Bi-unicondylar knee replacement laxity with changes to simulated soft tissue constraints

Brendan L. Roach, Jonathan B. Matheny, Michele Spinelli, Roy J. Rusly, Saverio Affatato, John D. DesJardins

Research output: Contribution to journalArticle

Abstract

Unicondylar knee replacement systems have been shown to perform comparably to total knee replacements, while being much less surgically invasive. Proper ligament balancing, as well as knee laxity, has been shown to play an important role in optimizing kinematic behavior of these implant systems and improving long-term survival of the implant. This study investigates the effect of different simulated ligament laxity conditions of the anterior cruciate ligament and the posterior cruciate ligament on the resulting anteroposterior and mediolateral contact kinematics for medial and lateral pairs of UKR implants with flat and symmetric ultrahigh-molecular-weight polyethylene inserts during force-controlled ISO-14243-1 knee testing simulation. A novel method of capturing the tibiofemoral lowest point contact path was used to calculate the shear plane lowest point contact path kinematics in both the anteroposterior and the mediolateral directions. The results illustrated that multiple clinically relevant soft tissue configurations produce statistically different measured knee kinematics in unicondylar knee replacement systems than is seen in accepted "standard" knee simulator protocols with 95% confidence interval. The observed kinematic differences in anteroposterior and mediolateral movement from what was observed using standard wear testing protocols could aid in the development of unicondylar knee replacement design enhancements that are resistant to varying soft tissue deficiencies.

Original languageEnglish
Pages (from-to)27-36
Number of pages10
JournalProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Volume227
Issue number1
DOIs
Publication statusPublished - Jan 2013

Fingerprint

Knee prostheses
Knee Replacement Arthroplasties
Biomechanical Phenomena
Ligaments
Kinematics
Tissue
Knee
Point contacts
Posterior Cruciate Ligament
Ultrahigh molecular weight polyethylenes
Anterior Cruciate Ligament
Testing
Simulators
Wear of materials
Confidence Intervals

Keywords

  • Contact pathway
  • Knee kinematics
  • Soft tissue laxity
  • Total knee arthroplasty
  • Unicondylar

ASJC Scopus subject areas

  • Mechanical Engineering
  • Medicine(all)

Cite this

Bi-unicondylar knee replacement laxity with changes to simulated soft tissue constraints. / Roach, Brendan L.; Matheny, Jonathan B.; Spinelli, Michele; Rusly, Roy J.; Affatato, Saverio; DesJardins, John D.

In: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, Vol. 227, No. 1, 01.2013, p. 27-36.

Research output: Contribution to journalArticle

Roach, Brendan L. ; Matheny, Jonathan B. ; Spinelli, Michele ; Rusly, Roy J. ; Affatato, Saverio ; DesJardins, John D. / Bi-unicondylar knee replacement laxity with changes to simulated soft tissue constraints. In: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2013 ; Vol. 227, No. 1. pp. 27-36.
@article{b70ea57ee1f0444a94a728694088f8f2,
title = "Bi-unicondylar knee replacement laxity with changes to simulated soft tissue constraints",
abstract = "Unicondylar knee replacement systems have been shown to perform comparably to total knee replacements, while being much less surgically invasive. Proper ligament balancing, as well as knee laxity, has been shown to play an important role in optimizing kinematic behavior of these implant systems and improving long-term survival of the implant. This study investigates the effect of different simulated ligament laxity conditions of the anterior cruciate ligament and the posterior cruciate ligament on the resulting anteroposterior and mediolateral contact kinematics for medial and lateral pairs of UKR implants with flat and symmetric ultrahigh-molecular-weight polyethylene inserts during force-controlled ISO-14243-1 knee testing simulation. A novel method of capturing the tibiofemoral lowest point contact path was used to calculate the shear plane lowest point contact path kinematics in both the anteroposterior and the mediolateral directions. The results illustrated that multiple clinically relevant soft tissue configurations produce statistically different measured knee kinematics in unicondylar knee replacement systems than is seen in accepted {"}standard{"} knee simulator protocols with 95{\%} confidence interval. The observed kinematic differences in anteroposterior and mediolateral movement from what was observed using standard wear testing protocols could aid in the development of unicondylar knee replacement design enhancements that are resistant to varying soft tissue deficiencies.",
keywords = "Contact pathway, Knee kinematics, Soft tissue laxity, Total knee arthroplasty, Unicondylar",
author = "Roach, {Brendan L.} and Matheny, {Jonathan B.} and Michele Spinelli and Rusly, {Roy J.} and Saverio Affatato and DesJardins, {John D.}",
year = "2013",
month = "1",
doi = "10.1177/0954411912459420",
language = "English",
volume = "227",
pages = "27--36",
journal = "Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine",
issn = "0954-4119",
publisher = "SAGE Publications Ltd",
number = "1",

}

TY - JOUR

T1 - Bi-unicondylar knee replacement laxity with changes to simulated soft tissue constraints

AU - Roach, Brendan L.

AU - Matheny, Jonathan B.

AU - Spinelli, Michele

AU - Rusly, Roy J.

AU - Affatato, Saverio

AU - DesJardins, John D.

PY - 2013/1

Y1 - 2013/1

N2 - Unicondylar knee replacement systems have been shown to perform comparably to total knee replacements, while being much less surgically invasive. Proper ligament balancing, as well as knee laxity, has been shown to play an important role in optimizing kinematic behavior of these implant systems and improving long-term survival of the implant. This study investigates the effect of different simulated ligament laxity conditions of the anterior cruciate ligament and the posterior cruciate ligament on the resulting anteroposterior and mediolateral contact kinematics for medial and lateral pairs of UKR implants with flat and symmetric ultrahigh-molecular-weight polyethylene inserts during force-controlled ISO-14243-1 knee testing simulation. A novel method of capturing the tibiofemoral lowest point contact path was used to calculate the shear plane lowest point contact path kinematics in both the anteroposterior and the mediolateral directions. The results illustrated that multiple clinically relevant soft tissue configurations produce statistically different measured knee kinematics in unicondylar knee replacement systems than is seen in accepted "standard" knee simulator protocols with 95% confidence interval. The observed kinematic differences in anteroposterior and mediolateral movement from what was observed using standard wear testing protocols could aid in the development of unicondylar knee replacement design enhancements that are resistant to varying soft tissue deficiencies.

AB - Unicondylar knee replacement systems have been shown to perform comparably to total knee replacements, while being much less surgically invasive. Proper ligament balancing, as well as knee laxity, has been shown to play an important role in optimizing kinematic behavior of these implant systems and improving long-term survival of the implant. This study investigates the effect of different simulated ligament laxity conditions of the anterior cruciate ligament and the posterior cruciate ligament on the resulting anteroposterior and mediolateral contact kinematics for medial and lateral pairs of UKR implants with flat and symmetric ultrahigh-molecular-weight polyethylene inserts during force-controlled ISO-14243-1 knee testing simulation. A novel method of capturing the tibiofemoral lowest point contact path was used to calculate the shear plane lowest point contact path kinematics in both the anteroposterior and the mediolateral directions. The results illustrated that multiple clinically relevant soft tissue configurations produce statistically different measured knee kinematics in unicondylar knee replacement systems than is seen in accepted "standard" knee simulator protocols with 95% confidence interval. The observed kinematic differences in anteroposterior and mediolateral movement from what was observed using standard wear testing protocols could aid in the development of unicondylar knee replacement design enhancements that are resistant to varying soft tissue deficiencies.

KW - Contact pathway

KW - Knee kinematics

KW - Soft tissue laxity

KW - Total knee arthroplasty

KW - Unicondylar

UR - http://www.scopus.com/inward/record.url?scp=84877101032&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84877101032&partnerID=8YFLogxK

U2 - 10.1177/0954411912459420

DO - 10.1177/0954411912459420

M3 - Article

C2 - 23516953

AN - SCOPUS:84877101032

VL - 227

SP - 27

EP - 36

JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine

JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine

SN - 0954-4119

IS - 1

ER -