Clinical evaluation of an automatic blood-pressure monitoring device

P. A. Modesti, G. F. Gensini, C. Conti, G. G. Neri Serneri

Research output: Contribution to journalArticle

Abstract

Reliability and reproducibility of an automatic ambulatory blood pressure recorder (ICR-5200, Squibb) were tested on 45 subjects (24 normotensives and 21 hypertensives) aged 32-64 (mean age 49). Of 6,566 recordings, 2,376 measurements were cross-checked with a Riva Rocci sphygmomanometer in supine, standing, or sitting position at the beginning and at the end of a 24-hour monitoring period. Moreover, in 11 normotensives and in 11 hypertensives, cross-checking was performed during a cycloergometer effort test. A highly significant linear correlation was found in all the conditions tested (r = 0.98, p <0.001 for systolic and r = 0.98, p <0.001 for diastolic pressure) either in hypertensives or in normotensives. Only 8% of diastolic measurements performed by the two methods differed more than 5 mmHg. For each subject 5.4 ± 3.8% of the preprogrammed measurements were lost. Only seven out of 45 subjects referred a disturbed sleep. ICR-5200 seems a suitable and reliable tool for automatic blood-pressure recording.

Original languageEnglish
Pages (from-to)631-644
Number of pages14
JournalJournal of clinical hypertension (Greenwich, Conn.)
Volume3
Issue number4
Publication statusPublished - 1987

ASJC Scopus subject areas

  • Internal Medicine

Fingerprint Dive into the research topics of 'Clinical evaluation of an automatic blood-pressure monitoring device'. Together they form a unique fingerprint.

  • Cite this

    Modesti, P. A., Gensini, G. F., Conti, C., & Neri Serneri, G. G. (1987). Clinical evaluation of an automatic blood-pressure monitoring device. Journal of clinical hypertension (Greenwich, Conn.), 3(4), 631-644.