An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures

Gianluigi Li Bassi, Otavio Tavares Ranzani, Joan Daniel Marti, Valeria Giunta, Nestor Luque, Valentina Isetta, Miquel Ferrer, Ramon Farre, Guilherme Leite Pimentel, Antoni Torres

Research output: Contribution to journalArticle

Abstract

OBJECTIVE: To assess the structural characteristics involved in the design of high-volume low-pressure endotracheal tube cuffs that are associated with fluid sealing effectiveness and to determine the extent of transmitted tracheal pressures upon cuff inflation. DESIGN: In vitro study. SETTINGS: Pneumology laboratories. INTERVENTIONS: Eight high-volume low-pressure cuffs of cylindrical or tapered shape, made of polyvinylchloride or polyurethane, were studied. Cuffs were tested within a tracheal model, oriented 30 above horizontal to assess 1 hr leakage of oropharyngeal secretions simulant at cuff internal pressures of 15-30 cm H2O. The four best performing cuffs were evaluated for 24 hrs using an internal pressure of 30 cm H2O. The extent of transmitted tracheal wall pressure throughout the cuff-trachea contact area was determined using an internal pressure sensor within a tracheal model upon cuff inflation up to 30 cm H2O. MEASUREMENTS AND MAIN RESULTS: Outer diameter, length, and compliance of each cuff were assessed. Multivariate regression analysis was performed to identify the main determinants of simulant leakage rate. The cuff-trachea contact area and the percentage of tracheal wall pressure measurements greater than 50 cm H2O were computed. Cuff design characteristics significantly differ among tubes. The cuffs made of polyurethane showed the best short-and long-term sealing efficacy. Nevertheless, in the multivariate analysis, the cuff outer diameter (n: 288, p = 0.003) and length (n: 288, p <0.001), along with the internal pressure (n: 288, p <0.001), were the only predictors of simulant leakage rate. The tapered cuffs showed the lowest tracheal wall contact area (n: 96, p <0.001). The tracheal wall pressure distribution pattern was heterogeneous, and the percentage of high tracheal wall pressure significantly differs among the cuffs (n: 96, p <0.001). CONCLUSIONS: The high-volume low-pressure cuffs' outer diameter, length, material, and internal pressure are the main determinants of sealing efficacy. Despite internal pressure within the safe range, transmitted tracheal pressure is extremely heterogeneous and differs among cuffs, occasionally reaching localized, very high, unsafe levels.

Original languageEnglish
Pages (from-to)518-526
Number of pages9
JournalCritical Care Medicine
Volume41
Issue number2
DOIs
Publication statusPublished - Feb 2013

Fingerprint

Pressure
Polyurethanes
Economic Inflation
In Vitro Techniques
Trachea
Multivariate Analysis
Pulmonary Medicine
Compliance
Regression Analysis

Keywords

  • endotracheal tube
  • polyurethane
  • polyvinylchloride
  • pulmonary aspiration
  • trachea
  • tracheal injury

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Cite this

An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures. / Li Bassi, Gianluigi; Ranzani, Otavio Tavares; Marti, Joan Daniel; Giunta, Valeria; Luque, Nestor; Isetta, Valentina; Ferrer, Miquel; Farre, Ramon; Pimentel, Guilherme Leite; Torres, Antoni.

In: Critical Care Medicine, Vol. 41, No. 2, 02.2013, p. 518-526.

Research output: Contribution to journalArticle

Li Bassi, G, Ranzani, OT, Marti, JD, Giunta, V, Luque, N, Isetta, V, Ferrer, M, Farre, R, Pimentel, GL & Torres, A 2013, 'An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures', Critical Care Medicine, vol. 41, no. 2, pp. 518-526. https://doi.org/10.1097/CCM.0b013e31826a4804
Li Bassi, Gianluigi ; Ranzani, Otavio Tavares ; Marti, Joan Daniel ; Giunta, Valeria ; Luque, Nestor ; Isetta, Valentina ; Ferrer, Miquel ; Farre, Ramon ; Pimentel, Guilherme Leite ; Torres, Antoni. / An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures. In: Critical Care Medicine. 2013 ; Vol. 41, No. 2. pp. 518-526.
@article{de163d522f6c4732b3edeab7363ac76c,
title = "An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures",
abstract = "OBJECTIVE: To assess the structural characteristics involved in the design of high-volume low-pressure endotracheal tube cuffs that are associated with fluid sealing effectiveness and to determine the extent of transmitted tracheal pressures upon cuff inflation. DESIGN: In vitro study. SETTINGS: Pneumology laboratories. INTERVENTIONS: Eight high-volume low-pressure cuffs of cylindrical or tapered shape, made of polyvinylchloride or polyurethane, were studied. Cuffs were tested within a tracheal model, oriented 30 above horizontal to assess 1 hr leakage of oropharyngeal secretions simulant at cuff internal pressures of 15-30 cm H2O. The four best performing cuffs were evaluated for 24 hrs using an internal pressure of 30 cm H2O. The extent of transmitted tracheal wall pressure throughout the cuff-trachea contact area was determined using an internal pressure sensor within a tracheal model upon cuff inflation up to 30 cm H2O. MEASUREMENTS AND MAIN RESULTS: Outer diameter, length, and compliance of each cuff were assessed. Multivariate regression analysis was performed to identify the main determinants of simulant leakage rate. The cuff-trachea contact area and the percentage of tracheal wall pressure measurements greater than 50 cm H2O were computed. Cuff design characteristics significantly differ among tubes. The cuffs made of polyurethane showed the best short-and long-term sealing efficacy. Nevertheless, in the multivariate analysis, the cuff outer diameter (n: 288, p = 0.003) and length (n: 288, p <0.001), along with the internal pressure (n: 288, p <0.001), were the only predictors of simulant leakage rate. The tapered cuffs showed the lowest tracheal wall contact area (n: 96, p <0.001). The tracheal wall pressure distribution pattern was heterogeneous, and the percentage of high tracheal wall pressure significantly differs among the cuffs (n: 96, p <0.001). CONCLUSIONS: The high-volume low-pressure cuffs' outer diameter, length, material, and internal pressure are the main determinants of sealing efficacy. Despite internal pressure within the safe range, transmitted tracheal pressure is extremely heterogeneous and differs among cuffs, occasionally reaching localized, very high, unsafe levels.",
keywords = "endotracheal tube, polyurethane, polyvinylchloride, pulmonary aspiration, trachea, tracheal injury",
author = "{Li Bassi}, Gianluigi and Ranzani, {Otavio Tavares} and Marti, {Joan Daniel} and Valeria Giunta and Nestor Luque and Valentina Isetta and Miquel Ferrer and Ramon Farre and Pimentel, {Guilherme Leite} and Antoni Torres",
year = "2013",
month = "2",
doi = "10.1097/CCM.0b013e31826a4804",
language = "English",
volume = "41",
pages = "518--526",
journal = "Critical Care Medicine",
issn = "0090-3493",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures

AU - Li Bassi, Gianluigi

AU - Ranzani, Otavio Tavares

AU - Marti, Joan Daniel

AU - Giunta, Valeria

AU - Luque, Nestor

AU - Isetta, Valentina

AU - Ferrer, Miquel

AU - Farre, Ramon

AU - Pimentel, Guilherme Leite

AU - Torres, Antoni

PY - 2013/2

Y1 - 2013/2

N2 - OBJECTIVE: To assess the structural characteristics involved in the design of high-volume low-pressure endotracheal tube cuffs that are associated with fluid sealing effectiveness and to determine the extent of transmitted tracheal pressures upon cuff inflation. DESIGN: In vitro study. SETTINGS: Pneumology laboratories. INTERVENTIONS: Eight high-volume low-pressure cuffs of cylindrical or tapered shape, made of polyvinylchloride or polyurethane, were studied. Cuffs were tested within a tracheal model, oriented 30 above horizontal to assess 1 hr leakage of oropharyngeal secretions simulant at cuff internal pressures of 15-30 cm H2O. The four best performing cuffs were evaluated for 24 hrs using an internal pressure of 30 cm H2O. The extent of transmitted tracheal wall pressure throughout the cuff-trachea contact area was determined using an internal pressure sensor within a tracheal model upon cuff inflation up to 30 cm H2O. MEASUREMENTS AND MAIN RESULTS: Outer diameter, length, and compliance of each cuff were assessed. Multivariate regression analysis was performed to identify the main determinants of simulant leakage rate. The cuff-trachea contact area and the percentage of tracheal wall pressure measurements greater than 50 cm H2O were computed. Cuff design characteristics significantly differ among tubes. The cuffs made of polyurethane showed the best short-and long-term sealing efficacy. Nevertheless, in the multivariate analysis, the cuff outer diameter (n: 288, p = 0.003) and length (n: 288, p <0.001), along with the internal pressure (n: 288, p <0.001), were the only predictors of simulant leakage rate. The tapered cuffs showed the lowest tracheal wall contact area (n: 96, p <0.001). The tracheal wall pressure distribution pattern was heterogeneous, and the percentage of high tracheal wall pressure significantly differs among the cuffs (n: 96, p <0.001). CONCLUSIONS: The high-volume low-pressure cuffs' outer diameter, length, material, and internal pressure are the main determinants of sealing efficacy. Despite internal pressure within the safe range, transmitted tracheal pressure is extremely heterogeneous and differs among cuffs, occasionally reaching localized, very high, unsafe levels.

AB - OBJECTIVE: To assess the structural characteristics involved in the design of high-volume low-pressure endotracheal tube cuffs that are associated with fluid sealing effectiveness and to determine the extent of transmitted tracheal pressures upon cuff inflation. DESIGN: In vitro study. SETTINGS: Pneumology laboratories. INTERVENTIONS: Eight high-volume low-pressure cuffs of cylindrical or tapered shape, made of polyvinylchloride or polyurethane, were studied. Cuffs were tested within a tracheal model, oriented 30 above horizontal to assess 1 hr leakage of oropharyngeal secretions simulant at cuff internal pressures of 15-30 cm H2O. The four best performing cuffs were evaluated for 24 hrs using an internal pressure of 30 cm H2O. The extent of transmitted tracheal wall pressure throughout the cuff-trachea contact area was determined using an internal pressure sensor within a tracheal model upon cuff inflation up to 30 cm H2O. MEASUREMENTS AND MAIN RESULTS: Outer diameter, length, and compliance of each cuff were assessed. Multivariate regression analysis was performed to identify the main determinants of simulant leakage rate. The cuff-trachea contact area and the percentage of tracheal wall pressure measurements greater than 50 cm H2O were computed. Cuff design characteristics significantly differ among tubes. The cuffs made of polyurethane showed the best short-and long-term sealing efficacy. Nevertheless, in the multivariate analysis, the cuff outer diameter (n: 288, p = 0.003) and length (n: 288, p <0.001), along with the internal pressure (n: 288, p <0.001), were the only predictors of simulant leakage rate. The tapered cuffs showed the lowest tracheal wall contact area (n: 96, p <0.001). The tracheal wall pressure distribution pattern was heterogeneous, and the percentage of high tracheal wall pressure significantly differs among the cuffs (n: 96, p <0.001). CONCLUSIONS: The high-volume low-pressure cuffs' outer diameter, length, material, and internal pressure are the main determinants of sealing efficacy. Despite internal pressure within the safe range, transmitted tracheal pressure is extremely heterogeneous and differs among cuffs, occasionally reaching localized, very high, unsafe levels.

KW - endotracheal tube

KW - polyurethane

KW - polyvinylchloride

KW - pulmonary aspiration

KW - trachea

KW - tracheal injury

UR - http://www.scopus.com/inward/record.url?scp=84873360319&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84873360319&partnerID=8YFLogxK

U2 - 10.1097/CCM.0b013e31826a4804

DO - 10.1097/CCM.0b013e31826a4804

M3 - Article

C2 - 23263575

AN - SCOPUS:84873360319

VL - 41

SP - 518

EP - 526

JO - Critical Care Medicine

JF - Critical Care Medicine

SN - 0090-3493

IS - 2

ER -