Failure of implantable cardioverter-defibrillator leads: a matter of lead size?

Roberto Rordorf, Luca Poggio, Simone Savastano, Alessandro Vicentini, Barbara Petracci, Enrico Chieffo, Catherine Klersy, Maurizio Landolina

Research output: Contribution to journalArticlepeer-review

Abstract

Small-diameter implantable cardioverter-defibrillator (ICD) Sprint Fidelis and Riata leads have been recalled owing to an increased risk of lead failure, thus arousing the suspicion that lead size might be a critical issue. To compare the incidence of failure of small-diameter (≤8 F) and standard-diameter (>8 F) ICD leads implanted in a single center. From January 2003 to December 2010, 190 Sprint Fidelis, 182 Riata/Riata ST, 99 Optim (Riata Optim/Durata), and 419 standard-diameter leads were implanted in our center. During a median follow-up of 33 months, the overall failure rate was 6.3%. Follow-up duration was similar for Sprint Fidelis, Riata, and standard-diameter leads but shorter for the Optim group. The failure rate was significantly higher in Sprint Fidelis leads than in both standard-diameter (4.8%/year vs 0.8%/year; P

Original languageEnglish
Pages (from-to)184-190
Number of pages7
JournalHeart Rhythm
Volume10
Issue number2
Publication statusPublished - Feb 2013

ASJC Scopus subject areas

  • Medicine(all)

Fingerprint Dive into the research topics of 'Failure of implantable cardioverter-defibrillator leads: a matter of lead size?'. Together they form a unique fingerprint.

Cite this