Bone volume-to-total volume ratio measured in trabecular bone by single-sided NMR devices.

L Brizi, M Barbieri, Fabio Baruffaldi, V Bortolotti, C. Fersini, H. Liu, M Nogueira d'Eurydice , S Obruchkov , F Zong, P Galvosas, P. Fantazzini

Research output: Contribution to journalArticle

Abstract

PURPOSE:
Reduced bone strength is associated with a loss of bone mass, usually evaluated by dual-energy X-ray absorptiometry, although it is known that the bone microstructure also affects the bone strength. Here, a method is proposed to measure (in laboratory) the bone volume-to-total volume ratio by single-sided NMR scanners, which is related to the microstructure of the trabecular bone.
METHODS:
Three single-sided scanners were used on animal bone samples. These low-field, mobile, low-cost devices are able to detect the NMR signal, regardless of the sample sizes, without the use of ionizing radiations, with the further advantage of signal localization offered by their intrinsic magnetic field gradients.
RESULTS:
The performance of the different single-sided scanners have been discussed. The results have been compared with bone volume-to-total volume ratio by micro CT and MRI, obtaining consistent values.
CONCLUSIONS:
Our results demonstrate the feasibility of the method for laboratory analyses, which are useful for measurements like porosity on bone specimens. This can be considered as the first step to develop an NMR method based on the use of a mobile single-sided device, for the diagnosis of osteoporosis, through the acquisition of the signal from the appendicular skeleton, allowing for low-cost, wide screening campaigns.
Original languageEnglish
Pages (from-to)501-510
Number of pages10
JournalMagnetic Resonance in Medicine
Volume79
Issue number1
DOIs
Publication statusPublished - Apr 10 2017

Keywords

  • BV/TV
  • osteoporosis
  • portable NMR
  • screening campaigns
  • single-sided NMR

Fingerprint Dive into the research topics of 'Bone volume-to-total volume ratio measured in trabecular bone by single-sided NMR devices.'. Together they form a unique fingerprint.

  • Cite this

    Brizi, L., Barbieri, M., Baruffaldi, F., Bortolotti, V., Fersini, C., Liu, H., Nogueira d'Eurydice , M., Obruchkov , S., Zong, F., Galvosas, P., & Fantazzini, P. (2017). Bone volume-to-total volume ratio measured in trabecular bone by single-sided NMR devices. Magnetic Resonance in Medicine, 79(1), 501-510. https://doi.org/10.1002/mrm.26697