Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model

Tasneem Z. Naqvi, Maurice Buchbinder, David Zarbatany, Julie Logan, Mia Molloy, Glenn Balke, Robert Ainsworth, John G. Webb, Ottavio Alfieri, Francesco Maisano

Research output: Contribution to journalArticle

Abstract

Background: The edge-to-edge (Alfieri) technique for mitral valve repair is a versatile method of treating mitral insufficiency. Because of its simplicity, it has been applied in minimally invasive surgery, and recently, in the design of endovascular closed-heart devices. Aim: The purpose of this study was to evaluate the acute in-vivo safety and feasibility of a novel percutaneous mitral valve repair system based on Alfieri technique in an animal model. Methods: Under general anesthesia, 11 pigs (90-100 kgs), under-went percutaneous Alfeiri procedure. The right femoral vein was punctured and the mitral valve was approached via a standard transeptal puncture. Combined intracardiac echo and fluoroscopic guidance was used. The procedure included: the positioning of a guide catheter for multiple access to the left atrium and for directing devices; the use of a therapy device to capture the free edge of the mitral valve leaflets using vacuum, and to deliver the suture to the valve and finally the fixation with a Nitinol suture clip, and trimming of the suture with a fastener catheter. Results: Leaflet capture, suture placement, and suture-clip deployment was successful in all 11 animals. There were no acute cardiac or access site complications. Procedural time (from wire in left atrium to completion of the procedure was 18 ± 9 min (range 9-30 min). Blood loss was 67 ± 44 ml (range 0-125ml). A double orifice configuration was visible by echocardiography at the end of the procedure in all animals. Conclusion: This acute animal study demonstrated the feasibility of a beating heart percutaneous Alfieri procedure in a non-diseased porcine valve using an endovascular suturing device to safely access the mitral valve, place a stitch through the mitral valve leaflets, and deploy a suture-clip that reproduces the surgical technique. Clinical application of this device in humans needs to be evaluated.

Original languageEnglish
Pages (from-to)525-531
Number of pages7
JournalCatheterization and Cardiovascular Interventions
Volume69
Issue number4
DOIs
Publication statusPublished - Mar 1 2007

Fingerprint

Mitral Valve
Sutures
Animal Models
Equipment and Supplies
Surgical Instruments
Heart Atria
Swine
Catheters
Femoral Vein
Minimally Invasive Surgical Procedures
Mitral Valve Insufficiency
Feasibility Studies
Vacuum
Punctures
General Anesthesia
Echocardiography
Safety

Keywords

  • Intracardiac echo
  • Transeptal cath
  • Valvular heart disease

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Radiology Nuclear Medicine and imaging

Cite this

Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model. / Naqvi, Tasneem Z.; Buchbinder, Maurice; Zarbatany, David; Logan, Julie; Molloy, Mia; Balke, Glenn; Ainsworth, Robert; Webb, John G.; Alfieri, Ottavio; Maisano, Francesco.

In: Catheterization and Cardiovascular Interventions, Vol. 69, No. 4, 01.03.2007, p. 525-531.

Research output: Contribution to journalArticle

Naqvi, TZ, Buchbinder, M, Zarbatany, D, Logan, J, Molloy, M, Balke, G, Ainsworth, R, Webb, JG, Alfieri, O & Maisano, F 2007, 'Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model', Catheterization and Cardiovascular Interventions, vol. 69, no. 4, pp. 525-531. https://doi.org/10.1002/ccd.21029
Naqvi, Tasneem Z. ; Buchbinder, Maurice ; Zarbatany, David ; Logan, Julie ; Molloy, Mia ; Balke, Glenn ; Ainsworth, Robert ; Webb, John G. ; Alfieri, Ottavio ; Maisano, Francesco. / Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model. In: Catheterization and Cardiovascular Interventions. 2007 ; Vol. 69, No. 4. pp. 525-531.
@article{4384de5c9a4d43608928451fc141a245,
title = "Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model",
abstract = "Background: The edge-to-edge (Alfieri) technique for mitral valve repair is a versatile method of treating mitral insufficiency. Because of its simplicity, it has been applied in minimally invasive surgery, and recently, in the design of endovascular closed-heart devices. Aim: The purpose of this study was to evaluate the acute in-vivo safety and feasibility of a novel percutaneous mitral valve repair system based on Alfieri technique in an animal model. Methods: Under general anesthesia, 11 pigs (90-100 kgs), under-went percutaneous Alfeiri procedure. The right femoral vein was punctured and the mitral valve was approached via a standard transeptal puncture. Combined intracardiac echo and fluoroscopic guidance was used. The procedure included: the positioning of a guide catheter for multiple access to the left atrium and for directing devices; the use of a therapy device to capture the free edge of the mitral valve leaflets using vacuum, and to deliver the suture to the valve and finally the fixation with a Nitinol suture clip, and trimming of the suture with a fastener catheter. Results: Leaflet capture, suture placement, and suture-clip deployment was successful in all 11 animals. There were no acute cardiac or access site complications. Procedural time (from wire in left atrium to completion of the procedure was 18 ± 9 min (range 9-30 min). Blood loss was 67 ± 44 ml (range 0-125ml). A double orifice configuration was visible by echocardiography at the end of the procedure in all animals. Conclusion: This acute animal study demonstrated the feasibility of a beating heart percutaneous Alfieri procedure in a non-diseased porcine valve using an endovascular suturing device to safely access the mitral valve, place a stitch through the mitral valve leaflets, and deploy a suture-clip that reproduces the surgical technique. Clinical application of this device in humans needs to be evaluated.",
keywords = "Intracardiac echo, Transeptal cath, Valvular heart disease",
author = "Naqvi, {Tasneem Z.} and Maurice Buchbinder and David Zarbatany and Julie Logan and Mia Molloy and Glenn Balke and Robert Ainsworth and Webb, {John G.} and Ottavio Alfieri and Francesco Maisano",
year = "2007",
month = "3",
day = "1",
doi = "10.1002/ccd.21029",
language = "English",
volume = "69",
pages = "525--531",
journal = "Catheterization and Cardiovascular Interventions",
issn = "1522-1946",
publisher = "John Wiley and Sons Inc.",
number = "4",

}

TY - JOUR

T1 - Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model

AU - Naqvi, Tasneem Z.

AU - Buchbinder, Maurice

AU - Zarbatany, David

AU - Logan, Julie

AU - Molloy, Mia

AU - Balke, Glenn

AU - Ainsworth, Robert

AU - Webb, John G.

AU - Alfieri, Ottavio

AU - Maisano, Francesco

PY - 2007/3/1

Y1 - 2007/3/1

N2 - Background: The edge-to-edge (Alfieri) technique for mitral valve repair is a versatile method of treating mitral insufficiency. Because of its simplicity, it has been applied in minimally invasive surgery, and recently, in the design of endovascular closed-heart devices. Aim: The purpose of this study was to evaluate the acute in-vivo safety and feasibility of a novel percutaneous mitral valve repair system based on Alfieri technique in an animal model. Methods: Under general anesthesia, 11 pigs (90-100 kgs), under-went percutaneous Alfeiri procedure. The right femoral vein was punctured and the mitral valve was approached via a standard transeptal puncture. Combined intracardiac echo and fluoroscopic guidance was used. The procedure included: the positioning of a guide catheter for multiple access to the left atrium and for directing devices; the use of a therapy device to capture the free edge of the mitral valve leaflets using vacuum, and to deliver the suture to the valve and finally the fixation with a Nitinol suture clip, and trimming of the suture with a fastener catheter. Results: Leaflet capture, suture placement, and suture-clip deployment was successful in all 11 animals. There were no acute cardiac or access site complications. Procedural time (from wire in left atrium to completion of the procedure was 18 ± 9 min (range 9-30 min). Blood loss was 67 ± 44 ml (range 0-125ml). A double orifice configuration was visible by echocardiography at the end of the procedure in all animals. Conclusion: This acute animal study demonstrated the feasibility of a beating heart percutaneous Alfieri procedure in a non-diseased porcine valve using an endovascular suturing device to safely access the mitral valve, place a stitch through the mitral valve leaflets, and deploy a suture-clip that reproduces the surgical technique. Clinical application of this device in humans needs to be evaluated.

AB - Background: The edge-to-edge (Alfieri) technique for mitral valve repair is a versatile method of treating mitral insufficiency. Because of its simplicity, it has been applied in minimally invasive surgery, and recently, in the design of endovascular closed-heart devices. Aim: The purpose of this study was to evaluate the acute in-vivo safety and feasibility of a novel percutaneous mitral valve repair system based on Alfieri technique in an animal model. Methods: Under general anesthesia, 11 pigs (90-100 kgs), under-went percutaneous Alfeiri procedure. The right femoral vein was punctured and the mitral valve was approached via a standard transeptal puncture. Combined intracardiac echo and fluoroscopic guidance was used. The procedure included: the positioning of a guide catheter for multiple access to the left atrium and for directing devices; the use of a therapy device to capture the free edge of the mitral valve leaflets using vacuum, and to deliver the suture to the valve and finally the fixation with a Nitinol suture clip, and trimming of the suture with a fastener catheter. Results: Leaflet capture, suture placement, and suture-clip deployment was successful in all 11 animals. There were no acute cardiac or access site complications. Procedural time (from wire in left atrium to completion of the procedure was 18 ± 9 min (range 9-30 min). Blood loss was 67 ± 44 ml (range 0-125ml). A double orifice configuration was visible by echocardiography at the end of the procedure in all animals. Conclusion: This acute animal study demonstrated the feasibility of a beating heart percutaneous Alfieri procedure in a non-diseased porcine valve using an endovascular suturing device to safely access the mitral valve, place a stitch through the mitral valve leaflets, and deploy a suture-clip that reproduces the surgical technique. Clinical application of this device in humans needs to be evaluated.

KW - Intracardiac echo

KW - Transeptal cath

KW - Valvular heart disease

UR - http://www.scopus.com/inward/record.url?scp=33847622562&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33847622562&partnerID=8YFLogxK

U2 - 10.1002/ccd.21029

DO - 10.1002/ccd.21029

M3 - Article

C2 - 17323355

AN - SCOPUS:33847622562

VL - 69

SP - 525

EP - 531

JO - Catheterization and Cardiovascular Interventions

JF - Catheterization and Cardiovascular Interventions

SN - 1522-1946

IS - 4

ER -