Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents

Davide Carnelli, Giancarlo Pennati, Tomaso Villa, Luigi Baglioni, Bernhard Reimers, Francesco Migliavacca

Research output: Contribution to journalArticle

18 Citations (Scopus)

Abstract

Mechanical properties of carotid stents are important for the selection of the implantable device. The efficacy of cell design is usually investigated on the basis of clinical follow-ups; however, for the specific carotid application the relative importance of the cell design on the stenting outcome is debated. Ideally, a carotid stent should have at the same time low bending stiffness (i.e., high flexibility) to correctly conform to the vessel curvatures in the stenotic region and high stiffness to sustain the vessel wall in the radial direction. In this study, experimental tests on six carotid open-cell design self-expandable stents were carried out to assess bending and radial stiffness. A four-point bending test was adopted to measure the bending stiffness. Radial stiffness was obtained by measuring the stent resistance to local compression in a V-shaped support. Geometrical features, such as the link positioning and their length, were examined as a possible determinant of the mechanical properties of the tested stents. The tested stents showed very different geometrical features and mechanical properties despite similar material and a common open-cell structure consisting in rings connected by roughly straight links. In particular, different strategies are evident regarding the link position. Three distinct behaviors were observed: stents with both low radial and bending stiffness, moderate bending and high radial stiffness, and high bending and moderate radial stiffness. In general, a large link length allows the stent to be more flexible under bending condition, whereas the mutual links positioning plays a key role in the radial stiffness of these devices. Furthermore, for a fixed strut layout, a correlation was found between the strut cross area, as well as metal-to-artery ratio and the resultant radial and bending stiffness.

Original languageEnglish
Pages (from-to)74-80
Number of pages7
JournalArtificial Organs
Volume35
Issue number1
DOIs
Publication statusPublished - Jan 2011

Fingerprint

Stents
Shape memory effect
Stiffness
Mechanical properties
Bending (forming)
Struts
Equipment and Supplies
Bending (deformation)
Bending tests
Arteries
Metals

Keywords

  • Bending stiffness
  • Carotid stents
  • Radial stiffness
  • Shape memory alloy

ASJC Scopus subject areas

  • Biomaterials
  • Biomedical Engineering
  • Bioengineering
  • Medicine (miscellaneous)

Cite this

Carnelli, D., Pennati, G., Villa, T., Baglioni, L., Reimers, B., & Migliavacca, F. (2011). Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents. Artificial Organs, 35(1), 74-80. https://doi.org/10.1111/j.1525-1594.2010.01018.x

Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents. / Carnelli, Davide; Pennati, Giancarlo; Villa, Tomaso; Baglioni, Luigi; Reimers, Bernhard; Migliavacca, Francesco.

In: Artificial Organs, Vol. 35, No. 1, 01.2011, p. 74-80.

Research output: Contribution to journalArticle

Carnelli, D, Pennati, G, Villa, T, Baglioni, L, Reimers, B & Migliavacca, F 2011, 'Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents', Artificial Organs, vol. 35, no. 1, pp. 74-80. https://doi.org/10.1111/j.1525-1594.2010.01018.x
Carnelli, Davide ; Pennati, Giancarlo ; Villa, Tomaso ; Baglioni, Luigi ; Reimers, Bernhard ; Migliavacca, Francesco. / Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents. In: Artificial Organs. 2011 ; Vol. 35, No. 1. pp. 74-80.
@article{501c5d11a9e5464f9427ac8eb5107a2d,
title = "Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents",
abstract = "Mechanical properties of carotid stents are important for the selection of the implantable device. The efficacy of cell design is usually investigated on the basis of clinical follow-ups; however, for the specific carotid application the relative importance of the cell design on the stenting outcome is debated. Ideally, a carotid stent should have at the same time low bending stiffness (i.e., high flexibility) to correctly conform to the vessel curvatures in the stenotic region and high stiffness to sustain the vessel wall in the radial direction. In this study, experimental tests on six carotid open-cell design self-expandable stents were carried out to assess bending and radial stiffness. A four-point bending test was adopted to measure the bending stiffness. Radial stiffness was obtained by measuring the stent resistance to local compression in a V-shaped support. Geometrical features, such as the link positioning and their length, were examined as a possible determinant of the mechanical properties of the tested stents. The tested stents showed very different geometrical features and mechanical properties despite similar material and a common open-cell structure consisting in rings connected by roughly straight links. In particular, different strategies are evident regarding the link position. Three distinct behaviors were observed: stents with both low radial and bending stiffness, moderate bending and high radial stiffness, and high bending and moderate radial stiffness. In general, a large link length allows the stent to be more flexible under bending condition, whereas the mutual links positioning plays a key role in the radial stiffness of these devices. Furthermore, for a fixed strut layout, a correlation was found between the strut cross area, as well as metal-to-artery ratio and the resultant radial and bending stiffness.",
keywords = "Bending stiffness, Carotid stents, Radial stiffness, Shape memory alloy",
author = "Davide Carnelli and Giancarlo Pennati and Tomaso Villa and Luigi Baglioni and Bernhard Reimers and Francesco Migliavacca",
year = "2011",
month = "1",
doi = "10.1111/j.1525-1594.2010.01018.x",
language = "English",
volume = "35",
pages = "74--80",
journal = "Artificial Organs",
issn = "0160-564X",
publisher = "Wiley-Blackwell",
number = "1",

}

TY - JOUR

T1 - Mechanical Properties of Open-Cell, Self-Expandable Shape Memory Alloy Carotid Stents

AU - Carnelli, Davide

AU - Pennati, Giancarlo

AU - Villa, Tomaso

AU - Baglioni, Luigi

AU - Reimers, Bernhard

AU - Migliavacca, Francesco

PY - 2011/1

Y1 - 2011/1

N2 - Mechanical properties of carotid stents are important for the selection of the implantable device. The efficacy of cell design is usually investigated on the basis of clinical follow-ups; however, for the specific carotid application the relative importance of the cell design on the stenting outcome is debated. Ideally, a carotid stent should have at the same time low bending stiffness (i.e., high flexibility) to correctly conform to the vessel curvatures in the stenotic region and high stiffness to sustain the vessel wall in the radial direction. In this study, experimental tests on six carotid open-cell design self-expandable stents were carried out to assess bending and radial stiffness. A four-point bending test was adopted to measure the bending stiffness. Radial stiffness was obtained by measuring the stent resistance to local compression in a V-shaped support. Geometrical features, such as the link positioning and their length, were examined as a possible determinant of the mechanical properties of the tested stents. The tested stents showed very different geometrical features and mechanical properties despite similar material and a common open-cell structure consisting in rings connected by roughly straight links. In particular, different strategies are evident regarding the link position. Three distinct behaviors were observed: stents with both low radial and bending stiffness, moderate bending and high radial stiffness, and high bending and moderate radial stiffness. In general, a large link length allows the stent to be more flexible under bending condition, whereas the mutual links positioning plays a key role in the radial stiffness of these devices. Furthermore, for a fixed strut layout, a correlation was found between the strut cross area, as well as metal-to-artery ratio and the resultant radial and bending stiffness.

AB - Mechanical properties of carotid stents are important for the selection of the implantable device. The efficacy of cell design is usually investigated on the basis of clinical follow-ups; however, for the specific carotid application the relative importance of the cell design on the stenting outcome is debated. Ideally, a carotid stent should have at the same time low bending stiffness (i.e., high flexibility) to correctly conform to the vessel curvatures in the stenotic region and high stiffness to sustain the vessel wall in the radial direction. In this study, experimental tests on six carotid open-cell design self-expandable stents were carried out to assess bending and radial stiffness. A four-point bending test was adopted to measure the bending stiffness. Radial stiffness was obtained by measuring the stent resistance to local compression in a V-shaped support. Geometrical features, such as the link positioning and their length, were examined as a possible determinant of the mechanical properties of the tested stents. The tested stents showed very different geometrical features and mechanical properties despite similar material and a common open-cell structure consisting in rings connected by roughly straight links. In particular, different strategies are evident regarding the link position. Three distinct behaviors were observed: stents with both low radial and bending stiffness, moderate bending and high radial stiffness, and high bending and moderate radial stiffness. In general, a large link length allows the stent to be more flexible under bending condition, whereas the mutual links positioning plays a key role in the radial stiffness of these devices. Furthermore, for a fixed strut layout, a correlation was found between the strut cross area, as well as metal-to-artery ratio and the resultant radial and bending stiffness.

KW - Bending stiffness

KW - Carotid stents

KW - Radial stiffness

KW - Shape memory alloy

UR - http://www.scopus.com/inward/record.url?scp=78651429006&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78651429006&partnerID=8YFLogxK

U2 - 10.1111/j.1525-1594.2010.01018.x

DO - 10.1111/j.1525-1594.2010.01018.x

M3 - Article

C2 - 20497165

AN - SCOPUS:78651429006

VL - 35

SP - 74

EP - 80

JO - Artificial Organs

JF - Artificial Organs

SN - 0160-564X

IS - 1

ER -