Evaluating the effect of pupil dilation on spectral-domain optical coherence tomography measurements and their quality score

Lucia Tanga, Gloria Roberti, Francesco Oddone, Luciano Quaranta, Manuela Ferrazza, Francesca Berardo, Gianluca Manni, Marco Centofanti

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

BACKGROUND: Spectral-domain optical coherence tomography (SD-OCT) provides fast scan speed and high scan resolution improving its diagnostic accuracy. The purpose of this study was to evaluate if SD-OCT measurements and their quality score are influenced by pupil dilation.

METHODS: Retinal nerve fiber layer thickness (RNFL), ganglion cell complex (GCC) and optic nerve head (ONH) were measured in one eye of 57 glaucoma patients and 36 healthy subjects using spectral domain optical coherence tomography (SD-OCT) before and after pupil dilation. Comparisons were made between measurements and their quality score pre- and post dilation (Signal Strength Index, SSI). Overall RNFL, average GCC and ONH rim volume were considered in the analysis.

RESULTS: No statistically significant differences were found between pre- and post-dilation measurements in both groups (glaucoma: RNFL 80 ± 15 μm vs 80 ± 16 μm, p = 0.87; GCC 81.35 ± 13.4 μm vs 81.10 ± 13.14 μm, p = 0.92; ONH 0.05 ± 0.11 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.74; controls RNFL 99 ± 12 μm vs 98 ± 14 μm, p = 0.70; GCC 92.12 ± 6.7 μm vs 91.54 ± 7.05 μm, p = 0.72; ONH 0.11 ± 0.1 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.36) nor between pre- and post-dilation quality score (glaucoma SSI RNFL 54.3 ± 10.3 vs 51.7 ± 18.1, p = 0.12; SSI GCC 58 ± 9.5 vs 57 ± 8.09, p = 0.55; SSI ONH 48.5 ± 7.6 vs 46.6 ± 7.2, p = 0.16; controls SSI RNFL 57 ± 10.3 vs 54 ± 9.31, p = 0.2; SSI GCC 60.9 ± 8.1 vs 58.8 ± 7.3, p = 0.3; SSI ONH 51.5 ± 8.9 vs 50.4 ± 8.3, p = 0.59).

CONCLUSION: Pupil dilation doesn't affect SD-OCT measurements and their quality score.

Original languageEnglish
Pages (from-to)175
Number of pages1
JournalBMC Ophthalmology
Volume15
DOIs
Publication statusPublished - 2015

Fingerprint

Optical Coherence Tomography
Pupil
Dilatation
Optic Disk
Nerve Fibers
Ganglia
Glaucoma
Healthy Volunteers

ASJC Scopus subject areas

  • Medicine(all)

Cite this

Evaluating the effect of pupil dilation on spectral-domain optical coherence tomography measurements and their quality score. / Tanga, Lucia; Roberti, Gloria; Oddone, Francesco; Quaranta, Luciano; Ferrazza, Manuela; Berardo, Francesca; Manni, Gianluca; Centofanti, Marco.

In: BMC Ophthalmology, Vol. 15, 2015, p. 175.

Research output: Contribution to journalArticle

@article{40f223df509c4654bca55076838c58ed,
title = "Evaluating the effect of pupil dilation on spectral-domain optical coherence tomography measurements and their quality score",
abstract = "BACKGROUND: Spectral-domain optical coherence tomography (SD-OCT) provides fast scan speed and high scan resolution improving its diagnostic accuracy. The purpose of this study was to evaluate if SD-OCT measurements and their quality score are influenced by pupil dilation.METHODS: Retinal nerve fiber layer thickness (RNFL), ganglion cell complex (GCC) and optic nerve head (ONH) were measured in one eye of 57 glaucoma patients and 36 healthy subjects using spectral domain optical coherence tomography (SD-OCT) before and after pupil dilation. Comparisons were made between measurements and their quality score pre- and post dilation (Signal Strength Index, SSI). Overall RNFL, average GCC and ONH rim volume were considered in the analysis.RESULTS: No statistically significant differences were found between pre- and post-dilation measurements in both groups (glaucoma: RNFL 80 ± 15 μm vs 80 ± 16 μm, p = 0.87; GCC 81.35 ± 13.4 μm vs 81.10 ± 13.14 μm, p = 0.92; ONH 0.05 ± 0.11 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.74; controls RNFL 99 ± 12 μm vs 98 ± 14 μm, p = 0.70; GCC 92.12 ± 6.7 μm vs 91.54 ± 7.05 μm, p = 0.72; ONH 0.11 ± 0.1 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.36) nor between pre- and post-dilation quality score (glaucoma SSI RNFL 54.3 ± 10.3 vs 51.7 ± 18.1, p = 0.12; SSI GCC 58 ± 9.5 vs 57 ± 8.09, p = 0.55; SSI ONH 48.5 ± 7.6 vs 46.6 ± 7.2, p = 0.16; controls SSI RNFL 57 ± 10.3 vs 54 ± 9.31, p = 0.2; SSI GCC 60.9 ± 8.1 vs 58.8 ± 7.3, p = 0.3; SSI ONH 51.5 ± 8.9 vs 50.4 ± 8.3, p = 0.59).CONCLUSION: Pupil dilation doesn't affect SD-OCT measurements and their quality score.",
author = "Lucia Tanga and Gloria Roberti and Francesco Oddone and Luciano Quaranta and Manuela Ferrazza and Francesca Berardo and Gianluca Manni and Marco Centofanti",
year = "2015",
doi = "10.1186/s12886-015-0168-y",
language = "English",
volume = "15",
pages = "175",
journal = "BMC Ophthalmology",
issn = "1471-2415",
publisher = "BioMed Central",

}

TY - JOUR

T1 - Evaluating the effect of pupil dilation on spectral-domain optical coherence tomography measurements and their quality score

AU - Tanga, Lucia

AU - Roberti, Gloria

AU - Oddone, Francesco

AU - Quaranta, Luciano

AU - Ferrazza, Manuela

AU - Berardo, Francesca

AU - Manni, Gianluca

AU - Centofanti, Marco

PY - 2015

Y1 - 2015

N2 - BACKGROUND: Spectral-domain optical coherence tomography (SD-OCT) provides fast scan speed and high scan resolution improving its diagnostic accuracy. The purpose of this study was to evaluate if SD-OCT measurements and their quality score are influenced by pupil dilation.METHODS: Retinal nerve fiber layer thickness (RNFL), ganglion cell complex (GCC) and optic nerve head (ONH) were measured in one eye of 57 glaucoma patients and 36 healthy subjects using spectral domain optical coherence tomography (SD-OCT) before and after pupil dilation. Comparisons were made between measurements and their quality score pre- and post dilation (Signal Strength Index, SSI). Overall RNFL, average GCC and ONH rim volume were considered in the analysis.RESULTS: No statistically significant differences were found between pre- and post-dilation measurements in both groups (glaucoma: RNFL 80 ± 15 μm vs 80 ± 16 μm, p = 0.87; GCC 81.35 ± 13.4 μm vs 81.10 ± 13.14 μm, p = 0.92; ONH 0.05 ± 0.11 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.74; controls RNFL 99 ± 12 μm vs 98 ± 14 μm, p = 0.70; GCC 92.12 ± 6.7 μm vs 91.54 ± 7.05 μm, p = 0.72; ONH 0.11 ± 0.1 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.36) nor between pre- and post-dilation quality score (glaucoma SSI RNFL 54.3 ± 10.3 vs 51.7 ± 18.1, p = 0.12; SSI GCC 58 ± 9.5 vs 57 ± 8.09, p = 0.55; SSI ONH 48.5 ± 7.6 vs 46.6 ± 7.2, p = 0.16; controls SSI RNFL 57 ± 10.3 vs 54 ± 9.31, p = 0.2; SSI GCC 60.9 ± 8.1 vs 58.8 ± 7.3, p = 0.3; SSI ONH 51.5 ± 8.9 vs 50.4 ± 8.3, p = 0.59).CONCLUSION: Pupil dilation doesn't affect SD-OCT measurements and their quality score.

AB - BACKGROUND: Spectral-domain optical coherence tomography (SD-OCT) provides fast scan speed and high scan resolution improving its diagnostic accuracy. The purpose of this study was to evaluate if SD-OCT measurements and their quality score are influenced by pupil dilation.METHODS: Retinal nerve fiber layer thickness (RNFL), ganglion cell complex (GCC) and optic nerve head (ONH) were measured in one eye of 57 glaucoma patients and 36 healthy subjects using spectral domain optical coherence tomography (SD-OCT) before and after pupil dilation. Comparisons were made between measurements and their quality score pre- and post dilation (Signal Strength Index, SSI). Overall RNFL, average GCC and ONH rim volume were considered in the analysis.RESULTS: No statistically significant differences were found between pre- and post-dilation measurements in both groups (glaucoma: RNFL 80 ± 15 μm vs 80 ± 16 μm, p = 0.87; GCC 81.35 ± 13.4 μm vs 81.10 ± 13.14 μm, p = 0.92; ONH 0.05 ± 0.11 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.74; controls RNFL 99 ± 12 μm vs 98 ± 14 μm, p = 0.70; GCC 92.12 ± 6.7 μm vs 91.54 ± 7.05 μm, p = 0.72; ONH 0.11 ± 0.1 mm(3) vs 0.04 ± 0.07 mm(3), p = 0.36) nor between pre- and post-dilation quality score (glaucoma SSI RNFL 54.3 ± 10.3 vs 51.7 ± 18.1, p = 0.12; SSI GCC 58 ± 9.5 vs 57 ± 8.09, p = 0.55; SSI ONH 48.5 ± 7.6 vs 46.6 ± 7.2, p = 0.16; controls SSI RNFL 57 ± 10.3 vs 54 ± 9.31, p = 0.2; SSI GCC 60.9 ± 8.1 vs 58.8 ± 7.3, p = 0.3; SSI ONH 51.5 ± 8.9 vs 50.4 ± 8.3, p = 0.59).CONCLUSION: Pupil dilation doesn't affect SD-OCT measurements and their quality score.

UR - http://www.scopus.com/inward/record.url?scp=84973427480&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84973427480&partnerID=8YFLogxK

U2 - 10.1186/s12886-015-0168-y

DO - 10.1186/s12886-015-0168-y

M3 - Article

C2 - 26654127

AN - SCOPUS:84973427480

VL - 15

SP - 175

JO - BMC Ophthalmology

JF - BMC Ophthalmology

SN - 1471-2415

ER -