A sensitive, rapid and inexpensive method to assess aluminium(III) ions in technetium eluates

Federica E. Buroni, Lorenzo Lodola, Marco G. Persico, Carlo Aprile

Research output: Contribution to journalArticlepeer-review

Abstract

The aim of the study was to validate a semiquantitative analytical method to identify the aluminium(III) [Al(III)] concentration in Mo/Tc generator eluates to check the European Pharmacopoeia (Ph. Eur.) requirement (

Original languageEnglish
Pages (from-to)777-780
Number of pages4
JournalNuclear Medicine Communications
Volume35
Issue number7
DOIs
Publication statusPublished - 2014

Keywords

  • aluminium spot test
  • chemical purity
  • quality control
  • sodium pertechnetate

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Medicine(all)

Fingerprint Dive into the research topics of 'A sensitive, rapid and inexpensive method to assess aluminium(III) ions in technetium eluates'. Together they form a unique fingerprint.

Cite this