Intraocular Pressure Measurement in Patients Wearing Filtering Facepiece Masks

Luciano Quaranta, Eleonora Micheletti, Ivano Riva, Robert N. Weinreb

Research output: Contribution to journalArticlepeer-review

Abstract

Ophthalmological examination requires a strict contact between caregivers and patients. In the COVID-19 era, this may be a risk factor for virus spread, and the use of facial masks for all in-office ophthalmological procedures has been recommended. In this case-series, we report about some errors in intraocular pressure measurement, that may occur during the slit-lamp examination of patients wearing filtering facepiece masks and N95 respirators. This is mainly due to the greater dimensions of these masks in comparison with the surgical standard ones, and to the presence of a preshaped rigid nose area that may press against the Goldmann tonometer. Special care should be taken when measuring intraocular pressure in these cases.
Original languageEnglish
Pages (from-to)999-1000
Number of pages2
JournalJ. Glaucoma
Volume29
Issue number10
DOIs
Publication statusPublished - Oct 1 2020

Fingerprint

Dive into the research topics of 'Intraocular Pressure Measurement in Patients Wearing Filtering Facepiece Masks'. Together they form a unique fingerprint.