DualTrack technology improves optical coherence tomography angiography image quality

Luigi Antonio De Vitis, R Sacconi, Adriano Carnevali, Teresa Centoducati, M. Cavalleri, L Querques, F Bandello, G Querques

Research output: Contribution to journalArticle

Abstract

BACKGROUND AND OBJECTIVE: To evaluate the effects of DualTrack Motion Correction Technology on the performance of AngioVue optical coherence tomography angiography (Optovue, Freemont, CA). PATIENTS AND METHODS: Fifty subjects underwent two consecutive examinations with AngioVue. The operator inactivated the eye-tracking (ET) technology before the second examination. Two operators measured execution time, excluded low-quality images, and counted the number of motion artifacts per image. RESULTS: The mean execution time was lower without ET in all the groups (P < .05), regardless of retinal diseases. Percentages of available images and low-quality images were 75.0% and 25.0%, respectively, with ET and 63.0% and 37.0%, respectively, without ET. Mean number of motion artifacts was significantly lower in images obtained using ET compared to images obtained without ET in all the groups (P < .0001), regardless of retinal diseases. CONCLUSION: Although ET technology increases the mean execution time, the reduction of motion artifacts and the increase of high-quality images may represent a great advantage in clinical practice. © 2017, Slack Incorporated. All rights reserved.
Original languageEnglish
Pages (from-to)918-926
Number of pages9
JournalOphthalmic Surgery Lasers and Imaging Retina
Volume48
Issue number11
DOIs
Publication statusPublished - 2017