Abstract
Background: Technological improvement has contributed to making veno-venous extracorporeal membrane oxygenation (VV-ECMO) safer and easier, spreading its use in acute respiratory failure (ARF).
Methods: This is a retrospective observational study carried out in the ECMO center at IRCCS-ISMETT, a medical center focused on end-stage organ failure treatment in Italy. We investigated the effect of different cannula designs on the amount of blood product transfused. Eighty-nine consecutive patients affected with ARF on VV-ECMO from 2008 to 2016 were compared according to type of cannulation: older percutaneous cannula (Standard group, 52 patients) and HLS© BIOLINE-coated, but with shorter drainage cannula (BIOLINE group, 37 patients).
Results: The two study groups were comparable in terms of baseline characteristics [age, body mass index (BMI), Simplified Acute Physiology Score (SAPS-II), Sequential Organ Failure Assessment (SOFA), Predicting Death For Severe ARDS on VV-ECMO (PRESERVE) score] and ECMO management [median hematocrit (Htc), platelet nadir, antithrombin III (AT III), heparin, activated partial thromboplastin time (APTT)]. In the BIOLINE group, a lower amount of packed red blood cells (pRBC) was transfused considering both total number [4 units, interquartile range (IQR) 1-9 vs. 12 units, IQR 5.5-21; P<0.01] and mL of pRBC/day of ECMO support (91, IQR 21-158 vs. 193.5, IQR 140.5-254; P<0.01). In the BIOLINE group, a trend in reduction of ECMO days (P=0.05) and length of intensive care unit (ICU) stay was found (P=0.06), but no differences in rates of ECMO weaning and ICU discharge were evidenced. The BIOLINE group constituted a saving of €1,295.20 per patient/treatment, counting the costs for cannulation and pRBC administration.
Conclusions: More biocompatible and shorter drainage cannula may represent one of the contributing factors to a reduction in transfusions and costs of VV-ECMO in the current ongoing technological improvement in ECMO.
Original language | English |
---|---|
Pages (from-to) | 5813-5821 |
Number of pages | 9 |
Journal | Journal of Thoracic Disease |
Volume | 10 |
Issue number | 10 |
DOIs | |
Publication status | Published - Oct 2018 |
Fingerprint
Cite this
Impact of cannula design on packed red blood cell transfusions : technical advancement to improve outcomes in extracorporeal membrane oxygenation. / Martucci, Gennaro; Panarello, Giovanna; Occhipinti, Giovanna; Raffa, Giuseppe; Tuzzolino, Fabio; Capitanio, Guido; Carollo, Tiziana; Lino, Giovanni; Bertani, Alessandro; Vitulo, Patrizio; Pilato, Michele; Lorusso, Roberto; Arcadipane, Antonio.
In: Journal of Thoracic Disease, Vol. 10, No. 10, 10.2018, p. 5813-5821.Research output: Contribution to journal › Article
}
TY - JOUR
T1 - Impact of cannula design on packed red blood cell transfusions
T2 - technical advancement to improve outcomes in extracorporeal membrane oxygenation
AU - Martucci, Gennaro
AU - Panarello, Giovanna
AU - Occhipinti, Giovanna
AU - Raffa, Giuseppe
AU - Tuzzolino, Fabio
AU - Capitanio, Guido
AU - Carollo, Tiziana
AU - Lino, Giovanni
AU - Bertani, Alessandro
AU - Vitulo, Patrizio
AU - Pilato, Michele
AU - Lorusso, Roberto
AU - Arcadipane, Antonio
PY - 2018/10
Y1 - 2018/10
N2 - Background: Technological improvement has contributed to making veno-venous extracorporeal membrane oxygenation (VV-ECMO) safer and easier, spreading its use in acute respiratory failure (ARF).Methods: This is a retrospective observational study carried out in the ECMO center at IRCCS-ISMETT, a medical center focused on end-stage organ failure treatment in Italy. We investigated the effect of different cannula designs on the amount of blood product transfused. Eighty-nine consecutive patients affected with ARF on VV-ECMO from 2008 to 2016 were compared according to type of cannulation: older percutaneous cannula (Standard group, 52 patients) and HLS© BIOLINE-coated, but with shorter drainage cannula (BIOLINE group, 37 patients).Results: The two study groups were comparable in terms of baseline characteristics [age, body mass index (BMI), Simplified Acute Physiology Score (SAPS-II), Sequential Organ Failure Assessment (SOFA), Predicting Death For Severe ARDS on VV-ECMO (PRESERVE) score] and ECMO management [median hematocrit (Htc), platelet nadir, antithrombin III (AT III), heparin, activated partial thromboplastin time (APTT)]. In the BIOLINE group, a lower amount of packed red blood cells (pRBC) was transfused considering both total number [4 units, interquartile range (IQR) 1-9 vs. 12 units, IQR 5.5-21; P<0.01] and mL of pRBC/day of ECMO support (91, IQR 21-158 vs. 193.5, IQR 140.5-254; P<0.01). In the BIOLINE group, a trend in reduction of ECMO days (P=0.05) and length of intensive care unit (ICU) stay was found (P=0.06), but no differences in rates of ECMO weaning and ICU discharge were evidenced. The BIOLINE group constituted a saving of €1,295.20 per patient/treatment, counting the costs for cannulation and pRBC administration.Conclusions: More biocompatible and shorter drainage cannula may represent one of the contributing factors to a reduction in transfusions and costs of VV-ECMO in the current ongoing technological improvement in ECMO.
AB - Background: Technological improvement has contributed to making veno-venous extracorporeal membrane oxygenation (VV-ECMO) safer and easier, spreading its use in acute respiratory failure (ARF).Methods: This is a retrospective observational study carried out in the ECMO center at IRCCS-ISMETT, a medical center focused on end-stage organ failure treatment in Italy. We investigated the effect of different cannula designs on the amount of blood product transfused. Eighty-nine consecutive patients affected with ARF on VV-ECMO from 2008 to 2016 were compared according to type of cannulation: older percutaneous cannula (Standard group, 52 patients) and HLS© BIOLINE-coated, but with shorter drainage cannula (BIOLINE group, 37 patients).Results: The two study groups were comparable in terms of baseline characteristics [age, body mass index (BMI), Simplified Acute Physiology Score (SAPS-II), Sequential Organ Failure Assessment (SOFA), Predicting Death For Severe ARDS on VV-ECMO (PRESERVE) score] and ECMO management [median hematocrit (Htc), platelet nadir, antithrombin III (AT III), heparin, activated partial thromboplastin time (APTT)]. In the BIOLINE group, a lower amount of packed red blood cells (pRBC) was transfused considering both total number [4 units, interquartile range (IQR) 1-9 vs. 12 units, IQR 5.5-21; P<0.01] and mL of pRBC/day of ECMO support (91, IQR 21-158 vs. 193.5, IQR 140.5-254; P<0.01). In the BIOLINE group, a trend in reduction of ECMO days (P=0.05) and length of intensive care unit (ICU) stay was found (P=0.06), but no differences in rates of ECMO weaning and ICU discharge were evidenced. The BIOLINE group constituted a saving of €1,295.20 per patient/treatment, counting the costs for cannulation and pRBC administration.Conclusions: More biocompatible and shorter drainage cannula may represent one of the contributing factors to a reduction in transfusions and costs of VV-ECMO in the current ongoing technological improvement in ECMO.
U2 - 10.21037/jtd.2018.09.119
DO - 10.21037/jtd.2018.09.119
M3 - Article
C2 - 30505489
VL - 10
SP - 5813
EP - 5821
JO - Journal of Thoracic Disease
JF - Journal of Thoracic Disease
SN - 2072-1439
IS - 10
ER -