Dalla TC alla stampante 3D: Un metodo innovativo per la pianificazione chirurgica di una protesi a conduzione ossea transcutanea

Translated title of the contribution: From CT scanning to 3D printing technology: A new method for the preoperative planning of a transcutaneous bone-conduction hearing device

Pietro Canzi, S. Marconi, M. Manfrin, M. Magnetto, C. Carelli, A. M. Simoncelli, D. Fresa, M. Beltrame, F. Auricchio, M. Benazzo

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

The aim of the present study was to assess the feasibility and utility of 3D printing technology in surgical planning of a transcutaneous bone-conduction hearing device (Bonebridge®) (BB), focusing on the identification of the proper location and placement of the transducer. 3D printed (3DP) models of three human cadaveric temporal bones, previously submitted to CT scan, were created with the representation of a topographic bone thickness map and the sinus pathway on the outer surface. The 3DP model was used to detect the most suitable location for the BB. A 3DP transparent mask that faithfully reproduced the surface of both the temporal bone and the 3DP model was also developed to correctly transfer the designated BB area. The accuracy of the procedure was verified by CT scan: a radiological marker was used to evaluate the degree of correspondence of the transducer site between the 3DP model and the human temporal bone. The BB positioning was successfully performed on all human temporal bones, with no difficulties in finding the proper location of the transducer. A mean error of 0.13 mm was found when the transducer site of the 3DP model was compared to that of the human temporal bone. The employment of 3D printing technology in surgical planning of BB positioning showed feasible results. Further studies will be required to evaluate its clinical applicability.

Original languageItalian
Pages (from-to)251-256
Number of pages6
JournalActa Otorhinolaryngologica Italica
Volume38
Issue number3
DOIs
Publication statusPublished - Jun 1 2018

Fingerprint

Bone Conduction
Temporal Bone
Transducers
Technology
Equipment and Supplies
Masks
Three Dimensional Printing
Bone and Bones

ASJC Scopus subject areas

  • Otorhinolaryngology

Cite this

Dalla TC alla stampante 3D : Un metodo innovativo per la pianificazione chirurgica di una protesi a conduzione ossea transcutanea. / Canzi, Pietro; Marconi, S.; Manfrin, M.; Magnetto, M.; Carelli, C.; Simoncelli, A. M.; Fresa, D.; Beltrame, M.; Auricchio, F.; Benazzo, M.

In: Acta Otorhinolaryngologica Italica, Vol. 38, No. 3, 01.06.2018, p. 251-256.

Research output: Contribution to journalArticle

@article{ccf3feca325b454490a07ad4a9a859f2,
title = "Dalla TC alla stampante 3D: Un metodo innovativo per la pianificazione chirurgica di una protesi a conduzione ossea transcutanea",
abstract = "The aim of the present study was to assess the feasibility and utility of 3D printing technology in surgical planning of a transcutaneous bone-conduction hearing device (Bonebridge{\circledR}) (BB), focusing on the identification of the proper location and placement of the transducer. 3D printed (3DP) models of three human cadaveric temporal bones, previously submitted to CT scan, were created with the representation of a topographic bone thickness map and the sinus pathway on the outer surface. The 3DP model was used to detect the most suitable location for the BB. A 3DP transparent mask that faithfully reproduced the surface of both the temporal bone and the 3DP model was also developed to correctly transfer the designated BB area. The accuracy of the procedure was verified by CT scan: a radiological marker was used to evaluate the degree of correspondence of the transducer site between the 3DP model and the human temporal bone. The BB positioning was successfully performed on all human temporal bones, with no difficulties in finding the proper location of the transducer. A mean error of 0.13 mm was found when the transducer site of the 3DP model was compared to that of the human temporal bone. The employment of 3D printing technology in surgical planning of BB positioning showed feasible results. Further studies will be required to evaluate its clinical applicability.",
keywords = "3D printing, Bone conduction implant, Bonebridge, Implantable hearing aid, Surgical planning",
author = "Pietro Canzi and S. Marconi and M. Manfrin and M. Magnetto and C. Carelli and Simoncelli, {A. M.} and D. Fresa and M. Beltrame and F. Auricchio and M. Benazzo",
year = "2018",
month = "6",
day = "1",
doi = "10.14639/0392-100X-1625",
language = "Italian",
volume = "38",
pages = "251--256",
journal = "Acta Otorhinolaryngologica Italica",
issn = "0392-100X",
publisher = "Pacini Editore s.r.l.",
number = "3",

}

TY - JOUR

T1 - Dalla TC alla stampante 3D

T2 - Un metodo innovativo per la pianificazione chirurgica di una protesi a conduzione ossea transcutanea

AU - Canzi, Pietro

AU - Marconi, S.

AU - Manfrin, M.

AU - Magnetto, M.

AU - Carelli, C.

AU - Simoncelli, A. M.

AU - Fresa, D.

AU - Beltrame, M.

AU - Auricchio, F.

AU - Benazzo, M.

PY - 2018/6/1

Y1 - 2018/6/1

N2 - The aim of the present study was to assess the feasibility and utility of 3D printing technology in surgical planning of a transcutaneous bone-conduction hearing device (Bonebridge®) (BB), focusing on the identification of the proper location and placement of the transducer. 3D printed (3DP) models of three human cadaveric temporal bones, previously submitted to CT scan, were created with the representation of a topographic bone thickness map and the sinus pathway on the outer surface. The 3DP model was used to detect the most suitable location for the BB. A 3DP transparent mask that faithfully reproduced the surface of both the temporal bone and the 3DP model was also developed to correctly transfer the designated BB area. The accuracy of the procedure was verified by CT scan: a radiological marker was used to evaluate the degree of correspondence of the transducer site between the 3DP model and the human temporal bone. The BB positioning was successfully performed on all human temporal bones, with no difficulties in finding the proper location of the transducer. A mean error of 0.13 mm was found when the transducer site of the 3DP model was compared to that of the human temporal bone. The employment of 3D printing technology in surgical planning of BB positioning showed feasible results. Further studies will be required to evaluate its clinical applicability.

AB - The aim of the present study was to assess the feasibility and utility of 3D printing technology in surgical planning of a transcutaneous bone-conduction hearing device (Bonebridge®) (BB), focusing on the identification of the proper location and placement of the transducer. 3D printed (3DP) models of three human cadaveric temporal bones, previously submitted to CT scan, were created with the representation of a topographic bone thickness map and the sinus pathway on the outer surface. The 3DP model was used to detect the most suitable location for the BB. A 3DP transparent mask that faithfully reproduced the surface of both the temporal bone and the 3DP model was also developed to correctly transfer the designated BB area. The accuracy of the procedure was verified by CT scan: a radiological marker was used to evaluate the degree of correspondence of the transducer site between the 3DP model and the human temporal bone. The BB positioning was successfully performed on all human temporal bones, with no difficulties in finding the proper location of the transducer. A mean error of 0.13 mm was found when the transducer site of the 3DP model was compared to that of the human temporal bone. The employment of 3D printing technology in surgical planning of BB positioning showed feasible results. Further studies will be required to evaluate its clinical applicability.

KW - 3D printing

KW - Bone conduction implant

KW - Bonebridge

KW - Implantable hearing aid

KW - Surgical planning

UR - http://www.scopus.com/inward/record.url?scp=85046994610&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85046994610&partnerID=8YFLogxK

U2 - 10.14639/0392-100X-1625

DO - 10.14639/0392-100X-1625

M3 - Articolo

AN - SCOPUS:85046994610

VL - 38

SP - 251

EP - 256

JO - Acta Otorhinolaryngologica Italica

JF - Acta Otorhinolaryngologica Italica

SN - 0392-100X

IS - 3

ER -