Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance

Antonella Fogliata, Jens Fleckenstein, Frank Schneider, Marc Pachoud, Sarah Ghandour, Harald Krauss, Giacomo Reggiori, Antonella Stravato, Frank Lohr, Marta Scorsetti, Luca Cozzi

Research output: Contribution to journalArticle

Abstract

Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators are today clinically used for stereotactical and non-stereotactical radiotherapy treatments. Such beams differ from the standard flattened beams (FF) in the high dose rate and the profile shape peaked on the beam central axis. Definition of new parameters as unflatness and slope for FFF beams has been proposed based on a renormalization factor for FFF profiles. The present study aims to assess the dosimetric differences between FFF beams generated by linear accelerators from different vendors and to provide renormalization and parameter data of the two kinds of units. Methods: Dosimetric data from two Varian TrueBeam and two Elekta Versa HD linear accelerators all with 6 and 10 MV nominal accelerating potentials, FF and FFF modes have been collected. Renormalization factors and related fit parameters according to Fogliata et al. ["Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy," Med. Phys. 39, 6455-6464 (2012)] have been evaluated for FFF beams of both units and energies. Unflatness and slope parameters from profile curves were evaluated. Dosimetric differences in terms of beam penetration and near-the-surface dose were also assessed. Results: FFF profile parameters have been updated; renormalization factors and unflatness from the Varian units are consistent with the published data. Elekta FFF beam qualities, different from the Varian generated beams, tend to express similar behaviour as the FF beam of the corresponding nominal energy. TPR20,10 for 6 and 10 MV FF and FFF TrueBeam beams are 0.665, 0.629 (6 MV) and 0.738, 0.703 (10 MV). The same figures for Versa HD units are 0.684, 0.678 (6 MV) and 0.734 0.721 (10 MV). Conclusions: Renormalization factor and unflatness parameters evaluated from Varian and Elekta FFF beams are provided, in particular renormalization factors table and fit parameters.

Original languageEnglish
Pages (from-to)205-212
Number of pages8
JournalMedical Physics
Volume43
Issue number1
DOIs
Publication statusPublished - Jan 1 2016

Fingerprint

Particle Accelerators
Radiotherapy
Photons
Therapeutics

Keywords

  • dosimetric parameters
  • flattening filter free beams

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Fogliata, A., Fleckenstein, J., Schneider, F., Pachoud, M., Ghandour, S., Krauss, H., ... Cozzi, L. (2016). Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance. Medical Physics, 43(1), 205-212. https://doi.org/10.1118/1.4938060

Flattening filter free beams from TrueBeam and Versa HD units : Evaluation of the parameters for quality assurance. / Fogliata, Antonella; Fleckenstein, Jens; Schneider, Frank; Pachoud, Marc; Ghandour, Sarah; Krauss, Harald; Reggiori, Giacomo; Stravato, Antonella; Lohr, Frank; Scorsetti, Marta; Cozzi, Luca.

In: Medical Physics, Vol. 43, No. 1, 01.01.2016, p. 205-212.

Research output: Contribution to journalArticle

Fogliata, A, Fleckenstein, J, Schneider, F, Pachoud, M, Ghandour, S, Krauss, H, Reggiori, G, Stravato, A, Lohr, F, Scorsetti, M & Cozzi, L 2016, 'Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance', Medical Physics, vol. 43, no. 1, pp. 205-212. https://doi.org/10.1118/1.4938060
Fogliata A, Fleckenstein J, Schneider F, Pachoud M, Ghandour S, Krauss H et al. Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance. Medical Physics. 2016 Jan 1;43(1):205-212. https://doi.org/10.1118/1.4938060
Fogliata, Antonella ; Fleckenstein, Jens ; Schneider, Frank ; Pachoud, Marc ; Ghandour, Sarah ; Krauss, Harald ; Reggiori, Giacomo ; Stravato, Antonella ; Lohr, Frank ; Scorsetti, Marta ; Cozzi, Luca. / Flattening filter free beams from TrueBeam and Versa HD units : Evaluation of the parameters for quality assurance. In: Medical Physics. 2016 ; Vol. 43, No. 1. pp. 205-212.
@article{92af7914615d45eaa8e10e01edc2e056,
title = "Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance",
abstract = "Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators are today clinically used for stereotactical and non-stereotactical radiotherapy treatments. Such beams differ from the standard flattened beams (FF) in the high dose rate and the profile shape peaked on the beam central axis. Definition of new parameters as unflatness and slope for FFF beams has been proposed based on a renormalization factor for FFF profiles. The present study aims to assess the dosimetric differences between FFF beams generated by linear accelerators from different vendors and to provide renormalization and parameter data of the two kinds of units. Methods: Dosimetric data from two Varian TrueBeam and two Elekta Versa HD linear accelerators all with 6 and 10 MV nominal accelerating potentials, FF and FFF modes have been collected. Renormalization factors and related fit parameters according to Fogliata et al. [{"}Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy,{"} Med. Phys. 39, 6455-6464 (2012)] have been evaluated for FFF beams of both units and energies. Unflatness and slope parameters from profile curves were evaluated. Dosimetric differences in terms of beam penetration and near-the-surface dose were also assessed. Results: FFF profile parameters have been updated; renormalization factors and unflatness from the Varian units are consistent with the published data. Elekta FFF beam qualities, different from the Varian generated beams, tend to express similar behaviour as the FF beam of the corresponding nominal energy. TPR20,10 for 6 and 10 MV FF and FFF TrueBeam beams are 0.665, 0.629 (6 MV) and 0.738, 0.703 (10 MV). The same figures for Versa HD units are 0.684, 0.678 (6 MV) and 0.734 0.721 (10 MV). Conclusions: Renormalization factor and unflatness parameters evaluated from Varian and Elekta FFF beams are provided, in particular renormalization factors table and fit parameters.",
keywords = "dosimetric parameters, flattening filter free beams",
author = "Antonella Fogliata and Jens Fleckenstein and Frank Schneider and Marc Pachoud and Sarah Ghandour and Harald Krauss and Giacomo Reggiori and Antonella Stravato and Frank Lohr and Marta Scorsetti and Luca Cozzi",
year = "2016",
month = "1",
day = "1",
doi = "10.1118/1.4938060",
language = "English",
volume = "43",
pages = "205--212",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "John Wiley and Sons Ltd",
number = "1",

}

TY - JOUR

T1 - Flattening filter free beams from TrueBeam and Versa HD units

T2 - Evaluation of the parameters for quality assurance

AU - Fogliata, Antonella

AU - Fleckenstein, Jens

AU - Schneider, Frank

AU - Pachoud, Marc

AU - Ghandour, Sarah

AU - Krauss, Harald

AU - Reggiori, Giacomo

AU - Stravato, Antonella

AU - Lohr, Frank

AU - Scorsetti, Marta

AU - Cozzi, Luca

PY - 2016/1/1

Y1 - 2016/1/1

N2 - Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators are today clinically used for stereotactical and non-stereotactical radiotherapy treatments. Such beams differ from the standard flattened beams (FF) in the high dose rate and the profile shape peaked on the beam central axis. Definition of new parameters as unflatness and slope for FFF beams has been proposed based on a renormalization factor for FFF profiles. The present study aims to assess the dosimetric differences between FFF beams generated by linear accelerators from different vendors and to provide renormalization and parameter data of the two kinds of units. Methods: Dosimetric data from two Varian TrueBeam and two Elekta Versa HD linear accelerators all with 6 and 10 MV nominal accelerating potentials, FF and FFF modes have been collected. Renormalization factors and related fit parameters according to Fogliata et al. ["Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy," Med. Phys. 39, 6455-6464 (2012)] have been evaluated for FFF beams of both units and energies. Unflatness and slope parameters from profile curves were evaluated. Dosimetric differences in terms of beam penetration and near-the-surface dose were also assessed. Results: FFF profile parameters have been updated; renormalization factors and unflatness from the Varian units are consistent with the published data. Elekta FFF beam qualities, different from the Varian generated beams, tend to express similar behaviour as the FF beam of the corresponding nominal energy. TPR20,10 for 6 and 10 MV FF and FFF TrueBeam beams are 0.665, 0.629 (6 MV) and 0.738, 0.703 (10 MV). The same figures for Versa HD units are 0.684, 0.678 (6 MV) and 0.734 0.721 (10 MV). Conclusions: Renormalization factor and unflatness parameters evaluated from Varian and Elekta FFF beams are provided, in particular renormalization factors table and fit parameters.

AB - Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators are today clinically used for stereotactical and non-stereotactical radiotherapy treatments. Such beams differ from the standard flattened beams (FF) in the high dose rate and the profile shape peaked on the beam central axis. Definition of new parameters as unflatness and slope for FFF beams has been proposed based on a renormalization factor for FFF profiles. The present study aims to assess the dosimetric differences between FFF beams generated by linear accelerators from different vendors and to provide renormalization and parameter data of the two kinds of units. Methods: Dosimetric data from two Varian TrueBeam and two Elekta Versa HD linear accelerators all with 6 and 10 MV nominal accelerating potentials, FF and FFF modes have been collected. Renormalization factors and related fit parameters according to Fogliata et al. ["Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy," Med. Phys. 39, 6455-6464 (2012)] have been evaluated for FFF beams of both units and energies. Unflatness and slope parameters from profile curves were evaluated. Dosimetric differences in terms of beam penetration and near-the-surface dose were also assessed. Results: FFF profile parameters have been updated; renormalization factors and unflatness from the Varian units are consistent with the published data. Elekta FFF beam qualities, different from the Varian generated beams, tend to express similar behaviour as the FF beam of the corresponding nominal energy. TPR20,10 for 6 and 10 MV FF and FFF TrueBeam beams are 0.665, 0.629 (6 MV) and 0.738, 0.703 (10 MV). The same figures for Versa HD units are 0.684, 0.678 (6 MV) and 0.734 0.721 (10 MV). Conclusions: Renormalization factor and unflatness parameters evaluated from Varian and Elekta FFF beams are provided, in particular renormalization factors table and fit parameters.

KW - dosimetric parameters

KW - flattening filter free beams

UR - http://www.scopus.com/inward/record.url?scp=84952683363&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84952683363&partnerID=8YFLogxK

U2 - 10.1118/1.4938060

DO - 10.1118/1.4938060

M3 - Article

AN - SCOPUS:84952683363

VL - 43

SP - 205

EP - 212

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 1

ER -