Early rupture of subclavian vein catheter: A case report and literature review

Giacomo Sarzo, C. Finco, F. Zustovich, P. Parise, S. Savastano, S. Degregori, M. Vecchiato, S. Merigliano

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Prolonged venous access devices (PVADs) have become indispensable in antiblastic protocols for the treatment of cancer patients, in anti-infection protocols for acquired immunodeficiency syndrome (AIDS) patients and in the management of chronic malabsorption syndromes. Using these catheters carries the risk of several complications, and some are potentially lethal, for example, cardiac embolization of catheter fragments. Rupture is a complication almost exclusive to catheters positioned percutaneously: after using this technique, device malfunction can occur due to catheter kinking after its excessively medial introduction in the subclavian vein. The early recognition of any pinch-off sign (POS) is fundamental in preventing catheter rupture that frequently follows this complication. Other factors can be involved in early rupture, for example, excessive force on a syringe used to clear a catheter that shows early signs of malfunction, or a strength defect in the materials used in the catheter construction. This report describes an early rupture case of an initially correctly positioned catheter and reviews 20 such cases in the recent literature.

Original languageEnglish
Pages (from-to)39-46
Number of pages8
JournalJournal of Vascular Access
Volume5
Issue number1
Publication statusPublished - Jan 2004

Fingerprint

Subclavian Vein
Rupture
Catheters
Antineoplastic Protocols
Cardiac Catheters
Malabsorption Syndromes
Equipment and Supplies
Syringes
Acquired Immunodeficiency Syndrome
Infection

Keywords

  • Catheters
  • Chemotherapy
  • Post-operative complications
  • Prolonged venous access devices
  • Venous access

ASJC Scopus subject areas

  • Surgery

Cite this

Sarzo, G., Finco, C., Zustovich, F., Parise, P., Savastano, S., Degregori, S., ... Merigliano, S. (2004). Early rupture of subclavian vein catheter: A case report and literature review. Journal of Vascular Access, 5(1), 39-46.

Early rupture of subclavian vein catheter : A case report and literature review. / Sarzo, Giacomo; Finco, C.; Zustovich, F.; Parise, P.; Savastano, S.; Degregori, S.; Vecchiato, M.; Merigliano, S.

In: Journal of Vascular Access, Vol. 5, No. 1, 01.2004, p. 39-46.

Research output: Contribution to journalArticle

Sarzo, G, Finco, C, Zustovich, F, Parise, P, Savastano, S, Degregori, S, Vecchiato, M & Merigliano, S 2004, 'Early rupture of subclavian vein catheter: A case report and literature review', Journal of Vascular Access, vol. 5, no. 1, pp. 39-46.
Sarzo G, Finco C, Zustovich F, Parise P, Savastano S, Degregori S et al. Early rupture of subclavian vein catheter: A case report and literature review. Journal of Vascular Access. 2004 Jan;5(1):39-46.
Sarzo, Giacomo ; Finco, C. ; Zustovich, F. ; Parise, P. ; Savastano, S. ; Degregori, S. ; Vecchiato, M. ; Merigliano, S. / Early rupture of subclavian vein catheter : A case report and literature review. In: Journal of Vascular Access. 2004 ; Vol. 5, No. 1. pp. 39-46.
@article{aa39a4ff335342ffaa337ba5ec0bf3b3,
title = "Early rupture of subclavian vein catheter: A case report and literature review",
abstract = "Prolonged venous access devices (PVADs) have become indispensable in antiblastic protocols for the treatment of cancer patients, in anti-infection protocols for acquired immunodeficiency syndrome (AIDS) patients and in the management of chronic malabsorption syndromes. Using these catheters carries the risk of several complications, and some are potentially lethal, for example, cardiac embolization of catheter fragments. Rupture is a complication almost exclusive to catheters positioned percutaneously: after using this technique, device malfunction can occur due to catheter kinking after its excessively medial introduction in the subclavian vein. The early recognition of any pinch-off sign (POS) is fundamental in preventing catheter rupture that frequently follows this complication. Other factors can be involved in early rupture, for example, excessive force on a syringe used to clear a catheter that shows early signs of malfunction, or a strength defect in the materials used in the catheter construction. This report describes an early rupture case of an initially correctly positioned catheter and reviews 20 such cases in the recent literature.",
keywords = "Catheters, Chemotherapy, Post-operative complications, Prolonged venous access devices, Venous access",
author = "Giacomo Sarzo and C. Finco and F. Zustovich and P. Parise and S. Savastano and S. Degregori and M. Vecchiato and S. Merigliano",
year = "2004",
month = "1",
language = "English",
volume = "5",
pages = "39--46",
journal = "Journal of Vascular Access",
issn = "1129-7298",
publisher = "Wichtig Publishing",
number = "1",

}

TY - JOUR

T1 - Early rupture of subclavian vein catheter

T2 - A case report and literature review

AU - Sarzo, Giacomo

AU - Finco, C.

AU - Zustovich, F.

AU - Parise, P.

AU - Savastano, S.

AU - Degregori, S.

AU - Vecchiato, M.

AU - Merigliano, S.

PY - 2004/1

Y1 - 2004/1

N2 - Prolonged venous access devices (PVADs) have become indispensable in antiblastic protocols for the treatment of cancer patients, in anti-infection protocols for acquired immunodeficiency syndrome (AIDS) patients and in the management of chronic malabsorption syndromes. Using these catheters carries the risk of several complications, and some are potentially lethal, for example, cardiac embolization of catheter fragments. Rupture is a complication almost exclusive to catheters positioned percutaneously: after using this technique, device malfunction can occur due to catheter kinking after its excessively medial introduction in the subclavian vein. The early recognition of any pinch-off sign (POS) is fundamental in preventing catheter rupture that frequently follows this complication. Other factors can be involved in early rupture, for example, excessive force on a syringe used to clear a catheter that shows early signs of malfunction, or a strength defect in the materials used in the catheter construction. This report describes an early rupture case of an initially correctly positioned catheter and reviews 20 such cases in the recent literature.

AB - Prolonged venous access devices (PVADs) have become indispensable in antiblastic protocols for the treatment of cancer patients, in anti-infection protocols for acquired immunodeficiency syndrome (AIDS) patients and in the management of chronic malabsorption syndromes. Using these catheters carries the risk of several complications, and some are potentially lethal, for example, cardiac embolization of catheter fragments. Rupture is a complication almost exclusive to catheters positioned percutaneously: after using this technique, device malfunction can occur due to catheter kinking after its excessively medial introduction in the subclavian vein. The early recognition of any pinch-off sign (POS) is fundamental in preventing catheter rupture that frequently follows this complication. Other factors can be involved in early rupture, for example, excessive force on a syringe used to clear a catheter that shows early signs of malfunction, or a strength defect in the materials used in the catheter construction. This report describes an early rupture case of an initially correctly positioned catheter and reviews 20 such cases in the recent literature.

KW - Catheters

KW - Chemotherapy

KW - Post-operative complications

KW - Prolonged venous access devices

KW - Venous access

UR - http://www.scopus.com/inward/record.url?scp=2542444392&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=2542444392&partnerID=8YFLogxK

M3 - Article

C2 - 16596539

AN - SCOPUS:2542444392

VL - 5

SP - 39

EP - 46

JO - Journal of Vascular Access

JF - Journal of Vascular Access

SN - 1129-7298

IS - 1

ER -