Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: Fundamental characterization in water

Antonella Fogliata, Giorgia Nicolini, Alessandro Clivio, Eugenio Vanetti, Pietro Mancosu, Luca Cozzi

Research output: Contribution to journalArticle

59 Citations (Scopus)

Abstract

A new algorithm, Acuros XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these pre-clinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.

Original languageEnglish
Pages (from-to)1879-1904
Number of pages26
JournalPhysics in Medicine and Biology
Volume56
Issue number6
DOIs
Publication statusPublished - Mar 21 2011

Fingerprint

Water
Particle Accelerators
Photons
General Practice
Radiotherapy
Radiation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm : Fundamental characterization in water. / Fogliata, Antonella; Nicolini, Giorgia; Clivio, Alessandro; Vanetti, Eugenio; Mancosu, Pietro; Cozzi, Luca.

In: Physics in Medicine and Biology, Vol. 56, No. 6, 21.03.2011, p. 1879-1904.

Research output: Contribution to journalArticle

Fogliata, Antonella ; Nicolini, Giorgia ; Clivio, Alessandro ; Vanetti, Eugenio ; Mancosu, Pietro ; Cozzi, Luca. / Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm : Fundamental characterization in water. In: Physics in Medicine and Biology. 2011 ; Vol. 56, No. 6. pp. 1879-1904.
@article{5cd8fb72795940da8a74ff3bfb38176d,
title = "Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: Fundamental characterization in water",
abstract = "A new algorithm, Acuros XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these pre-clinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1{\%} for open beams and within 2{\%} for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.",
author = "Antonella Fogliata and Giorgia Nicolini and Alessandro Clivio and Eugenio Vanetti and Pietro Mancosu and Luca Cozzi",
year = "2011",
month = "3",
day = "21",
doi = "10.1088/0031-9155/56/6/022",
language = "English",
volume = "56",
pages = "1879--1904",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "6",

}

TY - JOUR

T1 - Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm

T2 - Fundamental characterization in water

AU - Fogliata, Antonella

AU - Nicolini, Giorgia

AU - Clivio, Alessandro

AU - Vanetti, Eugenio

AU - Mancosu, Pietro

AU - Cozzi, Luca

PY - 2011/3/21

Y1 - 2011/3/21

N2 - A new algorithm, Acuros XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these pre-clinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.

AB - A new algorithm, Acuros XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these pre-clinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.

UR - http://www.scopus.com/inward/record.url?scp=79952919293&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79952919293&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/56/6/022

DO - 10.1088/0031-9155/56/6/022

M3 - Article

C2 - 21364257

AN - SCOPUS:79952919293

VL - 56

SP - 1879

EP - 1904

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 6

ER -