Limitations of the computerized laryngeal analyzer (CLA) in the assessment of the swallowing mechanism

R. Maestri, C. Pistarini, G. Bertotti, A. Ferri, C. Guarnaschelli, G. D. Pinna

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P

Original languageEnglish
Pages (from-to)223-227
Number of pages5
JournalMedical Engineering and Physics
Volume22
Issue number3
DOIs
Publication statusPublished - Apr 2000

Fingerprint

Deglutition
Thyroid Cartilage
Transducers
Saliva
Healthy Volunteers
Piezoelectric transducers
Cartilage
Water
Liquids

Keywords

  • Computerized laryngeal analyzer (CLA)
  • Repeatability
  • Swallowing disorders

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine
  • Psychology(all)

Cite this

Limitations of the computerized laryngeal analyzer (CLA) in the assessment of the swallowing mechanism. / Maestri, R.; Pistarini, C.; Bertotti, G.; Ferri, A.; Guarnaschelli, C.; Pinna, G. D.

In: Medical Engineering and Physics, Vol. 22, No. 3, 04.2000, p. 223-227.

Research output: Contribution to journalArticle

@article{b4a7e8fc85d5447a99218ac75e41302c,
title = "Limitations of the computerized laryngeal analyzer (CLA) in the assessment of the swallowing mechanism",
abstract = "This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P",
keywords = "Computerized laryngeal analyzer (CLA), Repeatability, Swallowing disorders",
author = "R. Maestri and C. Pistarini and G. Bertotti and A. Ferri and C. Guarnaschelli and Pinna, {G. D.}",
year = "2000",
month = "4",
doi = "10.1016/S1350-4533(00)00029-1",
language = "English",
volume = "22",
pages = "223--227",
journal = "Medical Engineering and Physics",
issn = "1350-4533",
publisher = "Elsevier BV",
number = "3",

}

TY - JOUR

T1 - Limitations of the computerized laryngeal analyzer (CLA) in the assessment of the swallowing mechanism

AU - Maestri, R.

AU - Pistarini, C.

AU - Bertotti, G.

AU - Ferri, A.

AU - Guarnaschelli, C.

AU - Pinna, G. D.

PY - 2000/4

Y1 - 2000/4

N2 - This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P

AB - This study was devised to assess the repeatability of measurements obtained from the Computerized Laryngeal Analyzer (CLA), a commercial system which monitors global laryngeal activity through a piezoelectric transducer positioned at the level of the thyroid cartilage. Twenty-two healthy subjects (37±7 yr) were asked to repeat three consecutive times the deglutition of three types of bolus: dry (saliva only), liquid (10 ml water) and solid (1 cm3 biscuit). The whole recording session was repeated 4±3 days apart. Two parameters were considered: amplitude and duration of the deglutition. Both intra- and inter-sessions repeatability were assessed by the standard error of measurement. The amplitude measurements were often found to be clipped at 100 mV, making duration the only parameter suitable for analysis. The duration was homogenous for the three types of bolus and no trend effect was found either intra- or inter-sessions. Inter-sessions repeatability was significantly lower than the intra-session one (P

KW - Computerized laryngeal analyzer (CLA)

KW - Repeatability

KW - Swallowing disorders

UR - http://www.scopus.com/inward/record.url?scp=0034173923&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034173923&partnerID=8YFLogxK