Functional evaluation of a CPAP circuit with a high compliance reservoir bag

A. Braschi, G. Iotti, A. Locatelli, G. Bellinzona

Research output: Contribution to journalArticle

Abstract

Continuous positive airway pressure is widely used in the treatment of ARF and an evaluation of the systems is important. The authors used an artificial model to test a continuous flow system with a high compliance reservoir bag. The results confirm that the system is effective in maintaining positive pressure stability within a wide range of inspiratory peak flow rates, even when a low fresh gas flow rate is employed. Nevertheless, rebreathing of expired gases is possible and may be noticeable at high expiratory flow rates, caused by the high compliance of the reservoir bag.

Original languageEnglish
Pages (from-to)85-89
Number of pages5
JournalIntensive Care Medicine
Volume11
Issue number2
DOIs
Publication statusPublished - Mar 1985

Keywords

  • Airway pressure
  • CPAP
  • Rebreathing

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Fingerprint Dive into the research topics of 'Functional evaluation of a CPAP circuit with a high compliance reservoir bag'. Together they form a unique fingerprint.