Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems

Tomaso Villa, Luigi La Barbera, Fabio Galbusera

Research output: Contribution to journalArticle

Abstract

Background context: Preclinical evaluation of the long-term reliability of devices for lumbar fixation is a mandatory activity before they are put into market. The experimental setups are described in two different standards edited by the International Organization for Standardization (ISO) and the American Society for Testing Materials (ASTM), but the evaluation of the suitability of such tests to simulate the actual loading with in vivo situations has never been performed. Purpose: To calculate through finite element (FE) simulations the stress in the rods of the fixator when subjected to ASTM and ISO standards. To compare the calculated stresses arising in the same fixator once it has been virtually mounted in a physiological environment and loaded with physiological forces and moments. Study design: FE simulations and validation experimental tests. Methods: FE models of the ISO and ASTM setups were created to conduct simulations of the tests prescribed by standards and calculate stresses in the rods. Validation of the simulations were performed through experimental tests; the same fixator was virtually mounted in an L2-L4 FE model of the lumbar spine and stresses in the rods were calculated when the spine was subjected to physiological forces and moments. Results: The comparison between FE simulations and experimental tests showed good agreement between results obtained using the two methodologies, thus confirming the suitability of the FE method to evaluate stresses in the device in different loading situations. The usage of a physiological load with ASTM standard is impossible due to the extreme severity of the ASTM configuration; in this circumstance, the presence of an anterior support is suggested. Also, ISO prescriptions, although the choice of the setup correctly simulates the mechanical contribution of the discs, seem to overstress the device as compared with a physiological loading condition. Some daily activities, other than walking, can induce a further state of stress in the device that should be taken into account in setting up new experimental procedures. Conclusions: ISO standard loading prescriptions seems to be more severe than the expected physiological ones. The ASTM standard should be completed by including some anterior supporting device and declaring the value of the load to be imposed. Moreover, a further enhancement of standards would be simulating other movements representative of daily activities different from walking.

Original languageEnglish
Pages (from-to)695-704
Number of pages10
JournalSpine Journal
Volume14
Issue number4
DOIs
Publication statusPublished - Apr 1 2014

Fingerprint

Fatigue
Materials Testing
Equipment and Supplies
Walking
Prescriptions
Spine

Keywords

  • Fatigue test
  • Finite element method
  • International standard
  • Spinal fixator
  • Validation

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. / Villa, Tomaso; La Barbera, Luigi; Galbusera, Fabio.

In: Spine Journal, Vol. 14, No. 4, 01.04.2014, p. 695-704.

Research output: Contribution to journalArticle

@article{0599bb31d81e4a40b769666a2361e9e8,
title = "Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems",
abstract = "Background context: Preclinical evaluation of the long-term reliability of devices for lumbar fixation is a mandatory activity before they are put into market. The experimental setups are described in two different standards edited by the International Organization for Standardization (ISO) and the American Society for Testing Materials (ASTM), but the evaluation of the suitability of such tests to simulate the actual loading with in vivo situations has never been performed. Purpose: To calculate through finite element (FE) simulations the stress in the rods of the fixator when subjected to ASTM and ISO standards. To compare the calculated stresses arising in the same fixator once it has been virtually mounted in a physiological environment and loaded with physiological forces and moments. Study design: FE simulations and validation experimental tests. Methods: FE models of the ISO and ASTM setups were created to conduct simulations of the tests prescribed by standards and calculate stresses in the rods. Validation of the simulations were performed through experimental tests; the same fixator was virtually mounted in an L2-L4 FE model of the lumbar spine and stresses in the rods were calculated when the spine was subjected to physiological forces and moments. Results: The comparison between FE simulations and experimental tests showed good agreement between results obtained using the two methodologies, thus confirming the suitability of the FE method to evaluate stresses in the device in different loading situations. The usage of a physiological load with ASTM standard is impossible due to the extreme severity of the ASTM configuration; in this circumstance, the presence of an anterior support is suggested. Also, ISO prescriptions, although the choice of the setup correctly simulates the mechanical contribution of the discs, seem to overstress the device as compared with a physiological loading condition. Some daily activities, other than walking, can induce a further state of stress in the device that should be taken into account in setting up new experimental procedures. Conclusions: ISO standard loading prescriptions seems to be more severe than the expected physiological ones. The ASTM standard should be completed by including some anterior supporting device and declaring the value of the load to be imposed. Moreover, a further enhancement of standards would be simulating other movements representative of daily activities different from walking.",
keywords = "Fatigue test, Finite element method, International standard, Spinal fixator, Validation",
author = "Tomaso Villa and {La Barbera}, Luigi and Fabio Galbusera",
year = "2014",
month = "4",
day = "1",
doi = "10.1016/j.spinee.2013.08.032",
language = "English",
volume = "14",
pages = "695--704",
journal = "Spine Journal",
issn = "1529-9430",
publisher = "Elsevier Inc.",
number = "4",

}

TY - JOUR

T1 - Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems

AU - Villa, Tomaso

AU - La Barbera, Luigi

AU - Galbusera, Fabio

PY - 2014/4/1

Y1 - 2014/4/1

N2 - Background context: Preclinical evaluation of the long-term reliability of devices for lumbar fixation is a mandatory activity before they are put into market. The experimental setups are described in two different standards edited by the International Organization for Standardization (ISO) and the American Society for Testing Materials (ASTM), but the evaluation of the suitability of such tests to simulate the actual loading with in vivo situations has never been performed. Purpose: To calculate through finite element (FE) simulations the stress in the rods of the fixator when subjected to ASTM and ISO standards. To compare the calculated stresses arising in the same fixator once it has been virtually mounted in a physiological environment and loaded with physiological forces and moments. Study design: FE simulations and validation experimental tests. Methods: FE models of the ISO and ASTM setups were created to conduct simulations of the tests prescribed by standards and calculate stresses in the rods. Validation of the simulations were performed through experimental tests; the same fixator was virtually mounted in an L2-L4 FE model of the lumbar spine and stresses in the rods were calculated when the spine was subjected to physiological forces and moments. Results: The comparison between FE simulations and experimental tests showed good agreement between results obtained using the two methodologies, thus confirming the suitability of the FE method to evaluate stresses in the device in different loading situations. The usage of a physiological load with ASTM standard is impossible due to the extreme severity of the ASTM configuration; in this circumstance, the presence of an anterior support is suggested. Also, ISO prescriptions, although the choice of the setup correctly simulates the mechanical contribution of the discs, seem to overstress the device as compared with a physiological loading condition. Some daily activities, other than walking, can induce a further state of stress in the device that should be taken into account in setting up new experimental procedures. Conclusions: ISO standard loading prescriptions seems to be more severe than the expected physiological ones. The ASTM standard should be completed by including some anterior supporting device and declaring the value of the load to be imposed. Moreover, a further enhancement of standards would be simulating other movements representative of daily activities different from walking.

AB - Background context: Preclinical evaluation of the long-term reliability of devices for lumbar fixation is a mandatory activity before they are put into market. The experimental setups are described in two different standards edited by the International Organization for Standardization (ISO) and the American Society for Testing Materials (ASTM), but the evaluation of the suitability of such tests to simulate the actual loading with in vivo situations has never been performed. Purpose: To calculate through finite element (FE) simulations the stress in the rods of the fixator when subjected to ASTM and ISO standards. To compare the calculated stresses arising in the same fixator once it has been virtually mounted in a physiological environment and loaded with physiological forces and moments. Study design: FE simulations and validation experimental tests. Methods: FE models of the ISO and ASTM setups were created to conduct simulations of the tests prescribed by standards and calculate stresses in the rods. Validation of the simulations were performed through experimental tests; the same fixator was virtually mounted in an L2-L4 FE model of the lumbar spine and stresses in the rods were calculated when the spine was subjected to physiological forces and moments. Results: The comparison between FE simulations and experimental tests showed good agreement between results obtained using the two methodologies, thus confirming the suitability of the FE method to evaluate stresses in the device in different loading situations. The usage of a physiological load with ASTM standard is impossible due to the extreme severity of the ASTM configuration; in this circumstance, the presence of an anterior support is suggested. Also, ISO prescriptions, although the choice of the setup correctly simulates the mechanical contribution of the discs, seem to overstress the device as compared with a physiological loading condition. Some daily activities, other than walking, can induce a further state of stress in the device that should be taken into account in setting up new experimental procedures. Conclusions: ISO standard loading prescriptions seems to be more severe than the expected physiological ones. The ASTM standard should be completed by including some anterior supporting device and declaring the value of the load to be imposed. Moreover, a further enhancement of standards would be simulating other movements representative of daily activities different from walking.

KW - Fatigue test

KW - Finite element method

KW - International standard

KW - Spinal fixator

KW - Validation

UR - http://www.scopus.com/inward/record.url?scp=84897100269&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84897100269&partnerID=8YFLogxK

U2 - 10.1016/j.spinee.2013.08.032

DO - 10.1016/j.spinee.2013.08.032

M3 - Article

C2 - 24268390

AN - SCOPUS:84897100269

VL - 14

SP - 695

EP - 704

JO - Spine Journal

JF - Spine Journal

SN - 1529-9430

IS - 4

ER -