Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM)

Marco Bertolini, Andrea Nitrosi, Giovanni Borasi, Andrea Botti, Davide Tassoni, Roberto Sghedoni, Giulio Zuccoli

Research output: Contribution to journalArticle

Abstract

The performance of a commercial digital mammographic system working in 2D planar versus tomosynthesis mode was evaluated in terms of the image signal difference to noise ratio (SDNR). A contrast detail phantom was obtained embedding 1 cm Plexiglas, including 49 holes of different diameter and depth, between two layers containing a breast-simulating material. The phantom was exposed with the details plane perpendicular to the X-ray beam using the manufacturer's standard clinical breast acquisition parameters. SDNR in the digital breast tomosynthesis (DBT) images was higher than that of the full-field digital mammography (FFDM) for 38 out of 49 details in complex background conditions. These differences (p

Original languageEnglish
Pages (from-to)58-65
Number of pages8
JournalJournal of Digital Imaging
Volume24
Issue number1
DOIs
Publication statusPublished - Feb 2011

Keywords

  • contrast detail phantom analysis
  • digital breast tomosynthesis
  • Digital mammography
  • full-field digital mammography (FFDM)
  • image quality analysis
  • imaging
  • phantoms

ASJC Scopus subject areas

  • Computer Science Applications
  • Medicine(all)
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Fingerprint Dive into the research topics of 'Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM)'. Together they form a unique fingerprint.