Reliability and reproducibility of an automatic ambulatory blood pressure recorder (ICR-5200, Squibb) were tested on 45 subjects (24 normotensives and 21 hypertensives) aged 32-64 (mean age 49). Of 6,566 recordings, 2,376 measurements were cross-checked with a Riva Rocci sphygmomanometer in supine, standing, or sitting position at the beginning and at the end of a 24-hour monitoring period. Moreover, in 11 normotensives and in 11 hypertensives, cross-checking was performed during a cycloergometer effort test. A highly significant linear correlation was found in all the conditions tested (r = 0.98, p <0.001 for systolic and r = 0.98, p <0.001 for diastolic pressure) either in hypertensives or in normotensives. Only 8% of diastolic measurements performed by the two methods differed more than 5 mmHg. For each subject 5.4 ± 3.8% of the preprogrammed measurements were lost. Only seven out of 45 subjects referred a disturbed sleep. ICR-5200 seems a suitable and reliable tool for automatic blood-pressure recording.

Original languageEnglish
Pages (from-to)631-644
Number of pages14
JournalJournal of clinical hypertension (Greenwich, Conn.)
Volume3
Issue number4
Publication statusPublished - 1987

Fingerprint

Blood Pressure
Posture
Equipment and Supplies
Sphygmomanometers
Supine Position
Sleep

ASJC Scopus subject areas

  • Internal Medicine

Cite this

Modesti, P. A., Gensini, G. F., Conti, C., & Neri Serneri, G. G. (1987). Clinical evaluation of an automatic blood-pressure monitoring device. Journal of clinical hypertension (Greenwich, Conn.), 3(4), 631-644.

Clinical evaluation of an automatic blood-pressure monitoring device. / Modesti, P. A.; Gensini, G. F.; Conti, C.; Neri Serneri, G. G.

In: Journal of clinical hypertension (Greenwich, Conn.), Vol. 3, No. 4, 1987, p. 631-644.

Research output: Contribution to journalArticle

Modesti, PA, Gensini, GF, Conti, C & Neri Serneri, GG 1987, 'Clinical evaluation of an automatic blood-pressure monitoring device', Journal of clinical hypertension (Greenwich, Conn.), vol. 3, no. 4, pp. 631-644.
Modesti, P. A. ; Gensini, G. F. ; Conti, C. ; Neri Serneri, G. G. / Clinical evaluation of an automatic blood-pressure monitoring device. In: Journal of clinical hypertension (Greenwich, Conn.). 1987 ; Vol. 3, No. 4. pp. 631-644.
@article{a4f9b56cc3d848e590b11805d647aca7,
title = "Clinical evaluation of an automatic blood-pressure monitoring device",
abstract = "Reliability and reproducibility of an automatic ambulatory blood pressure recorder (ICR-5200, Squibb) were tested on 45 subjects (24 normotensives and 21 hypertensives) aged 32-64 (mean age 49). Of 6,566 recordings, 2,376 measurements were cross-checked with a Riva Rocci sphygmomanometer in supine, standing, or sitting position at the beginning and at the end of a 24-hour monitoring period. Moreover, in 11 normotensives and in 11 hypertensives, cross-checking was performed during a cycloergometer effort test. A highly significant linear correlation was found in all the conditions tested (r = 0.98, p <0.001 for systolic and r = 0.98, p <0.001 for diastolic pressure) either in hypertensives or in normotensives. Only 8{\%} of diastolic measurements performed by the two methods differed more than 5 mmHg. For each subject 5.4 ± 3.8{\%} of the preprogrammed measurements were lost. Only seven out of 45 subjects referred a disturbed sleep. ICR-5200 seems a suitable and reliable tool for automatic blood-pressure recording.",
author = "Modesti, {P. A.} and Gensini, {G. F.} and C. Conti and {Neri Serneri}, {G. G.}",
year = "1987",
language = "English",
volume = "3",
pages = "631--644",
journal = "Journal of clinical hypertension",
issn = "1524-6175",
publisher = "Wiley-Blackwell",
number = "4",

}

TY - JOUR

T1 - Clinical evaluation of an automatic blood-pressure monitoring device

AU - Modesti, P. A.

AU - Gensini, G. F.

AU - Conti, C.

AU - Neri Serneri, G. G.

PY - 1987

Y1 - 1987

N2 - Reliability and reproducibility of an automatic ambulatory blood pressure recorder (ICR-5200, Squibb) were tested on 45 subjects (24 normotensives and 21 hypertensives) aged 32-64 (mean age 49). Of 6,566 recordings, 2,376 measurements were cross-checked with a Riva Rocci sphygmomanometer in supine, standing, or sitting position at the beginning and at the end of a 24-hour monitoring period. Moreover, in 11 normotensives and in 11 hypertensives, cross-checking was performed during a cycloergometer effort test. A highly significant linear correlation was found in all the conditions tested (r = 0.98, p <0.001 for systolic and r = 0.98, p <0.001 for diastolic pressure) either in hypertensives or in normotensives. Only 8% of diastolic measurements performed by the two methods differed more than 5 mmHg. For each subject 5.4 ± 3.8% of the preprogrammed measurements were lost. Only seven out of 45 subjects referred a disturbed sleep. ICR-5200 seems a suitable and reliable tool for automatic blood-pressure recording.

AB - Reliability and reproducibility of an automatic ambulatory blood pressure recorder (ICR-5200, Squibb) were tested on 45 subjects (24 normotensives and 21 hypertensives) aged 32-64 (mean age 49). Of 6,566 recordings, 2,376 measurements were cross-checked with a Riva Rocci sphygmomanometer in supine, standing, or sitting position at the beginning and at the end of a 24-hour monitoring period. Moreover, in 11 normotensives and in 11 hypertensives, cross-checking was performed during a cycloergometer effort test. A highly significant linear correlation was found in all the conditions tested (r = 0.98, p <0.001 for systolic and r = 0.98, p <0.001 for diastolic pressure) either in hypertensives or in normotensives. Only 8% of diastolic measurements performed by the two methods differed more than 5 mmHg. For each subject 5.4 ± 3.8% of the preprogrammed measurements were lost. Only seven out of 45 subjects referred a disturbed sleep. ICR-5200 seems a suitable and reliable tool for automatic blood-pressure recording.

UR - http://www.scopus.com/inward/record.url?scp=0023609591&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0023609591&partnerID=8YFLogxK

M3 - Article

VL - 3

SP - 631

EP - 644

JO - Journal of clinical hypertension

JF - Journal of clinical hypertension

SN - 1524-6175

IS - 4

ER -