TY - JOUR
T1 - In vitro effects on mobile polyethylene insert under highly demanding daily activities
T2 - stair climbing
AU - Jaber, Sami Abdel
AU - Taddei, Paola
AU - Tozzi, Silvia
AU - Sudanese, Alessandra
AU - Affatato, Saverio
PY - 2015
Y1 - 2015
N2 - Purpose: Wear and survival of total joint replacements do not depend on the duration of the implant in situ, but rather on the amount of its use, i.e. the patient’s activity level. With this in mind, the present study was driven by two questions: (1) How does total knee replacement (TKR) respond to the simulation of daily highly demanding activities? (2) Are certain activities to be advised against or, on the contrary, useful to implanted patients, in order to reduce wear of TKR and its related problems? Methods: One set of the same total knee prosthesis (TKP), equal in design and size, was tested on a three-plus-one knee joint simulator for two million cycles using a highly demanding daily load waveform, replicating a stair-climbing movement. The results were compared with a set of TKP previously tested with the ISO level walking task. A digital microscope was used to characterise the superficial structure of all the TKPs. Gravimetric and micro-Raman spectroscopic analyses were carried out on the polyethylene inserts. Visual comparison with in vivo explants was carried out. Results: The average volumetric mass loss after two million cycles was 44 ± 6 mm3. Microscope examinations showed some deep scratches along the flexion/extension movements for all the components. Also, the metallic backside surface showed intense non-linear scratches and the polyethylene counterface was characterised by some craters. A decrease in crystallinity, induced by mechanical stress was observed on all polyethylene components and was quantitatively confirmed by the orthorhombic fraction αovalue. Conclusions: The results of this study demonstrated that the forces and motion sustained by the knee are highly activity-dependent. Moreover, this test confirmed that under more severe conditions, the material properties change according to a different wear mechanism and a decrease in crystallinity occurs. Loading characteristics for specific activities should be considered for the design of functional and robust TKRs.
AB - Purpose: Wear and survival of total joint replacements do not depend on the duration of the implant in situ, but rather on the amount of its use, i.e. the patient’s activity level. With this in mind, the present study was driven by two questions: (1) How does total knee replacement (TKR) respond to the simulation of daily highly demanding activities? (2) Are certain activities to be advised against or, on the contrary, useful to implanted patients, in order to reduce wear of TKR and its related problems? Methods: One set of the same total knee prosthesis (TKP), equal in design and size, was tested on a three-plus-one knee joint simulator for two million cycles using a highly demanding daily load waveform, replicating a stair-climbing movement. The results were compared with a set of TKP previously tested with the ISO level walking task. A digital microscope was used to characterise the superficial structure of all the TKPs. Gravimetric and micro-Raman spectroscopic analyses were carried out on the polyethylene inserts. Visual comparison with in vivo explants was carried out. Results: The average volumetric mass loss after two million cycles was 44 ± 6 mm3. Microscope examinations showed some deep scratches along the flexion/extension movements for all the components. Also, the metallic backside surface showed intense non-linear scratches and the polyethylene counterface was characterised by some craters. A decrease in crystallinity, induced by mechanical stress was observed on all polyethylene components and was quantitatively confirmed by the orthorhombic fraction αovalue. Conclusions: The results of this study demonstrated that the forces and motion sustained by the knee are highly activity-dependent. Moreover, this test confirmed that under more severe conditions, the material properties change according to a different wear mechanism and a decrease in crystallinity occurs. Loading characteristics for specific activities should be considered for the design of functional and robust TKRs.
KW - Daily activity
KW - In vitro knee
KW - Knee simulator
KW - Mobile TKR
KW - Raman spectroscopy
UR - http://www.scopus.com/inward/record.url?scp=84932196864&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84932196864&partnerID=8YFLogxK
U2 - 10.1007/s00264-014-2622-5
DO - 10.1007/s00264-014-2622-5
M3 - Article
C2 - 25500956
AN - SCOPUS:84932196864
VL - 39
SP - 1433
EP - 1440
JO - International Orthopaedics
JF - International Orthopaedics
SN - 0341-2695
IS - 7
ER -