Incidence, predictors, and implications of access site complications with transfemoral transcatheter aortic valve implantation

Nicolas M. Van Mieghem, Didier Tchetche, Alaide Chieffo, Nicolas Dumonteil, David Messika-Zeitoun, Robert M A Van Der Boon, Olivier Vahdat, Gill L. Buchanan, Bertrand Marcheix, Dominique Himbert, Patrick W. Serruys, Jean Fajadet, Antonio Colombo, Didier Carrié, Alec Vahanian, Peter P T De Jaegere

Research output: Contribution to journalArticle

Abstract

Our study objective was to assess the incidence, predictors, and implications of access site complications related to transfemoral transcatheter aortic valve implantation (TAVI). We pooled the prospective TAVI databases of 5 experienced centers in Europe enrolling only transfemoral cases for this analysis. Access site complications were defined according to the Valve Academic Research Consortium end-point definitions. The global transfemoral TAVI database contained 986 patients. Percutaneous access and closure was performed in 803 patients (81%) and a surgical strategy in 183 (19%). Incidences of major vascular complications, life-threatening/disabling bleeding, and major bleeding were 14.2%, 11%, and 17.8% respectively. In the patient cohort with a completely percutaneous access strategy, major vascular complications and life-threatening/disabling bleedings were related to closure device failure in 64% and 29%, respectively. Female gender (odds ratio 1.63, 95% confidence interval 1.12 to 2.36) and use of >19Fr system (2.87, 1.68 to 4.91) were independent predictors for major vascular complications. Female gender (odds ratio 2.04, 95% confidence interval 1.31 to 3.17), use of >19Fr system (1.86, 1.02 to 3.38), peripheral arterial disease (2.14, 1.27 to 3.61), learning effect (0.45, 0.27 to 0.73), and percutaneous access strategy (2.39, 1.16 to 4.89) were independently associated with life-threatening/disabling bleedings. In conclusion, transfemoral TAVI is associated with a >10% incidence of major vascular-related complications. A considerable number of these events is related to arteriotomy closure failure. Arterial sheath size and female gender are important determinants of major vascular complications and life-threatening/disabling bleeding.

Original languageEnglish
Pages (from-to)1361-1367
Number of pages7
JournalThe American Journal of Cardiology
Volume110
Issue number9
DOIs
Publication statusPublished - Oct 1 2012

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint Dive into the research topics of 'Incidence, predictors, and implications of access site complications with transfemoral transcatheter aortic valve implantation'. Together they form a unique fingerprint.

  • Cite this

    Van Mieghem, N. M., Tchetche, D., Chieffo, A., Dumonteil, N., Messika-Zeitoun, D., Van Der Boon, R. M. A., Vahdat, O., Buchanan, G. L., Marcheix, B., Himbert, D., Serruys, P. W., Fajadet, J., Colombo, A., Carrié, D., Vahanian, A., & De Jaegere, P. P. T. (2012). Incidence, predictors, and implications of access site complications with transfemoral transcatheter aortic valve implantation. The American Journal of Cardiology, 110(9), 1361-1367. https://doi.org/10.1016/j.amjcard.2012.06.042