Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: Experience from five radiotherapy departments

Ann Van Esch, Jörg Bohsung, Pekka Sorvari, Mikko Tenhunen, Marta Paiusco, Mauro Iori, Per Engström, Håkan Nyström, Dominique Pierre Huyskens

Research output: Contribution to journalArticlepeer-review

Abstract

Background and purpose: An increasing number of radiotherapy centres is now aiming for clinical implementation of intensity modulated radiotherapy (IMRT), but - in contrast to conventional treatment - no national or international guidelines for commissioning of the treatment planning system (TPS) and acceptance tests of treatment equipment have yet been developed. This paper bundles the experience of five radiotherapy departments that have introduced IMRT into their clinical routine. Methods and materials: The five radiotherapy departments are using similar configurations since they adopted the commercially available Varian solution for IMRT, regarding treatment planning as well as treatment delivery. All are using the sliding window technique. Different approaches towards the derivation of the multileaf collimator (MLC) parameters required for the configuration of the TPS are described. A description of the quality control procedures for the dynamic MLC, including their respective frequencies, is given. For the acceptance of the TPS for IMRT multiple quality control plans were developed on a variety of phantoms, testing the flexibility of the inverse planning modules to produce the desired dose pattern as well as assessing the accuracy of the dose calculation. Regarding patient treatment verification, all five centres perform dosimetric pre-treatment verification of the treatment fields, be it on a single field or on a total plan procedure. During the actual treatment, the primary focus is on patient positioning rather than dosimetry. Intracavitary in vivo measurements were performed in special cases. Result and conclusion: The configurational MLC parameters obtained through different methods are not identical for all centres, but the observed variations have shown to be of no significant clinical relevance. The quality control (QC) procedures for the dMLC have not detected any discrepancies since their initiation, demonstrating the reliability of the MLC controller. The development of geometrically simple QC plans to test the inverse planning, the dynamic MLC modules and the final dose calculation has proven to be useful in pointing out the need to remodel the single pencil beam scatter kernels in some centres. The final correspondence between calculated and measured dose was found to be satisfactory by all centres, for QC test plans as well as for pre-treatment verification of clinical IMRT fields. An intercomparison of the man hours needed per patient plan verification reveals a substantial variation depending on the type of measurements performed.

Original languageEnglish
Pages (from-to)53-70
Number of pages18
JournalRadiotherapy and Oncology
Volume65
Issue number1
DOIs
Publication statusPublished - Oct 1 2002

Keywords

  • Acceptance tests
  • Intensity modulated radiotherapy
  • Inverse planning
  • Quality control procedures
  • Sliding window

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Urology

Fingerprint Dive into the research topics of 'Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: Experience from five radiotherapy departments'. Together they form a unique fingerprint.