Experimental determination of the radionuclide calibrator setting for Technesium-99m, by using a primary standardisation method

Vittorio Cannatà, Gianclaudio Ciofetta, Maria Carmen Garganese, Pierino De Felice, Marco Capogni, Aldo Fazio, Francesca Maria Marracino

Research output: Contribution to journalArticle

Abstract

BACKGROUND: A new Tc national standard has been developed at the Italian National Metrology Institute for Ionizing Radiation quantities of ENEA (ENEA-INMRI). METHODS: The stock radioactive solution was prepared at the Ospedale Pediatrico Bambino Gesù (OPBG) and standardized at the ENEA-INMRI by using the 4πγ absolute measurement method. The new standard was used to calibrate the dose calibrator of the OPBG Nuclear Medicine Unit (NMU) and the ENEA-INMRI portable ionization chamber which was employed as the secondary standard measurement system to calibrate the dose calibrator of another Italian nuclear medicine unit. RESULTS AND CONCLUSION: Remarkable improvement of measurement accuracy was achieved. The importance of traceable calibrations is underlined. Specific comments are made about quality control performance and measurement accuracy for the instrumentation used in nuclear medicine units of paediatric hospitals.

Original languageEnglish
Pages (from-to)321-326
Number of pages6
JournalNuclear Medicine Communications
Volume28
Issue number4
DOIs
Publication statusPublished - Apr 2007

    Fingerprint

Keywords

  • Tc radiopharmaceutical
  • Administered activity
  • Calibration factor
  • Metrological supervision
  • Paediatric DRL
  • Radionuclide calibrator

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this