Metallic wear debris in dual modular hip arthroplasty.

M. Baleani, M. Viceconti, K. Walchholz, A. Toni

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Two cementless hip arthroplasty models with a modular neck made titanium alloy were submitted to cyclical loading in air, in physiological solution, and in a solution for accelerated corrosion tests in order to study the damage process of modular couplings. Thereafter, on one of the two models a series of tests were conducted in order to evaluate the quality of debris produced by the couplings. The parameters of testing adopted allow us to estimate the production of metallic debris observable in vivo with this type of prosthetic stem. The results show that the couplings of modular prostheses may be the source of metallic debris. The amount of debris produced depends on the geometry of the coupling and of the entire implant. Good planning and realization of the couplings, however, reduce the phenomenon that is negligible as compared to other potential sources of debris such as the surface of the stem or of the prosthetic head.

Original languageEnglish
Pages (from-to)231-238
Number of pages8
JournalLa Chirurgia degli organi di movimento
Volume82
Issue number3
Publication statusPublished - 1997

Fingerprint

Arthroplasty
Hip
Corrosion
Titanium
Prostheses and Implants
Neck
Air
Head

Cite this

Metallic wear debris in dual modular hip arthroplasty. / Baleani, M.; Viceconti, M.; Walchholz, K.; Toni, A.

In: La Chirurgia degli organi di movimento, Vol. 82, No. 3, 1997, p. 231-238.

Research output: Contribution to journalArticle

@article{21ee6c04b89f4723af291df7926e7208,
title = "Metallic wear debris in dual modular hip arthroplasty.",
abstract = "Two cementless hip arthroplasty models with a modular neck made titanium alloy were submitted to cyclical loading in air, in physiological solution, and in a solution for accelerated corrosion tests in order to study the damage process of modular couplings. Thereafter, on one of the two models a series of tests were conducted in order to evaluate the quality of debris produced by the couplings. The parameters of testing adopted allow us to estimate the production of metallic debris observable in vivo with this type of prosthetic stem. The results show that the couplings of modular prostheses may be the source of metallic debris. The amount of debris produced depends on the geometry of the coupling and of the entire implant. Good planning and realization of the couplings, however, reduce the phenomenon that is negligible as compared to other potential sources of debris such as the surface of the stem or of the prosthetic head.",
author = "M. Baleani and M. Viceconti and K. Walchholz and A. Toni",
year = "1997",
language = "English",
volume = "82",
pages = "231--238",
journal = "Chirurgia degli Organi di Movimento",
issn = "0009-4749",
publisher = "Nouva Casa Edtrice Licino Cappelli GEM",
number = "3",

}

TY - JOUR

T1 - Metallic wear debris in dual modular hip arthroplasty.

AU - Baleani, M.

AU - Viceconti, M.

AU - Walchholz, K.

AU - Toni, A.

PY - 1997

Y1 - 1997

N2 - Two cementless hip arthroplasty models with a modular neck made titanium alloy were submitted to cyclical loading in air, in physiological solution, and in a solution for accelerated corrosion tests in order to study the damage process of modular couplings. Thereafter, on one of the two models a series of tests were conducted in order to evaluate the quality of debris produced by the couplings. The parameters of testing adopted allow us to estimate the production of metallic debris observable in vivo with this type of prosthetic stem. The results show that the couplings of modular prostheses may be the source of metallic debris. The amount of debris produced depends on the geometry of the coupling and of the entire implant. Good planning and realization of the couplings, however, reduce the phenomenon that is negligible as compared to other potential sources of debris such as the surface of the stem or of the prosthetic head.

AB - Two cementless hip arthroplasty models with a modular neck made titanium alloy were submitted to cyclical loading in air, in physiological solution, and in a solution for accelerated corrosion tests in order to study the damage process of modular couplings. Thereafter, on one of the two models a series of tests were conducted in order to evaluate the quality of debris produced by the couplings. The parameters of testing adopted allow us to estimate the production of metallic debris observable in vivo with this type of prosthetic stem. The results show that the couplings of modular prostheses may be the source of metallic debris. The amount of debris produced depends on the geometry of the coupling and of the entire implant. Good planning and realization of the couplings, however, reduce the phenomenon that is negligible as compared to other potential sources of debris such as the surface of the stem or of the prosthetic head.

UR - http://www.scopus.com/inward/record.url?scp=0031173664&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031173664&partnerID=8YFLogxK

M3 - Article

VL - 82

SP - 231

EP - 238

JO - Chirurgia degli Organi di Movimento

JF - Chirurgia degli Organi di Movimento

SN - 0009-4749

IS - 3

ER -