Fracture of a Supera Interwoven Nitinol Stent after Treatment of Popliteal Artery Stenosis

T Cambiaghi, A Spertino, L Bertoglio, R Chiesa

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To present a Supera stent fracture following treatment of popliteal artery stenosis. Case Report: A 60-year-old man previously treated with angioplasty/stenting of a popliteal artery lesion with a Supera stent presented at 8 months with in-stent occlusion that proved to be secondary to stent fracture. Conclusion: Supera interwoven nitinol stents, despite having higher radial force and conformability than classic tube stents, are not free from fracture. Moreover, due to their peculiar structure, a break in these stents results in complete loss of integrity and consequent device collapse, thus jeopardizing endovascular recanalization and relining. © SAGE Publications.
Original languageEnglish
Pages (from-to)447-449
Number of pages3
JournalJournal of Endovascular Therapy
Volume24
Issue number3
DOIs
Publication statusPublished - 2017

Fingerprint Dive into the research topics of 'Fracture of a Supera Interwoven Nitinol Stent after Treatment of Popliteal Artery Stenosis'. Together they form a unique fingerprint.