Human errors in the calculation of monitor units in clinical radiotherapy practice

Riccardo Calandrino, Giovanni Mauro Cattaneo, Antonella Del Vecchio, Claudio Morino, Barbara Longobardi, Patrizia Signorotto

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

Human mistakes are an important source of error in all steps of radiotherapy planning and their incidence should be investigated. As has been recognized by different authors and by the ICRU [4], the human error rate in the calculation of monitor units (MU) is relatively high. At our institute, we measured the human error rate in the calculation of MU by an independent check of the calculation. From September 1991 to June 1992 we identified and corrected 17 serious errors (deviation from the prescribed dose ≥ 5%) over 890 controls (l.9%) (daily dose errors). We also found a serious global dose error rate (i.e. the errors induced on the total reference dose for the complete course of the treatment) of l.3% ( 9 685) during the period November 1991-June 1992. These values suggest the importance of human errors in the calculation of MU and also confirm the validity of the independent check of MU calculation as one of the simplest ways of avoiding erroneous dose delivery by incorrect calculation of MU.

Original languageEnglish
Pages (from-to)86-88
Number of pages3
JournalRadiotherapy and Oncology
Volume28
Issue number1
DOIs
Publication statusPublished - 1993

Fingerprint

Radiotherapy
Research Design
Incidence
Therapeutics

Keywords

  • Human errors
  • Monitor unit calculation
  • Quality Assurance

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Urology

Cite this

Human errors in the calculation of monitor units in clinical radiotherapy practice. / Calandrino, Riccardo; Cattaneo, Giovanni Mauro; Del Vecchio, Antonella; Morino, Claudio; Longobardi, Barbara; Signorotto, Patrizia.

In: Radiotherapy and Oncology, Vol. 28, No. 1, 1993, p. 86-88.

Research output: Contribution to journalArticle

Calandrino, Riccardo ; Cattaneo, Giovanni Mauro ; Del Vecchio, Antonella ; Morino, Claudio ; Longobardi, Barbara ; Signorotto, Patrizia. / Human errors in the calculation of monitor units in clinical radiotherapy practice. In: Radiotherapy and Oncology. 1993 ; Vol. 28, No. 1. pp. 86-88.
@article{1ce202be4f1943cab457dbf9e1c39ddd,
title = "Human errors in the calculation of monitor units in clinical radiotherapy practice",
abstract = "Human mistakes are an important source of error in all steps of radiotherapy planning and their incidence should be investigated. As has been recognized by different authors and by the ICRU [4], the human error rate in the calculation of monitor units (MU) is relatively high. At our institute, we measured the human error rate in the calculation of MU by an independent check of the calculation. From September 1991 to June 1992 we identified and corrected 17 serious errors (deviation from the prescribed dose ≥ 5{\%}) over 890 controls (l.9{\%}) (daily dose errors). We also found a serious global dose error rate (i.e. the errors induced on the total reference dose for the complete course of the treatment) of l.3{\%} ( 9 685) during the period November 1991-June 1992. These values suggest the importance of human errors in the calculation of MU and also confirm the validity of the independent check of MU calculation as one of the simplest ways of avoiding erroneous dose delivery by incorrect calculation of MU.",
keywords = "Human errors, Monitor unit calculation, Quality Assurance",
author = "Riccardo Calandrino and Cattaneo, {Giovanni Mauro} and {Del Vecchio}, Antonella and Claudio Morino and Barbara Longobardi and Patrizia Signorotto",
year = "1993",
doi = "10.1016/0167-8140(93)90193-C",
language = "English",
volume = "28",
pages = "86--88",
journal = "Radiotherapy and Oncology",
issn = "0167-8140",
publisher = "Elsevier Ireland Ltd",
number = "1",

}

TY - JOUR

T1 - Human errors in the calculation of monitor units in clinical radiotherapy practice

AU - Calandrino, Riccardo

AU - Cattaneo, Giovanni Mauro

AU - Del Vecchio, Antonella

AU - Morino, Claudio

AU - Longobardi, Barbara

AU - Signorotto, Patrizia

PY - 1993

Y1 - 1993

N2 - Human mistakes are an important source of error in all steps of radiotherapy planning and their incidence should be investigated. As has been recognized by different authors and by the ICRU [4], the human error rate in the calculation of monitor units (MU) is relatively high. At our institute, we measured the human error rate in the calculation of MU by an independent check of the calculation. From September 1991 to June 1992 we identified and corrected 17 serious errors (deviation from the prescribed dose ≥ 5%) over 890 controls (l.9%) (daily dose errors). We also found a serious global dose error rate (i.e. the errors induced on the total reference dose for the complete course of the treatment) of l.3% ( 9 685) during the period November 1991-June 1992. These values suggest the importance of human errors in the calculation of MU and also confirm the validity of the independent check of MU calculation as one of the simplest ways of avoiding erroneous dose delivery by incorrect calculation of MU.

AB - Human mistakes are an important source of error in all steps of radiotherapy planning and their incidence should be investigated. As has been recognized by different authors and by the ICRU [4], the human error rate in the calculation of monitor units (MU) is relatively high. At our institute, we measured the human error rate in the calculation of MU by an independent check of the calculation. From September 1991 to June 1992 we identified and corrected 17 serious errors (deviation from the prescribed dose ≥ 5%) over 890 controls (l.9%) (daily dose errors). We also found a serious global dose error rate (i.e. the errors induced on the total reference dose for the complete course of the treatment) of l.3% ( 9 685) during the period November 1991-June 1992. These values suggest the importance of human errors in the calculation of MU and also confirm the validity of the independent check of MU calculation as one of the simplest ways of avoiding erroneous dose delivery by incorrect calculation of MU.

KW - Human errors

KW - Monitor unit calculation

KW - Quality Assurance

UR - http://www.scopus.com/inward/record.url?scp=0027482664&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027482664&partnerID=8YFLogxK

U2 - 10.1016/0167-8140(93)90193-C

DO - 10.1016/0167-8140(93)90193-C

M3 - Article

C2 - 8234876

AN - SCOPUS:0027482664

VL - 28

SP - 86

EP - 88

JO - Radiotherapy and Oncology

JF - Radiotherapy and Oncology

SN - 0167-8140

IS - 1

ER -