Biomechanical validation of a new nail-plate for the repair of stable proximal femoral fractures

M. Ramer, M. Viceconti, A. Toni, F. Pipino, A. Giunti

Research output: Contribution to journalArticle

Abstract

The mechanical performance and the stress shielding effects of the Howmedica Spherolock MKII implant were evaluated. Three types of stable proximal femoral fractures were created in composite femurs and fixed with the nail-plate. The femurs were loaded to failure, and the yield strength, stiffness, and failure modes were noted. Additional intact composite femurs were fitted proximally with strain gauges, and the strain was examined under load. The femurs then underwent intertrochanteric fracture, plating, and re-testing. Comparative analysis showed that while the Spherolock system is less strong and stiff than other commonly used implants, it provides excellent and uniform load transfer across the fracture site. Varus rotation of the femoral head under load tended to open the fracture gap and localize the resultant load medially. Stress shielding of the calcar was less than 25% of the applied load even in the worst case. This very low stress shielding behavior was attributed mainly to the implant's low stiffness.

Original languageEnglish
Pages (from-to)137-142
Number of pages6
JournalArchives of Orthopaedic and Trauma Surgery
Volume116
Issue number3
DOIs
Publication statusPublished - 1997

Fingerprint

Femoral Fractures
Nails
Femur
Mechanical Stress
Open Fractures
Hip Fractures
Thigh

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

Biomechanical validation of a new nail-plate for the repair of stable proximal femoral fractures. / Ramer, M.; Viceconti, M.; Toni, A.; Pipino, F.; Giunti, A.

In: Archives of Orthopaedic and Trauma Surgery, Vol. 116, No. 3, 1997, p. 137-142.

Research output: Contribution to journalArticle

@article{a1bece889b3346b5ae280da20db17855,
title = "Biomechanical validation of a new nail-plate for the repair of stable proximal femoral fractures",
abstract = "The mechanical performance and the stress shielding effects of the Howmedica Spherolock MKII implant were evaluated. Three types of stable proximal femoral fractures were created in composite femurs and fixed with the nail-plate. The femurs were loaded to failure, and the yield strength, stiffness, and failure modes were noted. Additional intact composite femurs were fitted proximally with strain gauges, and the strain was examined under load. The femurs then underwent intertrochanteric fracture, plating, and re-testing. Comparative analysis showed that while the Spherolock system is less strong and stiff than other commonly used implants, it provides excellent and uniform load transfer across the fracture site. Varus rotation of the femoral head under load tended to open the fracture gap and localize the resultant load medially. Stress shielding of the calcar was less than 25{\%} of the applied load even in the worst case. This very low stress shielding behavior was attributed mainly to the implant's low stiffness.",
author = "M. Ramer and M. Viceconti and A. Toni and F. Pipino and A. Giunti",
year = "1997",
doi = "10.1007/s004020050088",
language = "English",
volume = "116",
pages = "137--142",
journal = "Archiv fur orthopadische und Unfall-Chirurgie",
issn = "0936-8051",
publisher = "Springer Verlag",
number = "3",

}

TY - JOUR

T1 - Biomechanical validation of a new nail-plate for the repair of stable proximal femoral fractures

AU - Ramer, M.

AU - Viceconti, M.

AU - Toni, A.

AU - Pipino, F.

AU - Giunti, A.

PY - 1997

Y1 - 1997

N2 - The mechanical performance and the stress shielding effects of the Howmedica Spherolock MKII implant were evaluated. Three types of stable proximal femoral fractures were created in composite femurs and fixed with the nail-plate. The femurs were loaded to failure, and the yield strength, stiffness, and failure modes were noted. Additional intact composite femurs were fitted proximally with strain gauges, and the strain was examined under load. The femurs then underwent intertrochanteric fracture, plating, and re-testing. Comparative analysis showed that while the Spherolock system is less strong and stiff than other commonly used implants, it provides excellent and uniform load transfer across the fracture site. Varus rotation of the femoral head under load tended to open the fracture gap and localize the resultant load medially. Stress shielding of the calcar was less than 25% of the applied load even in the worst case. This very low stress shielding behavior was attributed mainly to the implant's low stiffness.

AB - The mechanical performance and the stress shielding effects of the Howmedica Spherolock MKII implant were evaluated. Three types of stable proximal femoral fractures were created in composite femurs and fixed with the nail-plate. The femurs were loaded to failure, and the yield strength, stiffness, and failure modes were noted. Additional intact composite femurs were fitted proximally with strain gauges, and the strain was examined under load. The femurs then underwent intertrochanteric fracture, plating, and re-testing. Comparative analysis showed that while the Spherolock system is less strong and stiff than other commonly used implants, it provides excellent and uniform load transfer across the fracture site. Varus rotation of the femoral head under load tended to open the fracture gap and localize the resultant load medially. Stress shielding of the calcar was less than 25% of the applied load even in the worst case. This very low stress shielding behavior was attributed mainly to the implant's low stiffness.

UR - http://www.scopus.com/inward/record.url?scp=0031045612&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031045612&partnerID=8YFLogxK

U2 - 10.1007/s004020050088

DO - 10.1007/s004020050088

M3 - Article

C2 - 9061168

AN - SCOPUS:0031045612

VL - 116

SP - 137

EP - 142

JO - Archiv fur orthopadische und Unfall-Chirurgie

JF - Archiv fur orthopadische und Unfall-Chirurgie

SN - 0936-8051