High flow biphasic positive airway pressure by helmet - Effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure

Onnen Moerer, Peter Herrmann, José Hinz, Paolo Severgnini, Edoardo Calderini, Michael Quintel, Paolo Pelosi

Research output: Contribution to journalArticlepeer-review

Abstract

Introduction: Non-invasive ventilation (NIV) with a helmet device is often associated with poor patient-ventilator synchrony and impaired carbon dioxide (CO2) removal, which might lead to failure. A possible solution is to use a high free flow system in combination with a time-cycled pressure valve placed into the expiratory circuit (HF-BiPAP). This system would be independent from triggering while providing a high flow to eliminate CO2. Methods: Conventional pressure support ventilation (PSV) and time-cycled biphasic pressure controlled ventilation (BiVent) delivered by an Intensive Care Unit ventilator were compared to HF-BiPAP in an in vitro lung model study. Variables included delta pressures of 5 and 15 cmH2O, respiratory rates of 15 and 30 breaths/min, inspiratory efforts (respiratory drive) of 2.5 and 10 cmH2O) and different lung characteristics. Additionally, CO2 removal and noise exposure were measured. Results: Pressurization during inspiration was more effective with pressure controlled modes compared to PSV (P <0.001) at similar tidal volumes. During the expiratory phase, BiVent and HF-BiPAP led to an increase in pressure burden compared to PSV. This was especially true at higher upper pressures (P <0.001). At high level of asynchrony both HF-BiPAP and BiVent were less effective. Only HF-BiPAP ventilation effectively removed CO2 (P <0.001) during all settings. Noise exposure was higher during HF-BiPAP (P <0.001). Conclusions: This study demonstrates that in a lung model, the efficiency of NIV by helmet can be improved by using HF-BiPAP. However, it imposes a higher pressure during the expiratory phase. CO2 was almost completely removed with HF-BiPAP during all settings.

Original languageEnglish
Article numberR85
JournalCritical Care
Volume13
Issue number3
DOIs
Publication statusPublished - Jun 5 2009

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine
  • Medicine(all)

Fingerprint Dive into the research topics of 'High flow biphasic positive airway pressure by helmet - Effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure'. Together they form a unique fingerprint.

Cite this