Efficiency of different hollow-fiber hemofilters in continuous arteriovenous hemodiafiltration

Santo Morabito, Alessandro Pierucci, Rocco Marinelli, Angela Cicciarelli, Isabella Guzzo, Giulio Alberto Cinotti, Benedetto Marino, Roberto Chiavarelli

Research output: Contribution to journalArticle

Abstract

Low dialysate to blood flow rate ratios are a unique characteristic of continuous arteriovenous hemodiafiltration (CAVHDF) that should allow complete saturation of dialysis fluid with small-molecular-weight blood solutes. The aim of the investigation was to evaluate the performance of different hemofilters in CAVHDF. In 10 critically ill patients with acute renal failure, the efficiency of four hollow-fiber hemofilters, polyamide 0.6 m2, polyacrylonitrile (PAN) 0.3 and 0.6 m2, acrylonitrile sodium methallylsulfonate (AN69HF) 0.6 m2, has been evaluated. For comparison, dialysate flow rates (Q(di)) were standardized to 16.6 and 25 ml/min. Samples for urea nitrogen were obtained from the arterial blood line (C(bi)) and from the dialysate exit port (C(do)) within 24-hour running time. Outflowing dialysate (Q(do)) was also measured at the same time. Blood flow (Q(b)) was calculated by the bubble transit time technique. Diffusive and total urea clearances were determined. AN69HF and PAN hemofilters provided higher clearances than the polyamide hemofilter. Despite the smaller surface area, PAN 0.3 m2 had a total urea clearance comparable to that of PAN 0.6 m2 and AN69HF at Q(di) = 16.6 ml/min. While at Q(di) = 16.6 ml/min equilibrium between blood and dialysate (C(do)/C(bi) congruent with 1) occurred with the AN69HF and PAN hemofilters, at Q(di) = 25 ml/min the equilibrium was obtained only with the AN69HF hemofilter. In conclusion, almost complete urea saturation of dialysis fluid has not been obtained with all hemofilters tested here. In our experience, membrane characteristics play an important role in determining diffusive efficiency in CAVHDF. Copyright (C) 2000 S. Karger AG, Basel.

Original languageEnglish
Pages (from-to)116-121
Number of pages6
JournalAmerican Journal of Nephrology
Volume20
Issue number2
Publication statusPublished - Mar 2000

Keywords

  • Acute renal failure
  • Continuous arteriovenous hemodiafiltration hollow-fiber hemofilters
  • Continuous renal replacement therapy
  • Polyacrylonitrile acrylonitrile sodium methallylsulfonate
  • Polyamide
  • Urea clearance

ASJC Scopus subject areas

  • Nephrology

Fingerprint Dive into the research topics of 'Efficiency of different hollow-fiber hemofilters in continuous arteriovenous hemodiafiltration'. Together they form a unique fingerprint.

  • Cite this

    Morabito, S., Pierucci, A., Marinelli, R., Cicciarelli, A., Guzzo, I., Cinotti, G. A., Marino, B., & Chiavarelli, R. (2000). Efficiency of different hollow-fiber hemofilters in continuous arteriovenous hemodiafiltration. American Journal of Nephrology, 20(2), 116-121.