Analysis of the viscoelastic properties of the human cornea using scheimpflug imaging in inflation experiment of eye globes

Giuseppe Lombardo, Sebastiano Serrao, Marianna Rosati, Marco Lombardo

Research output: Contribution to journalArticle

19 Citations (Scopus)

Abstract

Purpose: To demonstrate a Scheimpflug-based imaging procedure for investigating the depth- and time-dependent strain response of the human cornea to inflation testing of whole eye globes.

Methods: Six specimens, three of which with intact corneal epithelium, were mounted in a customized apparatus within a humidity and temperature-monitored wet chamber. Each specimen was subjected to two mechanical tests in order to measure corneal strain resulting from application of cyclic (cyclic regimen) and constant (creep regimen) stress by changing the intra-ocular pressure (IOP) within physiological ranges (18-42 mmHg). Corneal shape changes were analyzed as a function of IOP and both corneal stress-strain curves and creep curves were generated.

Results: The procedure was highly accurate and repeatable. Upon cyclic stress application, a biomechanical corneal elasticity gradient was found in the front-back direction. The average Young's modulus of the anterior cornea ranged between 2.28±0.87 MPa and 3.30±0.90 MPa in specimens with and without intact epithelium (P = 0.05) respectively. The Young's modulus of the posterior cornea was on average 0.21±0.09 MPa and 0.17±0.06 MPa (P>0.05) respectively. The time-dependent strain response of the cornea to creep testing was quantified by fitting data to a modified Zener model for extracting both the relaxation time and compliance function.

Conclusion: Cyclic and creep mechanical tests are valuable for investigating the strain response of the intact human cornea within physiological IOP ranges, providing meaningful results that can be translated to clinic. The presence of epithelium influences the results of anterior corneal shape changes when monitoring deformation via Scheimpflug imaging in inflation experiments of whole eye globes.

Original languageEnglish
Article number0112169
JournalPLoS One
Volume9
Issue number11
DOIs
Publication statusPublished - Nov 14 2014

Fingerprint

inflation
Economic Inflation
cornea
Cornea
eyes
image analysis
Imaging techniques
Creep
epithelium
Elastic Modulus
modulus of elasticity
Pressure
Elastic moduli
mechanical creep
Experiments
Creep testing
Epithelium
testing
Stress-strain curves
Corneal Epithelium

ASJC Scopus subject areas

  • Agricultural and Biological Sciences(all)
  • Biochemistry, Genetics and Molecular Biology(all)
  • Medicine(all)

Cite this

Analysis of the viscoelastic properties of the human cornea using scheimpflug imaging in inflation experiment of eye globes. / Lombardo, Giuseppe; Serrao, Sebastiano; Rosati, Marianna; Lombardo, Marco.

In: PLoS One, Vol. 9, No. 11, 0112169, 14.11.2014.

Research output: Contribution to journalArticle

Lombardo, Giuseppe ; Serrao, Sebastiano ; Rosati, Marianna ; Lombardo, Marco. / Analysis of the viscoelastic properties of the human cornea using scheimpflug imaging in inflation experiment of eye globes. In: PLoS One. 2014 ; Vol. 9, No. 11.
@article{8da084bc796248dc8fcf7f19fdcd10c3,
title = "Analysis of the viscoelastic properties of the human cornea using scheimpflug imaging in inflation experiment of eye globes",
abstract = "Purpose: To demonstrate a Scheimpflug-based imaging procedure for investigating the depth- and time-dependent strain response of the human cornea to inflation testing of whole eye globes.Methods: Six specimens, three of which with intact corneal epithelium, were mounted in a customized apparatus within a humidity and temperature-monitored wet chamber. Each specimen was subjected to two mechanical tests in order to measure corneal strain resulting from application of cyclic (cyclic regimen) and constant (creep regimen) stress by changing the intra-ocular pressure (IOP) within physiological ranges (18-42 mmHg). Corneal shape changes were analyzed as a function of IOP and both corneal stress-strain curves and creep curves were generated.Results: The procedure was highly accurate and repeatable. Upon cyclic stress application, a biomechanical corneal elasticity gradient was found in the front-back direction. The average Young's modulus of the anterior cornea ranged between 2.28±0.87 MPa and 3.30±0.90 MPa in specimens with and without intact epithelium (P = 0.05) respectively. The Young's modulus of the posterior cornea was on average 0.21±0.09 MPa and 0.17±0.06 MPa (P>0.05) respectively. The time-dependent strain response of the cornea to creep testing was quantified by fitting data to a modified Zener model for extracting both the relaxation time and compliance function.Conclusion: Cyclic and creep mechanical tests are valuable for investigating the strain response of the intact human cornea within physiological IOP ranges, providing meaningful results that can be translated to clinic. The presence of epithelium influences the results of anterior corneal shape changes when monitoring deformation via Scheimpflug imaging in inflation experiments of whole eye globes.",
author = "Giuseppe Lombardo and Sebastiano Serrao and Marianna Rosati and Marco Lombardo",
year = "2014",
month = "11",
day = "14",
doi = "10.1371/journal.pone.0112169",
language = "English",
volume = "9",
journal = "PLoS One",
issn = "1932-6203",
publisher = "Public Library of Science",
number = "11",

}

TY - JOUR

T1 - Analysis of the viscoelastic properties of the human cornea using scheimpflug imaging in inflation experiment of eye globes

AU - Lombardo, Giuseppe

AU - Serrao, Sebastiano

AU - Rosati, Marianna

AU - Lombardo, Marco

PY - 2014/11/14

Y1 - 2014/11/14

N2 - Purpose: To demonstrate a Scheimpflug-based imaging procedure for investigating the depth- and time-dependent strain response of the human cornea to inflation testing of whole eye globes.Methods: Six specimens, three of which with intact corneal epithelium, were mounted in a customized apparatus within a humidity and temperature-monitored wet chamber. Each specimen was subjected to two mechanical tests in order to measure corneal strain resulting from application of cyclic (cyclic regimen) and constant (creep regimen) stress by changing the intra-ocular pressure (IOP) within physiological ranges (18-42 mmHg). Corneal shape changes were analyzed as a function of IOP and both corneal stress-strain curves and creep curves were generated.Results: The procedure was highly accurate and repeatable. Upon cyclic stress application, a biomechanical corneal elasticity gradient was found in the front-back direction. The average Young's modulus of the anterior cornea ranged between 2.28±0.87 MPa and 3.30±0.90 MPa in specimens with and without intact epithelium (P = 0.05) respectively. The Young's modulus of the posterior cornea was on average 0.21±0.09 MPa and 0.17±0.06 MPa (P>0.05) respectively. The time-dependent strain response of the cornea to creep testing was quantified by fitting data to a modified Zener model for extracting both the relaxation time and compliance function.Conclusion: Cyclic and creep mechanical tests are valuable for investigating the strain response of the intact human cornea within physiological IOP ranges, providing meaningful results that can be translated to clinic. The presence of epithelium influences the results of anterior corneal shape changes when monitoring deformation via Scheimpflug imaging in inflation experiments of whole eye globes.

AB - Purpose: To demonstrate a Scheimpflug-based imaging procedure for investigating the depth- and time-dependent strain response of the human cornea to inflation testing of whole eye globes.Methods: Six specimens, three of which with intact corneal epithelium, were mounted in a customized apparatus within a humidity and temperature-monitored wet chamber. Each specimen was subjected to two mechanical tests in order to measure corneal strain resulting from application of cyclic (cyclic regimen) and constant (creep regimen) stress by changing the intra-ocular pressure (IOP) within physiological ranges (18-42 mmHg). Corneal shape changes were analyzed as a function of IOP and both corneal stress-strain curves and creep curves were generated.Results: The procedure was highly accurate and repeatable. Upon cyclic stress application, a biomechanical corneal elasticity gradient was found in the front-back direction. The average Young's modulus of the anterior cornea ranged between 2.28±0.87 MPa and 3.30±0.90 MPa in specimens with and without intact epithelium (P = 0.05) respectively. The Young's modulus of the posterior cornea was on average 0.21±0.09 MPa and 0.17±0.06 MPa (P>0.05) respectively. The time-dependent strain response of the cornea to creep testing was quantified by fitting data to a modified Zener model for extracting both the relaxation time and compliance function.Conclusion: Cyclic and creep mechanical tests are valuable for investigating the strain response of the intact human cornea within physiological IOP ranges, providing meaningful results that can be translated to clinic. The presence of epithelium influences the results of anterior corneal shape changes when monitoring deformation via Scheimpflug imaging in inflation experiments of whole eye globes.

UR - http://www.scopus.com/inward/record.url?scp=84911912170&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84911912170&partnerID=8YFLogxK

U2 - 10.1371/journal.pone.0112169

DO - 10.1371/journal.pone.0112169

M3 - Article

VL - 9

JO - PLoS One

JF - PLoS One

SN - 1932-6203

IS - 11

M1 - 0112169

ER -