Comparative analysis of explanted DBS electrodes

Michele Rizzi, Alessandro De Benedictis, Giuseppe Messina, Roberto Cordella, Davide Marchesi, Raffaella Messina, Federica Penner, Angelo Franzini, Carlo Efisio Marras

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Background: Hardware-related complications frequently occur in deep brain stimulation. Microscopy and spectroscopy techniques are effective methods for characterizing the morphological and chemical basis of malfunctioning DBS electrodes. A previous report by our team revealed the morphological and chemical alterations on a malfunctioning explanted electrode when it was compared to a new device. The aim of this preliminary study was to verify whether these morphological and chemical alterations in the materials were a direct result of the hardware malfunctioning or if the failure was correlated to a degradation process over time. Methods: Two DBS electrodes were removed from two patients for reasons other than DBS system impairment and were analyzed by a scanning electron microscope and by an energy-dispersive X-ray spectroscopy. The results were compared to a malfunctioning device and to a new device, previously analyzed by our group. Results: The analysis revealed that the wear of the polyurethane external part of all the electrodes was directly correlated with the duration of implantation period. Moreover, these alterations were independent from the electrodes functioning and from parameters used during therapy. Conclusions: This is the first study done that demonstrates a time-related degradation in the external layer of DBS electrodes. The analyses of morphological and chemical properties of the implanted devices are relevant for predicting the possibility of hardware’s impairment as well as to improve the bio-stability of DBS systems.

Original languageEnglish
Pages (from-to)2135-2141
Number of pages7
JournalActa Neurochirurgica
Volume157
Issue number12
DOIs
Publication statusPublished - Dec 1 2015

Fingerprint

Electrodes
Equipment and Supplies
X-Ray Emission Spectrometry
Deep Brain Stimulation
Polyurethanes
Microscopy
Spectrum Analysis
Electrons
Therapeutics

Keywords

  • DBS
  • Electrodes analysis
  • Energy dispersive X-ray
  • Environmental biodegradation
  • Explanted electrodes
  • Scanning electron microscope

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

Comparative analysis of explanted DBS electrodes. / Rizzi, Michele; De Benedictis, Alessandro; Messina, Giuseppe; Cordella, Roberto; Marchesi, Davide; Messina, Raffaella; Penner, Federica; Franzini, Angelo; Marras, Carlo Efisio.

In: Acta Neurochirurgica, Vol. 157, No. 12, 01.12.2015, p. 2135-2141.

Research output: Contribution to journalArticle

Rizzi, Michele ; De Benedictis, Alessandro ; Messina, Giuseppe ; Cordella, Roberto ; Marchesi, Davide ; Messina, Raffaella ; Penner, Federica ; Franzini, Angelo ; Marras, Carlo Efisio. / Comparative analysis of explanted DBS electrodes. In: Acta Neurochirurgica. 2015 ; Vol. 157, No. 12. pp. 2135-2141.
@article{26a1d1e36fd14ceba1c79a2d8c5d5f3f,
title = "Comparative analysis of explanted DBS electrodes",
abstract = "Background: Hardware-related complications frequently occur in deep brain stimulation. Microscopy and spectroscopy techniques are effective methods for characterizing the morphological and chemical basis of malfunctioning DBS electrodes. A previous report by our team revealed the morphological and chemical alterations on a malfunctioning explanted electrode when it was compared to a new device. The aim of this preliminary study was to verify whether these morphological and chemical alterations in the materials were a direct result of the hardware malfunctioning or if the failure was correlated to a degradation process over time. Methods: Two DBS electrodes were removed from two patients for reasons other than DBS system impairment and were analyzed by a scanning electron microscope and by an energy-dispersive X-ray spectroscopy. The results were compared to a malfunctioning device and to a new device, previously analyzed by our group. Results: The analysis revealed that the wear of the polyurethane external part of all the electrodes was directly correlated with the duration of implantation period. Moreover, these alterations were independent from the electrodes functioning and from parameters used during therapy. Conclusions: This is the first study done that demonstrates a time-related degradation in the external layer of DBS electrodes. The analyses of morphological and chemical properties of the implanted devices are relevant for predicting the possibility of hardware’s impairment as well as to improve the bio-stability of DBS systems.",
keywords = "DBS, Electrodes analysis, Energy dispersive X-ray, Environmental biodegradation, Explanted electrodes, Scanning electron microscope",
author = "Michele Rizzi and {De Benedictis}, Alessandro and Giuseppe Messina and Roberto Cordella and Davide Marchesi and Raffaella Messina and Federica Penner and Angelo Franzini and Marras, {Carlo Efisio}",
year = "2015",
month = "12",
day = "1",
doi = "10.1007/s00701-015-2572-8",
language = "English",
volume = "157",
pages = "2135--2141",
journal = "Acta Neurochirurgica",
issn = "0001-6268",
publisher = "Springer Wien",
number = "12",

}

TY - JOUR

T1 - Comparative analysis of explanted DBS electrodes

AU - Rizzi, Michele

AU - De Benedictis, Alessandro

AU - Messina, Giuseppe

AU - Cordella, Roberto

AU - Marchesi, Davide

AU - Messina, Raffaella

AU - Penner, Federica

AU - Franzini, Angelo

AU - Marras, Carlo Efisio

PY - 2015/12/1

Y1 - 2015/12/1

N2 - Background: Hardware-related complications frequently occur in deep brain stimulation. Microscopy and spectroscopy techniques are effective methods for characterizing the morphological and chemical basis of malfunctioning DBS electrodes. A previous report by our team revealed the morphological and chemical alterations on a malfunctioning explanted electrode when it was compared to a new device. The aim of this preliminary study was to verify whether these morphological and chemical alterations in the materials were a direct result of the hardware malfunctioning or if the failure was correlated to a degradation process over time. Methods: Two DBS electrodes were removed from two patients for reasons other than DBS system impairment and were analyzed by a scanning electron microscope and by an energy-dispersive X-ray spectroscopy. The results were compared to a malfunctioning device and to a new device, previously analyzed by our group. Results: The analysis revealed that the wear of the polyurethane external part of all the electrodes was directly correlated with the duration of implantation period. Moreover, these alterations were independent from the electrodes functioning and from parameters used during therapy. Conclusions: This is the first study done that demonstrates a time-related degradation in the external layer of DBS electrodes. The analyses of morphological and chemical properties of the implanted devices are relevant for predicting the possibility of hardware’s impairment as well as to improve the bio-stability of DBS systems.

AB - Background: Hardware-related complications frequently occur in deep brain stimulation. Microscopy and spectroscopy techniques are effective methods for characterizing the morphological and chemical basis of malfunctioning DBS electrodes. A previous report by our team revealed the morphological and chemical alterations on a malfunctioning explanted electrode when it was compared to a new device. The aim of this preliminary study was to verify whether these morphological and chemical alterations in the materials were a direct result of the hardware malfunctioning or if the failure was correlated to a degradation process over time. Methods: Two DBS electrodes were removed from two patients for reasons other than DBS system impairment and were analyzed by a scanning electron microscope and by an energy-dispersive X-ray spectroscopy. The results were compared to a malfunctioning device and to a new device, previously analyzed by our group. Results: The analysis revealed that the wear of the polyurethane external part of all the electrodes was directly correlated with the duration of implantation period. Moreover, these alterations were independent from the electrodes functioning and from parameters used during therapy. Conclusions: This is the first study done that demonstrates a time-related degradation in the external layer of DBS electrodes. The analyses of morphological and chemical properties of the implanted devices are relevant for predicting the possibility of hardware’s impairment as well as to improve the bio-stability of DBS systems.

KW - DBS

KW - Electrodes analysis

KW - Energy dispersive X-ray

KW - Environmental biodegradation

KW - Explanted electrodes

KW - Scanning electron microscope

UR - http://www.scopus.com/inward/record.url?scp=84947128277&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84947128277&partnerID=8YFLogxK

U2 - 10.1007/s00701-015-2572-8

DO - 10.1007/s00701-015-2572-8

M3 - Article

C2 - 26347045

AN - SCOPUS:84947128277

VL - 157

SP - 2135

EP - 2141

JO - Acta Neurochirurgica

JF - Acta Neurochirurgica

SN - 0001-6268

IS - 12

ER -