Clinical variability in keratometry, ultrasound biometry measurements, and emmetropic intraocular lens power calculation

L. Pierro, G. Modorati, R. Brancato

Research output: Contribution to journalArticle

Abstract

We calculated inter-examiner and intra-examiner variability in keratometry and ultrasound biometry when performed by two different examiners. The potential effect on intraocular lens power calculation was also considered. Intra-examiner variability was statistically and clinically significant for axial measurements, particularly in eyes with axial length greater than 24 mm. Thus, variability in biometric examination affected the emmetropic intraocular lens power, especially in myopic eyes.

Original languageEnglish
Pages (from-to)91-94
Number of pages4
JournalJournal of Cataract and Refractive Surgery
Volume17
Issue number1
Publication statusPublished - 1991

ASJC Scopus subject areas

  • Ophthalmology

Fingerprint Dive into the research topics of 'Clinical variability in keratometry, ultrasound biometry measurements, and emmetropic intraocular lens power calculation'. Together they form a unique fingerprint.