Abstract
An evaluation of AWOSEM with a Point Spread Function model included in the system matrix for spatial resolution improvement in Positron Emission Tomography was presented. The algorithm performances were evaluated on an experimentally acquired anthropomorphic phantom and on two lung cancer patient data. The study confirmed that the proposed strategy is able to improve quantitative accuracy, but that the amount of the achievable improvement and the corresponding parameter settings are object-dependent.
Original language | English |
---|---|
Pages (from-to) | 6548-6551 |
Number of pages | 4 |
Journal | Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Conference |
Publication status | Published - 2007 |
ASJC Scopus subject areas
- Computer Vision and Pattern Recognition
- Signal Processing
- Biomedical Engineering
- Health Informatics