TY - JOUR
T1 - Limitations of the computerized laryngeal analyzer (CLA) in the assessment of the swallowing mechanism
AU - Maestri, R.
AU - Pistarini, C.
AU - Bertotti, G.
AU - Ferri, A.
AU - Guarnaschelli, C.
AU - Pinna, G. D.
PY - 2000/4
Y1 - 2000/4
N2 - This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P
AB - This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P
KW - Computerized laryngeal analyzer (CLA)
KW - Repeatability
KW - Swallowing disorders
UR - http://www.scopus.com/inward/record.url?scp=0034173923&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0034173923&partnerID=8YFLogxK
U2 - 10.1016/S1350-4533(00)00029-1
DO - 10.1016/S1350-4533(00)00029-1
M3 - Article
C2 - 10964042
AN - SCOPUS:0034173923
VL - 22
SP - 223
EP - 227
JO - Medical Engineering and Physics
JF - Medical Engineering and Physics
SN - 1350-4533
IS - 3
ER -