, Savino Cilla, Sergio Zucca, Aniello Russo, Elisabetta Di Castro, Mariateresa Russo, Rocchina Caivano, Vincenzo Fusco, Alessio Morganti, Andrea Fidanzio

Research output: Contribution to journalArticle

25 Citations (Scopus)

Abstract

The paper reports the results of a National project financed by the Istituto Nazionale di Fisica Nucleare (INFN) for the development of in vivo dosimetric procedures in radiotherapy. In particular, a generalized procedure for the in vivo reconstruction of the isocenter dose, D iso, has been developed for 3D conformal radiotherapy treatments with open and wedged X-ray beams supplied by linacs of different manufacturers and equipped with aSi Electronic Portal Imaging Devices (EPIDs). In this way, the commissioning procedure is very simplified and applicable to Elekta, Siemens and Varian linacs. The method here reported is based on measurements in solid-water phantoms of different thicknesses, w, irradiated by square field sizes, L. Generalized mid-plane doses, D 0, and transit signal by EPIDs, st0, obtained by 19 open and 38 wedged beams of 8 different linacs, were determined taking into account X-ray beam and EPID calibrations. Generalized ratios F 0 = st0/D 0 for open and wedged beams were fitted by surface equations and used by a dedicated software, for the D iso reconstruction. Moreover, for each beam the software supplied a set of transit signal profiles crossing the beam central axis to test the beam irradiation reproducibility. The tolerance level of the comparison between the D iso and the dose computed by the TPS, D iso,TPS, was estimated 5%. The generalized in vivo dosimetry procedure was adopted by 3 centers that used different linacs. The results of 480 tests showed that in absence of errors, the comparison between D iso and D iso,TPS resulted well-within the tolerance level. The presence of errors was detected in 10% of the tests and were due to essentially an incorrect set-up, presence of an attenuator on the beams and patient morphological changes. Moreover, the dedicated software used the information of the Record and Verify system of the centres and consequently the extra-time needed to obtain, for each beam, the D iso reconstruction after the dose delivery, was about 20 s.

Original languageEnglish
Pages (from-to)42-50
Number of pages9
JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
Volume274
DOIs
Publication statusPublished - Mar 1 2012

Fingerprint

Radiotherapy
Dosimetry
dosimeters
radiation therapy
Imaging techniques
X rays
dosage
Irradiation
Calibration
transit
computer programs
electronics
Water
attenuators
delivery
x rays
irradiation
profiles

Keywords

  • Electronic portal imaging device
  • In vivo dosimetry
  • Quality control in radiotherapy

ASJC Scopus subject areas

  • Instrumentation
  • Nuclear and High Energy Physics

Cite this

A National project for in vivo dosimetry procedures in radiotherapy : First results. / Piermattei, Angelo; Greco, Francesca; Azario, Luigi; Porcelli, Andrea; Cilla, Savino; Zucca, Sergio; Russo, Aniello; Castro, Elisabetta Di; Russo, Mariateresa; Caivano, Rocchina; Fusco, Vincenzo; Morganti, Alessio; Fidanzio, Andrea.

In: Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms, Vol. 274, 01.03.2012, p. 42-50.

Research output: Contribution to journalArticle

Piermattei, Angelo ; Greco, Francesca ; Azario, Luigi ; Porcelli, Andrea ; Cilla, Savino ; Zucca, Sergio ; Russo, Aniello ; Castro, Elisabetta Di ; Russo, Mariateresa ; Caivano, Rocchina ; Fusco, Vincenzo ; Morganti, Alessio ; Fidanzio, Andrea. / A National project for in vivo dosimetry procedures in radiotherapy : First results. In: Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms. 2012 ; Vol. 274. pp. 42-50.
@article{5e100ea424024495ae4f7e397bdedb3c,
title = "A National project for in vivo dosimetry procedures in radiotherapy: First results",
abstract = "The paper reports the results of a National project financed by the Istituto Nazionale di Fisica Nucleare (INFN) for the development of in vivo dosimetric procedures in radiotherapy. In particular, a generalized procedure for the in vivo reconstruction of the isocenter dose, D iso, has been developed for 3D conformal radiotherapy treatments with open and wedged X-ray beams supplied by linacs of different manufacturers and equipped with aSi Electronic Portal Imaging Devices (EPIDs). In this way, the commissioning procedure is very simplified and applicable to Elekta, Siemens and Varian linacs. The method here reported is based on measurements in solid-water phantoms of different thicknesses, w, irradiated by square field sizes, L. Generalized mid-plane doses, D 0, and transit signal by EPIDs, st0, obtained by 19 open and 38 wedged beams of 8 different linacs, were determined taking into account X-ray beam and EPID calibrations. Generalized ratios F 0 = st0/D 0 for open and wedged beams were fitted by surface equations and used by a dedicated software, for the D iso reconstruction. Moreover, for each beam the software supplied a set of transit signal profiles crossing the beam central axis to test the beam irradiation reproducibility. The tolerance level of the comparison between the D iso and the dose computed by the TPS, D iso,TPS, was estimated 5{\%}. The generalized in vivo dosimetry procedure was adopted by 3 centers that used different linacs. The results of 480 tests showed that in absence of errors, the comparison between D iso and D iso,TPS resulted well-within the tolerance level. The presence of errors was detected in 10{\%} of the tests and were due to essentially an incorrect set-up, presence of an attenuator on the beams and patient morphological changes. Moreover, the dedicated software used the information of the Record and Verify system of the centres and consequently the extra-time needed to obtain, for each beam, the D iso reconstruction after the dose delivery, was about 20 s.",
keywords = "Electronic portal imaging device, In vivo dosimetry, Quality control in radiotherapy",
author = "Angelo Piermattei and Francesca Greco and Luigi Azario and Andrea Porcelli and Savino Cilla and Sergio Zucca and Aniello Russo and Castro, {Elisabetta Di} and Mariateresa Russo and Rocchina Caivano and Vincenzo Fusco and Alessio Morganti and Andrea Fidanzio",
year = "2012",
month = "3",
day = "1",
doi = "10.1016/j.nimb.2011.12.004",
language = "English",
volume = "274",
pages = "42--50",
journal = "Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms",
issn = "0168-583X",
publisher = "Elsevier",

}

TY - JOUR

T1 - A National project for in vivo dosimetry procedures in radiotherapy

T2 - First results

AU - Piermattei, Angelo

AU - Greco, Francesca

AU - Azario, Luigi

AU - Porcelli, Andrea

AU - Cilla, Savino

AU - Zucca, Sergio

AU - Russo, Aniello

AU - Castro, Elisabetta Di

AU - Russo, Mariateresa

AU - Caivano, Rocchina

AU - Fusco, Vincenzo

AU - Morganti, Alessio

AU - Fidanzio, Andrea

PY - 2012/3/1

Y1 - 2012/3/1

N2 - The paper reports the results of a National project financed by the Istituto Nazionale di Fisica Nucleare (INFN) for the development of in vivo dosimetric procedures in radiotherapy. In particular, a generalized procedure for the in vivo reconstruction of the isocenter dose, D iso, has been developed for 3D conformal radiotherapy treatments with open and wedged X-ray beams supplied by linacs of different manufacturers and equipped with aSi Electronic Portal Imaging Devices (EPIDs). In this way, the commissioning procedure is very simplified and applicable to Elekta, Siemens and Varian linacs. The method here reported is based on measurements in solid-water phantoms of different thicknesses, w, irradiated by square field sizes, L. Generalized mid-plane doses, D 0, and transit signal by EPIDs, st0, obtained by 19 open and 38 wedged beams of 8 different linacs, were determined taking into account X-ray beam and EPID calibrations. Generalized ratios F 0 = st0/D 0 for open and wedged beams were fitted by surface equations and used by a dedicated software, for the D iso reconstruction. Moreover, for each beam the software supplied a set of transit signal profiles crossing the beam central axis to test the beam irradiation reproducibility. The tolerance level of the comparison between the D iso and the dose computed by the TPS, D iso,TPS, was estimated 5%. The generalized in vivo dosimetry procedure was adopted by 3 centers that used different linacs. The results of 480 tests showed that in absence of errors, the comparison between D iso and D iso,TPS resulted well-within the tolerance level. The presence of errors was detected in 10% of the tests and were due to essentially an incorrect set-up, presence of an attenuator on the beams and patient morphological changes. Moreover, the dedicated software used the information of the Record and Verify system of the centres and consequently the extra-time needed to obtain, for each beam, the D iso reconstruction after the dose delivery, was about 20 s.

AB - The paper reports the results of a National project financed by the Istituto Nazionale di Fisica Nucleare (INFN) for the development of in vivo dosimetric procedures in radiotherapy. In particular, a generalized procedure for the in vivo reconstruction of the isocenter dose, D iso, has been developed for 3D conformal radiotherapy treatments with open and wedged X-ray beams supplied by linacs of different manufacturers and equipped with aSi Electronic Portal Imaging Devices (EPIDs). In this way, the commissioning procedure is very simplified and applicable to Elekta, Siemens and Varian linacs. The method here reported is based on measurements in solid-water phantoms of different thicknesses, w, irradiated by square field sizes, L. Generalized mid-plane doses, D 0, and transit signal by EPIDs, st0, obtained by 19 open and 38 wedged beams of 8 different linacs, were determined taking into account X-ray beam and EPID calibrations. Generalized ratios F 0 = st0/D 0 for open and wedged beams were fitted by surface equations and used by a dedicated software, for the D iso reconstruction. Moreover, for each beam the software supplied a set of transit signal profiles crossing the beam central axis to test the beam irradiation reproducibility. The tolerance level of the comparison between the D iso and the dose computed by the TPS, D iso,TPS, was estimated 5%. The generalized in vivo dosimetry procedure was adopted by 3 centers that used different linacs. The results of 480 tests showed that in absence of errors, the comparison between D iso and D iso,TPS resulted well-within the tolerance level. The presence of errors was detected in 10% of the tests and were due to essentially an incorrect set-up, presence of an attenuator on the beams and patient morphological changes. Moreover, the dedicated software used the information of the Record and Verify system of the centres and consequently the extra-time needed to obtain, for each beam, the D iso reconstruction after the dose delivery, was about 20 s.

KW - Electronic portal imaging device

KW - In vivo dosimetry

KW - Quality control in radiotherapy

UR - http://www.scopus.com/inward/record.url?scp=84855172597&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84855172597&partnerID=8YFLogxK

U2 - 10.1016/j.nimb.2011.12.004

DO - 10.1016/j.nimb.2011.12.004

M3 - Article

AN - SCOPUS:84855172597

VL - 274

SP - 42

EP - 50

JO - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

JF - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms

SN - 0168-583X

ER -