Contact stresses and fatigue life in a knee prosthesis: Comparison between in vitro measurements and computational simulations

Tomaso Villa, Francesco Migliavacca, Dario Gastaldi, Maurizio Colombo, Riccardo Pietrabissa

Research output: Contribution to journalArticle

77 Citations (Scopus)

Abstract

The evaluation of contact areas and pressures in total knee prosthesis is a key issue to prevent early failure. The first part of this study is based on the hypothesis that the patterns of contact stresses on the tibial insert of a knee prosthesis at different stages of the gait cycle could be an indicator of the wear performances of a knee prosthesis. Contact stresses were calculated for a mobile bearing knee prosthesis by means of finite element method (FEM). Contact areas and stresses were also measured through in vitro tests using Fuji Prescale film in order to support the FEM findings. The second part of this study addresses the long-term structural integrity of metal tibial components in terms of fatigue life by means of experimental tests and FEM simulations. Fatigue experimental evaluations were performed on Cr-Co alloy tibial tray, based on ISO standards. FEM models were used to calculate the stress patterns. The failure risk was estimated with a standard fatigue criterion on the basis of the results obtained from the FEM calculations. Experimental and computational results showed a positive matching.

Original languageEnglish
Pages (from-to)45-53
Number of pages9
JournalJournal of Biomechanics
Volume37
Issue number1
DOIs
Publication statusPublished - Jan 2004

Fingerprint

Knee prostheses
Knee Prosthesis
Psychological Stress
Fatigue
Fatigue of materials
Finite element method
Bearings (structural)
Structural integrity
Gait
Metals
Wear of materials
In Vitro Techniques
Pressure

Keywords

  • Finite element method
  • Fuji film
  • ISO standard
  • Rotating platform knee prosthesis
  • Tibial tray

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Contact stresses and fatigue life in a knee prosthesis : Comparison between in vitro measurements and computational simulations. / Villa, Tomaso; Migliavacca, Francesco; Gastaldi, Dario; Colombo, Maurizio; Pietrabissa, Riccardo.

In: Journal of Biomechanics, Vol. 37, No. 1, 01.2004, p. 45-53.

Research output: Contribution to journalArticle

Villa, Tomaso ; Migliavacca, Francesco ; Gastaldi, Dario ; Colombo, Maurizio ; Pietrabissa, Riccardo. / Contact stresses and fatigue life in a knee prosthesis : Comparison between in vitro measurements and computational simulations. In: Journal of Biomechanics. 2004 ; Vol. 37, No. 1. pp. 45-53.
@article{035626a71f07458db43d309694240b1b,
title = "Contact stresses and fatigue life in a knee prosthesis: Comparison between in vitro measurements and computational simulations",
abstract = "The evaluation of contact areas and pressures in total knee prosthesis is a key issue to prevent early failure. The first part of this study is based on the hypothesis that the patterns of contact stresses on the tibial insert of a knee prosthesis at different stages of the gait cycle could be an indicator of the wear performances of a knee prosthesis. Contact stresses were calculated for a mobile bearing knee prosthesis by means of finite element method (FEM). Contact areas and stresses were also measured through in vitro tests using Fuji Prescale film in order to support the FEM findings. The second part of this study addresses the long-term structural integrity of metal tibial components in terms of fatigue life by means of experimental tests and FEM simulations. Fatigue experimental evaluations were performed on Cr-Co alloy tibial tray, based on ISO standards. FEM models were used to calculate the stress patterns. The failure risk was estimated with a standard fatigue criterion on the basis of the results obtained from the FEM calculations. Experimental and computational results showed a positive matching.",
keywords = "Finite element method, Fuji film, ISO standard, Rotating platform knee prosthesis, Tibial tray",
author = "Tomaso Villa and Francesco Migliavacca and Dario Gastaldi and Maurizio Colombo and Riccardo Pietrabissa",
year = "2004",
month = "1",
doi = "10.1016/S0021-9290(03)00255-0",
language = "English",
volume = "37",
pages = "45--53",
journal = "Journal of Biomechanics",
issn = "0021-9290",
publisher = "Elsevier Limited",
number = "1",

}

TY - JOUR

T1 - Contact stresses and fatigue life in a knee prosthesis

T2 - Comparison between in vitro measurements and computational simulations

AU - Villa, Tomaso

AU - Migliavacca, Francesco

AU - Gastaldi, Dario

AU - Colombo, Maurizio

AU - Pietrabissa, Riccardo

PY - 2004/1

Y1 - 2004/1

N2 - The evaluation of contact areas and pressures in total knee prosthesis is a key issue to prevent early failure. The first part of this study is based on the hypothesis that the patterns of contact stresses on the tibial insert of a knee prosthesis at different stages of the gait cycle could be an indicator of the wear performances of a knee prosthesis. Contact stresses were calculated for a mobile bearing knee prosthesis by means of finite element method (FEM). Contact areas and stresses were also measured through in vitro tests using Fuji Prescale film in order to support the FEM findings. The second part of this study addresses the long-term structural integrity of metal tibial components in terms of fatigue life by means of experimental tests and FEM simulations. Fatigue experimental evaluations were performed on Cr-Co alloy tibial tray, based on ISO standards. FEM models were used to calculate the stress patterns. The failure risk was estimated with a standard fatigue criterion on the basis of the results obtained from the FEM calculations. Experimental and computational results showed a positive matching.

AB - The evaluation of contact areas and pressures in total knee prosthesis is a key issue to prevent early failure. The first part of this study is based on the hypothesis that the patterns of contact stresses on the tibial insert of a knee prosthesis at different stages of the gait cycle could be an indicator of the wear performances of a knee prosthesis. Contact stresses were calculated for a mobile bearing knee prosthesis by means of finite element method (FEM). Contact areas and stresses were also measured through in vitro tests using Fuji Prescale film in order to support the FEM findings. The second part of this study addresses the long-term structural integrity of metal tibial components in terms of fatigue life by means of experimental tests and FEM simulations. Fatigue experimental evaluations were performed on Cr-Co alloy tibial tray, based on ISO standards. FEM models were used to calculate the stress patterns. The failure risk was estimated with a standard fatigue criterion on the basis of the results obtained from the FEM calculations. Experimental and computational results showed a positive matching.

KW - Finite element method

KW - Fuji film

KW - ISO standard

KW - Rotating platform knee prosthesis

KW - Tibial tray

UR - http://www.scopus.com/inward/record.url?scp=0345735675&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0345735675&partnerID=8YFLogxK

U2 - 10.1016/S0021-9290(03)00255-0

DO - 10.1016/S0021-9290(03)00255-0

M3 - Article

C2 - 14672567

AN - SCOPUS:0345735675

VL - 37

SP - 45

EP - 53

JO - Journal of Biomechanics

JF - Journal of Biomechanics

SN - 0021-9290

IS - 1

ER -