Medical technology assessment photic stimulation - Standardization of screening methods

D. G A Kasteleijn-Nolst Trenité, C. D. Binnie, G. F A Harding, A. Wilkins, T. Covanis, O. Eeg-Olofsson, L. Goosens, O. Henriksen, G. Krämer, F. Leyten, F. H. Lopes Da Silva, A. Martins Da Silva, R. Naquet, B. Pedersen, S. Ricci, G. Rubboli, H. Spekreijse, S. Waltz

Research output: Contribution to journalArticle

38 Citations (Scopus)

Abstract

Rationale - In many EEG laboratories in Europe, intermittent photic stimulation (IPS) is not performed routinely, and consequently, great variation exists in the type of photo stimulator used, the methodology employed, and the interpretation of the EEC curves, thus leading to different outcomes. Methodology -It was decided to hold a consensus meeting with experts in the field of photic stimulation from various European countries. This meeting was held at the Stichting Epilepsie Instellingen Nederland, Heemstede, the Netherlands. The consensus reached was presented and discussed at the 9th European Congress of Clinical Neurophysiology in Ljubjana in June 1998. Results - Patients should be positioned at a distance of 30 cm from the photic stimulator (nasion to lamp) with dim surrounding lights, just enough to see the patient. Flashes should be delivered in separate trains of 10 s for each frequency, with intervals of 7 s minimum. First stimulation occurs with eyes open followed after 5 s by eye closure, while starting at 1 Hz progressing to 20 Hz, unless generalised epileptiform discharges are evoked at a lower frequency. Then, frequencies should start at 60 Hz decreasing to 25 Hz. The following frequencies should be used: 1, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 60, 50, 40, 30 and 25 Hz. The total duration is a maximum of 6 min (patients without a reaction to IPS). In interpreting the evoked responses, a clear distinction should be made between epileptiform responses confined to the occipital area (OSW), starting occipitally and spreading to frontal regions (OGSW), or generalised from the start (GSW). Other responses include generalised spikes (OR). Conclusion - This standard is safe, relatively quick, simple and reliable. Comparison of data within patients and between patients of various laboratories will also be possible. This will improve the quality of the care of the individual patient and make collaborative research possible.

Original languageEnglish
Pages (from-to)318-324
Number of pages7
JournalNeurophysiologie Clinique / Clinical Neurophysiology
Volume29
Issue number4
Publication statusPublished - 1999

Fingerprint

Biomedical Technology Assessment
Photic Stimulation
Neurophysiology
Quality of Health Care
European Union
Netherlands
Electroencephalography
Light
Research

Keywords

  • Diagnostic
  • EEG
  • Epilepsy
  • Photic stimulation

ASJC Scopus subject areas

  • Clinical Neurology
  • Neuroscience(all)

Cite this

Kasteleijn-Nolst Trenité, D. G. A., Binnie, C. D., Harding, G. F. A., Wilkins, A., Covanis, T., Eeg-Olofsson, O., ... Waltz, S. (1999). Medical technology assessment photic stimulation - Standardization of screening methods. Neurophysiologie Clinique / Clinical Neurophysiology, 29(4), 318-324.

Medical technology assessment photic stimulation - Standardization of screening methods. / Kasteleijn-Nolst Trenité, D. G A; Binnie, C. D.; Harding, G. F A; Wilkins, A.; Covanis, T.; Eeg-Olofsson, O.; Goosens, L.; Henriksen, O.; Krämer, G.; Leyten, F.; Lopes Da Silva, F. H.; Martins Da Silva, A.; Naquet, R.; Pedersen, B.; Ricci, S.; Rubboli, G.; Spekreijse, H.; Waltz, S.

In: Neurophysiologie Clinique / Clinical Neurophysiology, Vol. 29, No. 4, 1999, p. 318-324.

Research output: Contribution to journalArticle

Kasteleijn-Nolst Trenité, DGA, Binnie, CD, Harding, GFA, Wilkins, A, Covanis, T, Eeg-Olofsson, O, Goosens, L, Henriksen, O, Krämer, G, Leyten, F, Lopes Da Silva, FH, Martins Da Silva, A, Naquet, R, Pedersen, B, Ricci, S, Rubboli, G, Spekreijse, H & Waltz, S 1999, 'Medical technology assessment photic stimulation - Standardization of screening methods', Neurophysiologie Clinique / Clinical Neurophysiology, vol. 29, no. 4, pp. 318-324.
Kasteleijn-Nolst Trenité DGA, Binnie CD, Harding GFA, Wilkins A, Covanis T, Eeg-Olofsson O et al. Medical technology assessment photic stimulation - Standardization of screening methods. Neurophysiologie Clinique / Clinical Neurophysiology. 1999;29(4):318-324.
Kasteleijn-Nolst Trenité, D. G A ; Binnie, C. D. ; Harding, G. F A ; Wilkins, A. ; Covanis, T. ; Eeg-Olofsson, O. ; Goosens, L. ; Henriksen, O. ; Krämer, G. ; Leyten, F. ; Lopes Da Silva, F. H. ; Martins Da Silva, A. ; Naquet, R. ; Pedersen, B. ; Ricci, S. ; Rubboli, G. ; Spekreijse, H. ; Waltz, S. / Medical technology assessment photic stimulation - Standardization of screening methods. In: Neurophysiologie Clinique / Clinical Neurophysiology. 1999 ; Vol. 29, No. 4. pp. 318-324.
@article{952957f2a1ab4eb9ba22b0029b2dcd0e,
title = "Medical technology assessment photic stimulation - Standardization of screening methods",
abstract = "Rationale - In many EEG laboratories in Europe, intermittent photic stimulation (IPS) is not performed routinely, and consequently, great variation exists in the type of photo stimulator used, the methodology employed, and the interpretation of the EEC curves, thus leading to different outcomes. Methodology -It was decided to hold a consensus meeting with experts in the field of photic stimulation from various European countries. This meeting was held at the Stichting Epilepsie Instellingen Nederland, Heemstede, the Netherlands. The consensus reached was presented and discussed at the 9th European Congress of Clinical Neurophysiology in Ljubjana in June 1998. Results - Patients should be positioned at a distance of 30 cm from the photic stimulator (nasion to lamp) with dim surrounding lights, just enough to see the patient. Flashes should be delivered in separate trains of 10 s for each frequency, with intervals of 7 s minimum. First stimulation occurs with eyes open followed after 5 s by eye closure, while starting at 1 Hz progressing to 20 Hz, unless generalised epileptiform discharges are evoked at a lower frequency. Then, frequencies should start at 60 Hz decreasing to 25 Hz. The following frequencies should be used: 1, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 60, 50, 40, 30 and 25 Hz. The total duration is a maximum of 6 min (patients without a reaction to IPS). In interpreting the evoked responses, a clear distinction should be made between epileptiform responses confined to the occipital area (OSW), starting occipitally and spreading to frontal regions (OGSW), or generalised from the start (GSW). Other responses include generalised spikes (OR). Conclusion - This standard is safe, relatively quick, simple and reliable. Comparison of data within patients and between patients of various laboratories will also be possible. This will improve the quality of the care of the individual patient and make collaborative research possible.",
keywords = "Diagnostic, EEG, Epilepsy, Photic stimulation",
author = "{Kasteleijn-Nolst Trenit{\'e}}, {D. G A} and Binnie, {C. D.} and Harding, {G. F A} and A. Wilkins and T. Covanis and O. Eeg-Olofsson and L. Goosens and O. Henriksen and G. Kr{\"a}mer and F. Leyten and {Lopes Da Silva}, {F. H.} and {Martins Da Silva}, A. and R. Naquet and B. Pedersen and S. Ricci and G. Rubboli and H. Spekreijse and S. Waltz",
year = "1999",
language = "English",
volume = "29",
pages = "318--324",
journal = "Neurophysiologie Clinique",
issn = "0987-7053",
publisher = "Elsevier Masson",
number = "4",

}

TY - JOUR

T1 - Medical technology assessment photic stimulation - Standardization of screening methods

AU - Kasteleijn-Nolst Trenité, D. G A

AU - Binnie, C. D.

AU - Harding, G. F A

AU - Wilkins, A.

AU - Covanis, T.

AU - Eeg-Olofsson, O.

AU - Goosens, L.

AU - Henriksen, O.

AU - Krämer, G.

AU - Leyten, F.

AU - Lopes Da Silva, F. H.

AU - Martins Da Silva, A.

AU - Naquet, R.

AU - Pedersen, B.

AU - Ricci, S.

AU - Rubboli, G.

AU - Spekreijse, H.

AU - Waltz, S.

PY - 1999

Y1 - 1999

N2 - Rationale - In many EEG laboratories in Europe, intermittent photic stimulation (IPS) is not performed routinely, and consequently, great variation exists in the type of photo stimulator used, the methodology employed, and the interpretation of the EEC curves, thus leading to different outcomes. Methodology -It was decided to hold a consensus meeting with experts in the field of photic stimulation from various European countries. This meeting was held at the Stichting Epilepsie Instellingen Nederland, Heemstede, the Netherlands. The consensus reached was presented and discussed at the 9th European Congress of Clinical Neurophysiology in Ljubjana in June 1998. Results - Patients should be positioned at a distance of 30 cm from the photic stimulator (nasion to lamp) with dim surrounding lights, just enough to see the patient. Flashes should be delivered in separate trains of 10 s for each frequency, with intervals of 7 s minimum. First stimulation occurs with eyes open followed after 5 s by eye closure, while starting at 1 Hz progressing to 20 Hz, unless generalised epileptiform discharges are evoked at a lower frequency. Then, frequencies should start at 60 Hz decreasing to 25 Hz. The following frequencies should be used: 1, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 60, 50, 40, 30 and 25 Hz. The total duration is a maximum of 6 min (patients without a reaction to IPS). In interpreting the evoked responses, a clear distinction should be made between epileptiform responses confined to the occipital area (OSW), starting occipitally and spreading to frontal regions (OGSW), or generalised from the start (GSW). Other responses include generalised spikes (OR). Conclusion - This standard is safe, relatively quick, simple and reliable. Comparison of data within patients and between patients of various laboratories will also be possible. This will improve the quality of the care of the individual patient and make collaborative research possible.

AB - Rationale - In many EEG laboratories in Europe, intermittent photic stimulation (IPS) is not performed routinely, and consequently, great variation exists in the type of photo stimulator used, the methodology employed, and the interpretation of the EEC curves, thus leading to different outcomes. Methodology -It was decided to hold a consensus meeting with experts in the field of photic stimulation from various European countries. This meeting was held at the Stichting Epilepsie Instellingen Nederland, Heemstede, the Netherlands. The consensus reached was presented and discussed at the 9th European Congress of Clinical Neurophysiology in Ljubjana in June 1998. Results - Patients should be positioned at a distance of 30 cm from the photic stimulator (nasion to lamp) with dim surrounding lights, just enough to see the patient. Flashes should be delivered in separate trains of 10 s for each frequency, with intervals of 7 s minimum. First stimulation occurs with eyes open followed after 5 s by eye closure, while starting at 1 Hz progressing to 20 Hz, unless generalised epileptiform discharges are evoked at a lower frequency. Then, frequencies should start at 60 Hz decreasing to 25 Hz. The following frequencies should be used: 1, 2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 60, 50, 40, 30 and 25 Hz. The total duration is a maximum of 6 min (patients without a reaction to IPS). In interpreting the evoked responses, a clear distinction should be made between epileptiform responses confined to the occipital area (OSW), starting occipitally and spreading to frontal regions (OGSW), or generalised from the start (GSW). Other responses include generalised spikes (OR). Conclusion - This standard is safe, relatively quick, simple and reliable. Comparison of data within patients and between patients of various laboratories will also be possible. This will improve the quality of the care of the individual patient and make collaborative research possible.

KW - Diagnostic

KW - EEG

KW - Epilepsy

KW - Photic stimulation

UR - http://www.scopus.com/inward/record.url?scp=0033192136&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033192136&partnerID=8YFLogxK

M3 - Article

C2 - 10546250

AN - SCOPUS:0033192136

VL - 29

SP - 318

EP - 324

JO - Neurophysiologie Clinique

JF - Neurophysiologie Clinique

SN - 0987-7053

IS - 4

ER -