A new hip-knee-ankle-foot sling: Kinematic comparison with a traditional ankle-foot orthosis

Arturo Zancan, Maria V. Beretta, Micaela Schmid, Marco Schieppati

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

In this study, we performed a kinematic analysis of a new, low-cost sling for the lower limb, compared to a common ankle-foot orthosis (AFO). Gait with no orthosis, with the AFO, and with the new sling was analyzed in one hemiplegic subject. Both the AFO and the sling reduced the mean angle and ROM (range of movement) of the ankle and the vertical displacement of the center of mass. The sling, but not the AFO, restored the normal sequence heel-strike, forefoot contact of the affected side. The sling, but not the AFO, reduced the affected limb stance and stride duration, increased stride length, and improved walking speed. In conclusion, the proposed sling for the lower limb equally improved the affected ankle kinematics in contrast to the traditional AFO, and it also improved some gait variables in this hemiplegic subject.

Original languageEnglish
Pages (from-to)707-712
Number of pages6
JournalJournal of Rehabilitation Research and Development
Volume41
Issue number5
DOIs
Publication statusPublished - Sep 2004

Fingerprint

Slings
Foot Orthoses
Biomechanical Phenomena
Ankle
Hip
Foot
Knee
Kinematics
Gait
Lower Extremity
Orthotic Devices
Heel
Extremities
Costs and Cost Analysis

Keywords

  • AFO
  • Gait
  • Hemiplegia
  • Kinematic
  • Lower limb
  • Orthosis
  • Sling

ASJC Scopus subject areas

  • Rehabilitation
  • Health Professions(all)
  • Engineering(all)

Cite this

A new hip-knee-ankle-foot sling : Kinematic comparison with a traditional ankle-foot orthosis. / Zancan, Arturo; Beretta, Maria V.; Schmid, Micaela; Schieppati, Marco.

In: Journal of Rehabilitation Research and Development, Vol. 41, No. 5, 09.2004, p. 707-712.

Research output: Contribution to journalArticle

@article{e267fb96351e48e985125e99ded1c71f,
title = "A new hip-knee-ankle-foot sling: Kinematic comparison with a traditional ankle-foot orthosis",
abstract = "In this study, we performed a kinematic analysis of a new, low-cost sling for the lower limb, compared to a common ankle-foot orthosis (AFO). Gait with no orthosis, with the AFO, and with the new sling was analyzed in one hemiplegic subject. Both the AFO and the sling reduced the mean angle and ROM (range of movement) of the ankle and the vertical displacement of the center of mass. The sling, but not the AFO, restored the normal sequence heel-strike, forefoot contact of the affected side. The sling, but not the AFO, reduced the affected limb stance and stride duration, increased stride length, and improved walking speed. In conclusion, the proposed sling for the lower limb equally improved the affected ankle kinematics in contrast to the traditional AFO, and it also improved some gait variables in this hemiplegic subject.",
keywords = "AFO, Gait, Hemiplegia, Kinematic, Lower limb, Orthosis, Sling",
author = "Arturo Zancan and Beretta, {Maria V.} and Micaela Schmid and Marco Schieppati",
year = "2004",
month = "9",
doi = "10.1682/JRRD.2003.06.0101",
language = "English",
volume = "41",
pages = "707--712",
journal = "Journal of rehabilitation R&D",
issn = "0748-7711",
publisher = "Rehabilitation Research and Development Service",
number = "5",

}

TY - JOUR

T1 - A new hip-knee-ankle-foot sling

T2 - Kinematic comparison with a traditional ankle-foot orthosis

AU - Zancan, Arturo

AU - Beretta, Maria V.

AU - Schmid, Micaela

AU - Schieppati, Marco

PY - 2004/9

Y1 - 2004/9

N2 - In this study, we performed a kinematic analysis of a new, low-cost sling for the lower limb, compared to a common ankle-foot orthosis (AFO). Gait with no orthosis, with the AFO, and with the new sling was analyzed in one hemiplegic subject. Both the AFO and the sling reduced the mean angle and ROM (range of movement) of the ankle and the vertical displacement of the center of mass. The sling, but not the AFO, restored the normal sequence heel-strike, forefoot contact of the affected side. The sling, but not the AFO, reduced the affected limb stance and stride duration, increased stride length, and improved walking speed. In conclusion, the proposed sling for the lower limb equally improved the affected ankle kinematics in contrast to the traditional AFO, and it also improved some gait variables in this hemiplegic subject.

AB - In this study, we performed a kinematic analysis of a new, low-cost sling for the lower limb, compared to a common ankle-foot orthosis (AFO). Gait with no orthosis, with the AFO, and with the new sling was analyzed in one hemiplegic subject. Both the AFO and the sling reduced the mean angle and ROM (range of movement) of the ankle and the vertical displacement of the center of mass. The sling, but not the AFO, restored the normal sequence heel-strike, forefoot contact of the affected side. The sling, but not the AFO, reduced the affected limb stance and stride duration, increased stride length, and improved walking speed. In conclusion, the proposed sling for the lower limb equally improved the affected ankle kinematics in contrast to the traditional AFO, and it also improved some gait variables in this hemiplegic subject.

KW - AFO

KW - Gait

KW - Hemiplegia

KW - Kinematic

KW - Lower limb

KW - Orthosis

KW - Sling

UR - http://www.scopus.com/inward/record.url?scp=8644281084&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=8644281084&partnerID=8YFLogxK

U2 - 10.1682/JRRD.2003.06.0101

DO - 10.1682/JRRD.2003.06.0101

M3 - Article

C2 - 15558400

AN - SCOPUS:8644281084

VL - 41

SP - 707

EP - 712

JO - Journal of rehabilitation R&D

JF - Journal of rehabilitation R&D

SN - 0748-7711

IS - 5

ER -