Maximum entropy and 2DFT NMR images

B. C. De Simone, F. De Luca, B. Maraviglia

Research output: Contribution to journalArticle

Abstract

A comparison between contrast and signal-to-noise ratio (SNR), as obtained by both maximum entropy (ME) and 2DFT reconstruction, is reported for NMR images. Our results show that in the ME elaboration the contrast and SNR are generally increased with respect to the 2DFT elaboration. The magnitude of such an increase depends on the ME statistical parameter and also, quite strongly, on the values of contrast and SNR of the 2DFT image.

Original languageEnglish
Pages (from-to)332-339
Number of pages8
JournalMagnetic Resonance in Medicine
Volume8
Issue number3
Publication statusPublished - 1988