Finite element study of human lumbar disc nucleus replacements

Hendrik Schmidt, Maxim Bashkuev, Fabio Galbusera, Hans Joachim Wilke, Aboulfazl Shirazi-Adl

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

Currently, there are a number of nucleus replacements under development. The important concern is how well these implants duplicate the mechanical function of the native nucleus. This finite element model study aimed to investigate the influence of different nucleus replacements on the mechanical response of the disc. Models included partial, full, over-sized, partially saturated, elastic and poroelastic solid replacements. Over-sized nucleus replacements up to 25% yielded results that were comparable to those in the intact state. Differences were much greater in cases with under-sized nucleus replacements. The effect was most pronounced for the 75% under-sized replacement that resembled the condition with a full nucleotomy. Nucleus implants with elastic properties substantially altered load transmission when 10% under-sized and over-sized replacements were considered. Compared to intact, the under-sized implants should be avoided when using biphasic materials with properties similar to the native nucleus, whereas for elastic replacements both under- and over-sized implants should not be used.

Original languageEnglish
Pages (from-to)1762-1776
Number of pages15
JournalComputer Methods in Biomechanics and Biomedical Engineering
Volume17
Issue number16
DOIs
Publication statusPublished - 2014

Keywords

  • finite element analysis
  • implants
  • intervertebral disc
  • nucleus replacement
  • poroelastic

ASJC Scopus subject areas

  • Bioengineering
  • Biomedical Engineering
  • Computer Science Applications
  • Human-Computer Interaction
  • Medicine(all)

Cite this

Finite element study of human lumbar disc nucleus replacements. / Schmidt, Hendrik; Bashkuev, Maxim; Galbusera, Fabio; Wilke, Hans Joachim; Shirazi-Adl, Aboulfazl.

In: Computer Methods in Biomechanics and Biomedical Engineering, Vol. 17, No. 16, 2014, p. 1762-1776.

Research output: Contribution to journalArticle

Schmidt, Hendrik ; Bashkuev, Maxim ; Galbusera, Fabio ; Wilke, Hans Joachim ; Shirazi-Adl, Aboulfazl. / Finite element study of human lumbar disc nucleus replacements. In: Computer Methods in Biomechanics and Biomedical Engineering. 2014 ; Vol. 17, No. 16. pp. 1762-1776.
@article{8d07168568314e7e981544171f81067e,
title = "Finite element study of human lumbar disc nucleus replacements",
abstract = "Currently, there are a number of nucleus replacements under development. The important concern is how well these implants duplicate the mechanical function of the native nucleus. This finite element model study aimed to investigate the influence of different nucleus replacements on the mechanical response of the disc. Models included partial, full, over-sized, partially saturated, elastic and poroelastic solid replacements. Over-sized nucleus replacements up to 25{\%} yielded results that were comparable to those in the intact state. Differences were much greater in cases with under-sized nucleus replacements. The effect was most pronounced for the 75{\%} under-sized replacement that resembled the condition with a full nucleotomy. Nucleus implants with elastic properties substantially altered load transmission when 10{\%} under-sized and over-sized replacements were considered. Compared to intact, the under-sized implants should be avoided when using biphasic materials with properties similar to the native nucleus, whereas for elastic replacements both under- and over-sized implants should not be used.",
keywords = "finite element analysis, implants, intervertebral disc, nucleus replacement, poroelastic",
author = "Hendrik Schmidt and Maxim Bashkuev and Fabio Galbusera and Wilke, {Hans Joachim} and Aboulfazl Shirazi-Adl",
year = "2014",
doi = "10.1080/10255842.2013.766722",
language = "English",
volume = "17",
pages = "1762--1776",
journal = "Computer Methods in Biomechanics and Biomedical Engineering",
issn = "1025-5842",
publisher = "Informa Healthcare",
number = "16",

}

TY - JOUR

T1 - Finite element study of human lumbar disc nucleus replacements

AU - Schmidt, Hendrik

AU - Bashkuev, Maxim

AU - Galbusera, Fabio

AU - Wilke, Hans Joachim

AU - Shirazi-Adl, Aboulfazl

PY - 2014

Y1 - 2014

N2 - Currently, there are a number of nucleus replacements under development. The important concern is how well these implants duplicate the mechanical function of the native nucleus. This finite element model study aimed to investigate the influence of different nucleus replacements on the mechanical response of the disc. Models included partial, full, over-sized, partially saturated, elastic and poroelastic solid replacements. Over-sized nucleus replacements up to 25% yielded results that were comparable to those in the intact state. Differences were much greater in cases with under-sized nucleus replacements. The effect was most pronounced for the 75% under-sized replacement that resembled the condition with a full nucleotomy. Nucleus implants with elastic properties substantially altered load transmission when 10% under-sized and over-sized replacements were considered. Compared to intact, the under-sized implants should be avoided when using biphasic materials with properties similar to the native nucleus, whereas for elastic replacements both under- and over-sized implants should not be used.

AB - Currently, there are a number of nucleus replacements under development. The important concern is how well these implants duplicate the mechanical function of the native nucleus. This finite element model study aimed to investigate the influence of different nucleus replacements on the mechanical response of the disc. Models included partial, full, over-sized, partially saturated, elastic and poroelastic solid replacements. Over-sized nucleus replacements up to 25% yielded results that were comparable to those in the intact state. Differences were much greater in cases with under-sized nucleus replacements. The effect was most pronounced for the 75% under-sized replacement that resembled the condition with a full nucleotomy. Nucleus implants with elastic properties substantially altered load transmission when 10% under-sized and over-sized replacements were considered. Compared to intact, the under-sized implants should be avoided when using biphasic materials with properties similar to the native nucleus, whereas for elastic replacements both under- and over-sized implants should not be used.

KW - finite element analysis

KW - implants

KW - intervertebral disc

KW - nucleus replacement

KW - poroelastic

UR - http://www.scopus.com/inward/record.url?scp=84901652393&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84901652393&partnerID=8YFLogxK

U2 - 10.1080/10255842.2013.766722

DO - 10.1080/10255842.2013.766722

M3 - Article

C2 - 23477684

AN - SCOPUS:84901652393

VL - 17

SP - 1762

EP - 1776

JO - Computer Methods in Biomechanics and Biomedical Engineering

JF - Computer Methods in Biomechanics and Biomedical Engineering

SN - 1025-5842

IS - 16