Attenuation assessment of medical protective eyewear: The AVEN experience

M. Bertolini, G. Benecchi, M. Amici, A. Piola, V. Piccagli, C. Giordano, L. Nocetti

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The goal of this paper is to test the attenuation capability of seven models of protective eyewear used in routine clinical practice. Scattered radiation from a standard patient was simulated by using a water tank located over the treatment couch of a GE Innova 3100 x-ray angiography system. Seven protective eyewear models were tested using an anthropomorphic phantom mimicking the first operator. At each test, 4 thermoluminiscent dosimeters were placed on the phantom (respectively in front of the protective eyewear, under the eyewear, on the left earpiece and at chest level) in order to have an eyewear-independent reference. A test session without glasses was also acquired. Each model was tested with standard posterior-anterior (PA) projections and the two most common protective eyewear were tested using LAO90° and LAO45°CRA30° projections. A worst-case scenario was created to be sure of having an upper limit for the assessment of eyewear attenuation in routine clinical practice. In PA projections, the absolute attenuation value ranged between 71% and 81%, while relative attenuation between dose measured at eye lens and that measured at eyewear earpiece ranged from 67% to 85%. The slightly wider range was probably due to scatter radiation variability; anyway, differences are still included in the variable uncertainty of experimental measurements. It is worth noting that #3 eyewear model (the one without lateral protection) allows an attenuation similar to that of #5 eyewear model (with 0.5 mm lead lateral protection) in LAO90° and LAO45°CRA30° projections. Despite the experimental limitations, a description of the radiation properties of protective eyewear concerning radiation attenuation can be useful to rely on protection devices which can be used in routine clinical practice.

Original languageEnglish
Pages (from-to)279-289
Number of pages11
JournalJournal of Radiological Protection
Volume36
Issue number2
DOIs
Publication statusPublished - Jun 1 2016

Fingerprint

Radiation
Crystalline Lens
Uncertainty
Glass
Angiography
Thorax
X-Rays
Equipment and Supplies
glass
Water
radiation
test
Therapeutics
water
Radiation Dosimeters

Keywords

  • Eye lens protection
  • Protective Eyewear
  • Radiation protection

ASJC Scopus subject areas

  • Public Health, Environmental and Occupational Health
  • Waste Management and Disposal

Cite this

Bertolini, M., Benecchi, G., Amici, M., Piola, A., Piccagli, V., Giordano, C., & Nocetti, L. (2016). Attenuation assessment of medical protective eyewear: The AVEN experience. Journal of Radiological Protection, 36(2), 279-289. https://doi.org/10.1088/0952-4746/36/2/279

Attenuation assessment of medical protective eyewear : The AVEN experience. / Bertolini, M.; Benecchi, G.; Amici, M.; Piola, A.; Piccagli, V.; Giordano, C.; Nocetti, L.

In: Journal of Radiological Protection, Vol. 36, No. 2, 01.06.2016, p. 279-289.

Research output: Contribution to journalArticle

Bertolini, M, Benecchi, G, Amici, M, Piola, A, Piccagli, V, Giordano, C & Nocetti, L 2016, 'Attenuation assessment of medical protective eyewear: The AVEN experience', Journal of Radiological Protection, vol. 36, no. 2, pp. 279-289. https://doi.org/10.1088/0952-4746/36/2/279
Bertolini, M. ; Benecchi, G. ; Amici, M. ; Piola, A. ; Piccagli, V. ; Giordano, C. ; Nocetti, L. / Attenuation assessment of medical protective eyewear : The AVEN experience. In: Journal of Radiological Protection. 2016 ; Vol. 36, No. 2. pp. 279-289.
@article{03ef1bf5ed6848fcab99866d7f7ccb5a,
title = "Attenuation assessment of medical protective eyewear: The AVEN experience",
abstract = "The goal of this paper is to test the attenuation capability of seven models of protective eyewear used in routine clinical practice. Scattered radiation from a standard patient was simulated by using a water tank located over the treatment couch of a GE Innova 3100 x-ray angiography system. Seven protective eyewear models were tested using an anthropomorphic phantom mimicking the first operator. At each test, 4 thermoluminiscent dosimeters were placed on the phantom (respectively in front of the protective eyewear, under the eyewear, on the left earpiece and at chest level) in order to have an eyewear-independent reference. A test session without glasses was also acquired. Each model was tested with standard posterior-anterior (PA) projections and the two most common protective eyewear were tested using LAO90° and LAO45°CRA30° projections. A worst-case scenario was created to be sure of having an upper limit for the assessment of eyewear attenuation in routine clinical practice. In PA projections, the absolute attenuation value ranged between 71{\%} and 81{\%}, while relative attenuation between dose measured at eye lens and that measured at eyewear earpiece ranged from 67{\%} to 85{\%}. The slightly wider range was probably due to scatter radiation variability; anyway, differences are still included in the variable uncertainty of experimental measurements. It is worth noting that #3 eyewear model (the one without lateral protection) allows an attenuation similar to that of #5 eyewear model (with 0.5 mm lead lateral protection) in LAO90° and LAO45°CRA30° projections. Despite the experimental limitations, a description of the radiation properties of protective eyewear concerning radiation attenuation can be useful to rely on protection devices which can be used in routine clinical practice.",
keywords = "Eye lens protection, Protective Eyewear, Radiation protection",
author = "M. Bertolini and G. Benecchi and M. Amici and A. Piola and V. Piccagli and C. Giordano and L. Nocetti",
year = "2016",
month = "6",
day = "1",
doi = "10.1088/0952-4746/36/2/279",
language = "English",
volume = "36",
pages = "279--289",
journal = "Journal of Radiological Protection",
issn = "0952-4746",
publisher = "IOP Publishing Ltd.",
number = "2",

}

TY - JOUR

T1 - Attenuation assessment of medical protective eyewear

T2 - The AVEN experience

AU - Bertolini, M.

AU - Benecchi, G.

AU - Amici, M.

AU - Piola, A.

AU - Piccagli, V.

AU - Giordano, C.

AU - Nocetti, L.

PY - 2016/6/1

Y1 - 2016/6/1

N2 - The goal of this paper is to test the attenuation capability of seven models of protective eyewear used in routine clinical practice. Scattered radiation from a standard patient was simulated by using a water tank located over the treatment couch of a GE Innova 3100 x-ray angiography system. Seven protective eyewear models were tested using an anthropomorphic phantom mimicking the first operator. At each test, 4 thermoluminiscent dosimeters were placed on the phantom (respectively in front of the protective eyewear, under the eyewear, on the left earpiece and at chest level) in order to have an eyewear-independent reference. A test session without glasses was also acquired. Each model was tested with standard posterior-anterior (PA) projections and the two most common protective eyewear were tested using LAO90° and LAO45°CRA30° projections. A worst-case scenario was created to be sure of having an upper limit for the assessment of eyewear attenuation in routine clinical practice. In PA projections, the absolute attenuation value ranged between 71% and 81%, while relative attenuation between dose measured at eye lens and that measured at eyewear earpiece ranged from 67% to 85%. The slightly wider range was probably due to scatter radiation variability; anyway, differences are still included in the variable uncertainty of experimental measurements. It is worth noting that #3 eyewear model (the one without lateral protection) allows an attenuation similar to that of #5 eyewear model (with 0.5 mm lead lateral protection) in LAO90° and LAO45°CRA30° projections. Despite the experimental limitations, a description of the radiation properties of protective eyewear concerning radiation attenuation can be useful to rely on protection devices which can be used in routine clinical practice.

AB - The goal of this paper is to test the attenuation capability of seven models of protective eyewear used in routine clinical practice. Scattered radiation from a standard patient was simulated by using a water tank located over the treatment couch of a GE Innova 3100 x-ray angiography system. Seven protective eyewear models were tested using an anthropomorphic phantom mimicking the first operator. At each test, 4 thermoluminiscent dosimeters were placed on the phantom (respectively in front of the protective eyewear, under the eyewear, on the left earpiece and at chest level) in order to have an eyewear-independent reference. A test session without glasses was also acquired. Each model was tested with standard posterior-anterior (PA) projections and the two most common protective eyewear were tested using LAO90° and LAO45°CRA30° projections. A worst-case scenario was created to be sure of having an upper limit for the assessment of eyewear attenuation in routine clinical practice. In PA projections, the absolute attenuation value ranged between 71% and 81%, while relative attenuation between dose measured at eye lens and that measured at eyewear earpiece ranged from 67% to 85%. The slightly wider range was probably due to scatter radiation variability; anyway, differences are still included in the variable uncertainty of experimental measurements. It is worth noting that #3 eyewear model (the one without lateral protection) allows an attenuation similar to that of #5 eyewear model (with 0.5 mm lead lateral protection) in LAO90° and LAO45°CRA30° projections. Despite the experimental limitations, a description of the radiation properties of protective eyewear concerning radiation attenuation can be useful to rely on protection devices which can be used in routine clinical practice.

KW - Eye lens protection

KW - Protective Eyewear

KW - Radiation protection

UR - http://www.scopus.com/inward/record.url?scp=84974602106&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84974602106&partnerID=8YFLogxK

U2 - 10.1088/0952-4746/36/2/279

DO - 10.1088/0952-4746/36/2/279

M3 - Article

AN - SCOPUS:84974602106

VL - 36

SP - 279

EP - 289

JO - Journal of Radiological Protection

JF - Journal of Radiological Protection

SN - 0952-4746

IS - 2

ER -