Comparative in vitro study on the long term performance of cemented hip stems: Validation of a protocol to discriminate between "good" and "bad" designs

Luca Cristofolini, Amelia Saponara Teutonico, Luisa Monti, Angelo Cappello, Aldo Toni

Research output: Contribution to journalArticle

75 Citations (Scopus)

Abstract

The long-term clinical success of cemented hip stems is influenced both by the implant design, and by the surgical procedure. A methodology is proposed for discriminating between implant designs with different clinical outcomes. The protocol was designed with industrial pre-clinical validation in mind. Two cemented stem types were tested, one (Lubinus SPII) having good and the other (Müller Curved) having poor clinical outcomes. Three implants for each type were subjected to a mechanical in vitro test of one million loading cycles. Each cycle reproduced the load components of stair climbing. Interface shear micromotion was measured during the test in the direction of rotation and along the stem axis. The stem roughness before and after the test was compared. After the test, the cement mantles were retrieved and inspected through dye penetrants to detect evidences of micro-damage. For each specimen, the events of the loosening process were examined, based on the in vitro data available, so as to analyze the whole failure mechanism. The protocol developed was sensitive to the implant design, with significantly different results being found for the two stem types, both in terms of stem-cement micromotions, surface roughness alteration, and cement mantle damage. The information yielded by the three different investigation techniques was consistent for each of the two groups of specimens tested, allowing a better understanding of the failure process. In vitro inducible micromotion and permanent migration measurements, together with cement-stem interface fretting damage and cement fatigue damage, can help predicting the clinical performance of cemented stems.

Original languageEnglish
Pages (from-to)1603-1615
Number of pages13
JournalJournal of Biomechanics
Volume36
Issue number11
DOIs
Publication statusPublished - Nov 1 2003

Fingerprint

Hip
Cements
Surface roughness
Fatigue
Coloring Agents
Stairs
Fatigue damage
Dyes
In Vitro Techniques
Direction compound
Stair Climbing

Keywords

  • Cement damage
  • Fatigue
  • Hip prostheses
  • In vitro testing
  • Migration

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Comparative in vitro study on the long term performance of cemented hip stems : Validation of a protocol to discriminate between "good" and "bad" designs. / Cristofolini, Luca; Teutonico, Amelia Saponara; Monti, Luisa; Cappello, Angelo; Toni, Aldo.

In: Journal of Biomechanics, Vol. 36, No. 11, 01.11.2003, p. 1603-1615.

Research output: Contribution to journalArticle

Cristofolini, Luca ; Teutonico, Amelia Saponara ; Monti, Luisa ; Cappello, Angelo ; Toni, Aldo. / Comparative in vitro study on the long term performance of cemented hip stems : Validation of a protocol to discriminate between "good" and "bad" designs. In: Journal of Biomechanics. 2003 ; Vol. 36, No. 11. pp. 1603-1615.
@article{08e8528711bb45ddbcd9cf36cd8dc584,
title = "Comparative in vitro study on the long term performance of cemented hip stems: Validation of a protocol to discriminate between {"}good{"} and {"}bad{"} designs",
abstract = "The long-term clinical success of cemented hip stems is influenced both by the implant design, and by the surgical procedure. A methodology is proposed for discriminating between implant designs with different clinical outcomes. The protocol was designed with industrial pre-clinical validation in mind. Two cemented stem types were tested, one (Lubinus SPII) having good and the other (M{\"u}ller Curved) having poor clinical outcomes. Three implants for each type were subjected to a mechanical in vitro test of one million loading cycles. Each cycle reproduced the load components of stair climbing. Interface shear micromotion was measured during the test in the direction of rotation and along the stem axis. The stem roughness before and after the test was compared. After the test, the cement mantles were retrieved and inspected through dye penetrants to detect evidences of micro-damage. For each specimen, the events of the loosening process were examined, based on the in vitro data available, so as to analyze the whole failure mechanism. The protocol developed was sensitive to the implant design, with significantly different results being found for the two stem types, both in terms of stem-cement micromotions, surface roughness alteration, and cement mantle damage. The information yielded by the three different investigation techniques was consistent for each of the two groups of specimens tested, allowing a better understanding of the failure process. In vitro inducible micromotion and permanent migration measurements, together with cement-stem interface fretting damage and cement fatigue damage, can help predicting the clinical performance of cemented stems.",
keywords = "Cement damage, Fatigue, Hip prostheses, In vitro testing, Migration",
author = "Luca Cristofolini and Teutonico, {Amelia Saponara} and Luisa Monti and Angelo Cappello and Aldo Toni",
year = "2003",
month = "11",
day = "1",
doi = "10.1016/S0021-9290(03)00191-X",
language = "English",
volume = "36",
pages = "1603--1615",
journal = "Journal of Biomechanics",
issn = "0021-9290",
publisher = "Elsevier Limited",
number = "11",

}

TY - JOUR

T1 - Comparative in vitro study on the long term performance of cemented hip stems

T2 - Validation of a protocol to discriminate between "good" and "bad" designs

AU - Cristofolini, Luca

AU - Teutonico, Amelia Saponara

AU - Monti, Luisa

AU - Cappello, Angelo

AU - Toni, Aldo

PY - 2003/11/1

Y1 - 2003/11/1

N2 - The long-term clinical success of cemented hip stems is influenced both by the implant design, and by the surgical procedure. A methodology is proposed for discriminating between implant designs with different clinical outcomes. The protocol was designed with industrial pre-clinical validation in mind. Two cemented stem types were tested, one (Lubinus SPII) having good and the other (Müller Curved) having poor clinical outcomes. Three implants for each type were subjected to a mechanical in vitro test of one million loading cycles. Each cycle reproduced the load components of stair climbing. Interface shear micromotion was measured during the test in the direction of rotation and along the stem axis. The stem roughness before and after the test was compared. After the test, the cement mantles were retrieved and inspected through dye penetrants to detect evidences of micro-damage. For each specimen, the events of the loosening process were examined, based on the in vitro data available, so as to analyze the whole failure mechanism. The protocol developed was sensitive to the implant design, with significantly different results being found for the two stem types, both in terms of stem-cement micromotions, surface roughness alteration, and cement mantle damage. The information yielded by the three different investigation techniques was consistent for each of the two groups of specimens tested, allowing a better understanding of the failure process. In vitro inducible micromotion and permanent migration measurements, together with cement-stem interface fretting damage and cement fatigue damage, can help predicting the clinical performance of cemented stems.

AB - The long-term clinical success of cemented hip stems is influenced both by the implant design, and by the surgical procedure. A methodology is proposed for discriminating between implant designs with different clinical outcomes. The protocol was designed with industrial pre-clinical validation in mind. Two cemented stem types were tested, one (Lubinus SPII) having good and the other (Müller Curved) having poor clinical outcomes. Three implants for each type were subjected to a mechanical in vitro test of one million loading cycles. Each cycle reproduced the load components of stair climbing. Interface shear micromotion was measured during the test in the direction of rotation and along the stem axis. The stem roughness before and after the test was compared. After the test, the cement mantles were retrieved and inspected through dye penetrants to detect evidences of micro-damage. For each specimen, the events of the loosening process were examined, based on the in vitro data available, so as to analyze the whole failure mechanism. The protocol developed was sensitive to the implant design, with significantly different results being found for the two stem types, both in terms of stem-cement micromotions, surface roughness alteration, and cement mantle damage. The information yielded by the three different investigation techniques was consistent for each of the two groups of specimens tested, allowing a better understanding of the failure process. In vitro inducible micromotion and permanent migration measurements, together with cement-stem interface fretting damage and cement fatigue damage, can help predicting the clinical performance of cemented stems.

KW - Cement damage

KW - Fatigue

KW - Hip prostheses

KW - In vitro testing

KW - Migration

UR - http://www.scopus.com/inward/record.url?scp=0141792462&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0141792462&partnerID=8YFLogxK

U2 - 10.1016/S0021-9290(03)00191-X

DO - 10.1016/S0021-9290(03)00191-X

M3 - Article

C2 - 14522201

AN - SCOPUS:0141792462

VL - 36

SP - 1603

EP - 1615

JO - Journal of Biomechanics

JF - Journal of Biomechanics

SN - 0021-9290

IS - 11

ER -