Intra- and inter-reader reproducibility of blood flow measurements on the ascending aorta and pulmonary artery using cardiac magnetic resonance

Giovanni Di Leo, Ida Daniela D’Angelo, Marco Alì, Paola Maria Cannaò, Giovanni Mauri, Francesco Secchi, Francesco Sardanelli

Research output: Contribution to journalArticle

Abstract

The aim of our study was to estimate the intra- and inter-reader reproducibility of blood flow measurements in the ascending aorta and main pulmonary artery using cardiac magnetic resonance (CMR) and a semi-automated segmentation method. The ethics committee approved this retrospective study. A total of 50 consecutive patients (35 males and 15 females; mean age±standard deviation 27±13 years) affected by congenital heart disease were reviewed. They underwent CMR for flow analysis of the ascending aorta and main pulmonary artery (1.5 T, through-plane phase-contrast sequences). Two independent readers (R1, trained radiology resident; R2, lower-trained technician student) obtained segmented images twice (>10-day interval), using a semi-automated method of segmentation. Peak velocity, forward and backward flows were obtained. Bland–Altman analysis was used and reproducibility was reported as complement to 100% of the ratio between the coefficient of repeatability and the mean. R1 intra-reader reproducibility for the aorta was 99% (peak velocity), 95% (forward flow) and 49% (backward flow); for the pulmonary artery, 99%, 91% and 90%, respectively. R2 intra-reader reproducibility was 92%, 91% and 38%; 98%, 86% and 87%, respectively. Inter-reader reproducibility for the aorta was 91%, 85% and 20%; for the pulmonary artery 96%, 75%, and 82%, respectively. Our results showed a good to excellent reproducibility of blood flow measurements of CMR together with a semiautomated method of segmentation, for all variables except the backward flow of the ascending aorta, with a limited impact of operator’s training.

Original languageEnglish
Pages (from-to)1-7
Number of pages7
JournalRadiologia Medica
DOIs
Publication statusAccepted/In press - Nov 22 2016

Fingerprint

Pulmonary Artery
Aorta
Magnetic Resonance Spectroscopy
Ethics Committees
Radiology
Heart Diseases
Retrospective Studies
Students

Keywords

  • Blood flow
  • Cardiac magnetic resonance
  • Heart valve
  • Phase-contrast sequences
  • Reproducibility

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Intra- and inter-reader reproducibility of blood flow measurements on the ascending aorta and pulmonary artery using cardiac magnetic resonance. / Di Leo, Giovanni; D’Angelo, Ida Daniela; Alì, Marco; Cannaò, Paola Maria; Mauri, Giovanni; Secchi, Francesco; Sardanelli, Francesco.

In: Radiologia Medica, 22.11.2016, p. 1-7.

Research output: Contribution to journalArticle

@article{287049c8a3704edc833a2b09951ac48e,
title = "Intra- and inter-reader reproducibility of blood flow measurements on the ascending aorta and pulmonary artery using cardiac magnetic resonance",
abstract = "The aim of our study was to estimate the intra- and inter-reader reproducibility of blood flow measurements in the ascending aorta and main pulmonary artery using cardiac magnetic resonance (CMR) and a semi-automated segmentation method. The ethics committee approved this retrospective study. A total of 50 consecutive patients (35 males and 15 females; mean age±standard deviation 27±13 years) affected by congenital heart disease were reviewed. They underwent CMR for flow analysis of the ascending aorta and main pulmonary artery (1.5 T, through-plane phase-contrast sequences). Two independent readers (R1, trained radiology resident; R2, lower-trained technician student) obtained segmented images twice (>10-day interval), using a semi-automated method of segmentation. Peak velocity, forward and backward flows were obtained. Bland–Altman analysis was used and reproducibility was reported as complement to 100{\%} of the ratio between the coefficient of repeatability and the mean. R1 intra-reader reproducibility for the aorta was 99{\%} (peak velocity), 95{\%} (forward flow) and 49{\%} (backward flow); for the pulmonary artery, 99{\%}, 91{\%} and 90{\%}, respectively. R2 intra-reader reproducibility was 92{\%}, 91{\%} and 38{\%}; 98{\%}, 86{\%} and 87{\%}, respectively. Inter-reader reproducibility for the aorta was 91{\%}, 85{\%} and 20{\%}; for the pulmonary artery 96{\%}, 75{\%}, and 82{\%}, respectively. Our results showed a good to excellent reproducibility of blood flow measurements of CMR together with a semiautomated method of segmentation, for all variables except the backward flow of the ascending aorta, with a limited impact of operator’s training.",
keywords = "Blood flow, Cardiac magnetic resonance, Heart valve, Phase-contrast sequences, Reproducibility",
author = "{Di Leo}, Giovanni and D’Angelo, {Ida Daniela} and Marco Al{\`i} and Canna{\`o}, {Paola Maria} and Giovanni Mauri and Francesco Secchi and Francesco Sardanelli",
year = "2016",
month = "11",
day = "22",
doi = "10.1007/s11547-016-0706-6",
language = "English",
pages = "1--7",
journal = "Radiologia Medica",
issn = "0033-8362",
publisher = "Springer-Verlag Italia s.r.l.",

}

TY - JOUR

T1 - Intra- and inter-reader reproducibility of blood flow measurements on the ascending aorta and pulmonary artery using cardiac magnetic resonance

AU - Di Leo, Giovanni

AU - D’Angelo, Ida Daniela

AU - Alì, Marco

AU - Cannaò, Paola Maria

AU - Mauri, Giovanni

AU - Secchi, Francesco

AU - Sardanelli, Francesco

PY - 2016/11/22

Y1 - 2016/11/22

N2 - The aim of our study was to estimate the intra- and inter-reader reproducibility of blood flow measurements in the ascending aorta and main pulmonary artery using cardiac magnetic resonance (CMR) and a semi-automated segmentation method. The ethics committee approved this retrospective study. A total of 50 consecutive patients (35 males and 15 females; mean age±standard deviation 27±13 years) affected by congenital heart disease were reviewed. They underwent CMR for flow analysis of the ascending aorta and main pulmonary artery (1.5 T, through-plane phase-contrast sequences). Two independent readers (R1, trained radiology resident; R2, lower-trained technician student) obtained segmented images twice (>10-day interval), using a semi-automated method of segmentation. Peak velocity, forward and backward flows were obtained. Bland–Altman analysis was used and reproducibility was reported as complement to 100% of the ratio between the coefficient of repeatability and the mean. R1 intra-reader reproducibility for the aorta was 99% (peak velocity), 95% (forward flow) and 49% (backward flow); for the pulmonary artery, 99%, 91% and 90%, respectively. R2 intra-reader reproducibility was 92%, 91% and 38%; 98%, 86% and 87%, respectively. Inter-reader reproducibility for the aorta was 91%, 85% and 20%; for the pulmonary artery 96%, 75%, and 82%, respectively. Our results showed a good to excellent reproducibility of blood flow measurements of CMR together with a semiautomated method of segmentation, for all variables except the backward flow of the ascending aorta, with a limited impact of operator’s training.

AB - The aim of our study was to estimate the intra- and inter-reader reproducibility of blood flow measurements in the ascending aorta and main pulmonary artery using cardiac magnetic resonance (CMR) and a semi-automated segmentation method. The ethics committee approved this retrospective study. A total of 50 consecutive patients (35 males and 15 females; mean age±standard deviation 27±13 years) affected by congenital heart disease were reviewed. They underwent CMR for flow analysis of the ascending aorta and main pulmonary artery (1.5 T, through-plane phase-contrast sequences). Two independent readers (R1, trained radiology resident; R2, lower-trained technician student) obtained segmented images twice (>10-day interval), using a semi-automated method of segmentation. Peak velocity, forward and backward flows were obtained. Bland–Altman analysis was used and reproducibility was reported as complement to 100% of the ratio between the coefficient of repeatability and the mean. R1 intra-reader reproducibility for the aorta was 99% (peak velocity), 95% (forward flow) and 49% (backward flow); for the pulmonary artery, 99%, 91% and 90%, respectively. R2 intra-reader reproducibility was 92%, 91% and 38%; 98%, 86% and 87%, respectively. Inter-reader reproducibility for the aorta was 91%, 85% and 20%; for the pulmonary artery 96%, 75%, and 82%, respectively. Our results showed a good to excellent reproducibility of blood flow measurements of CMR together with a semiautomated method of segmentation, for all variables except the backward flow of the ascending aorta, with a limited impact of operator’s training.

KW - Blood flow

KW - Cardiac magnetic resonance

KW - Heart valve

KW - Phase-contrast sequences

KW - Reproducibility

UR - http://www.scopus.com/inward/record.url?scp=84996528888&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84996528888&partnerID=8YFLogxK

U2 - 10.1007/s11547-016-0706-6

DO - 10.1007/s11547-016-0706-6

M3 - Article

AN - SCOPUS:84996528888

SP - 1

EP - 7

JO - Radiologia Medica

JF - Radiologia Medica

SN - 0033-8362

ER -