Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A

N. Ciavarella, S. Coccheri, P. M. Mannucci

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.

Original languageEnglish
Pages (from-to)308-317
Number of pages10
JournalScandinavian Journal of Haematology
Volume25
Issue number4
Publication statusPublished - 1980

Fingerprint

Partial Thromboplastin Time
Hemophilia A
Factor VIII
Analysis of Variance
Reading
Reference Values

ASJC Scopus subject areas

  • Hematology

Cite this

Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A. / Ciavarella, N.; Coccheri, S.; Mannucci, P. M.

In: Scandinavian Journal of Haematology, Vol. 25, No. 4, 1980, p. 308-317.

Research output: Contribution to journalArticle

@article{94801f23017649e1bed627c35027f477,
title = "Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A",
abstract = "An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.",
author = "N. Ciavarella and S. Coccheri and Mannucci, {P. M.}",
year = "1980",
language = "English",
volume = "25",
pages = "308--317",
journal = "European Journal of Haematology",
issn = "0902-4441",
publisher = "Wiley-Blackwell Publishing Ltd",
number = "4",

}

TY - JOUR

T1 - Activated partial thromboplastin time. A multicenter evaluation of 11 reagents in the screening of mild haemophilia A

AU - Ciavarella, N.

AU - Coccheri, S.

AU - Mannucci, P. M.

PY - 1980

Y1 - 1980

N2 - An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.

AB - An internationally standardized preparation and 10 commercial kits widely used to perform the activated partial thromboplastin time (APTT) were compared in 4 laboratories for the purpose of assessing their ability to detect mild deficiencies of factor VIII activity. The participating laboratories were asked to carry out with each APTT reagent quadruplicate readings of 3 coded lyophilized plasmas containing varying levels of factor VIII (109, 26 and 17 U/dl respectively). An analysis of variance of clotting times showed significant differences between reagents and laboratories. All the reagents detected the abnormality of the plasma containing 17 U/dl, whereas a number of failures were found when the plasma with 26 U/dl was tested. When analysis of variance was carried out on ratios of factor-VIII deficient to normal plasma clotting times, the results showed less difference between laboratories and reagents. Clotting times of plasma with normal factor VIII level (109 U/dl) usually fell within the normal range indicated by manufacturers of the commercial reagents.

UR - http://www.scopus.com/inward/record.url?scp=0019166316&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0019166316&partnerID=8YFLogxK

M3 - Article

VL - 25

SP - 308

EP - 317

JO - European Journal of Haematology

JF - European Journal of Haematology

SN - 0902-4441

IS - 4

ER -