Evaluation and management of malfunctioning artificial urinary sphincters

M. Schettini, M. Diana, M. Gallucci

Research output: Contribution to journalArticlepeer-review

Abstract

We report our experience of mechanical malfunctions after AMS 800 artificial urinary sphincter placement. From 1989 to 1997, 42 patients (39 men, 3 women with a mean age of 69 years) underwent artificial sphincter implantation. 5 required explantation of the prosthesis because of infection. The following mechanical malfunctions occurred: 2 cases of air bubbles, 3 cases of pump overturning with tube kinking, 16 cases of persistent or recurrent incontinence 6 to 61 months postoperatively, 1 case of cuff opening during sexual intercourse. Several complications may arise after artificial sphincter placement (infection, urethral erosion, mechanical malfunction). In the last instances, we emphasize the importance of an accurate diagnosis for the correct choice of treatment.

Original languageEnglish
Pages (from-to)301-304
Number of pages4
JournalActa Urologica Italica
Volume12
Issue number6
Publication statusPublished - 1998

Keywords

  • Artificial urinary sphincter
  • Malfunctioning
  • Urinary incontinence

ASJC Scopus subject areas

  • Urology

Fingerprint Dive into the research topics of 'Evaluation and management of malfunctioning artificial urinary sphincters'. Together they form a unique fingerprint.

Cite this