Morphometric and microanalytical analyses of alumina wear particles in hip prostheses

A. Pizzoferrato, S. Stea, A. Sudanese, A. Toni, M. Nigrisoli, G. Gualtieri, S. Squarzoni

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

We analysed tissues harvested during 24 retrievals of hip joint prostheses, with one or both articular components made of alumina. We describe the morphology of wear particles, measure their size and analyse their chemical composition. We relate histopathological aspects to the parameters that characterize ceramic wear particles, and notice that tissue reaction relates to the physical aspect and amount of wear debris and does not necessarily depend on their chemical composition.

Original languageEnglish
Pages (from-to)583-587
Number of pages5
JournalBiomaterials
Volume14
Issue number8
DOIs
Publication statusPublished - 1993

Fingerprint

Hip prostheses
Hip Prosthesis
Aluminum Oxide
Alumina
Wear of materials
Joint Prosthesis
Hip Joint
Ceramics
Joint prostheses
Tissue
Joints
Chemical analysis
Debris

Keywords

  • alumina
  • Hip joint prosthesis
  • tissue reaction
  • wear

ASJC Scopus subject areas

  • Biotechnology
  • Bioengineering
  • Biomedical Engineering

Cite this

Pizzoferrato, A., Stea, S., Sudanese, A., Toni, A., Nigrisoli, M., Gualtieri, G., & Squarzoni, S. (1993). Morphometric and microanalytical analyses of alumina wear particles in hip prostheses. Biomaterials, 14(8), 583-587. https://doi.org/10.1016/0142-9612(93)90175-2

Morphometric and microanalytical analyses of alumina wear particles in hip prostheses. / Pizzoferrato, A.; Stea, S.; Sudanese, A.; Toni, A.; Nigrisoli, M.; Gualtieri, G.; Squarzoni, S.

In: Biomaterials, Vol. 14, No. 8, 1993, p. 583-587.

Research output: Contribution to journalArticle

Pizzoferrato, A, Stea, S, Sudanese, A, Toni, A, Nigrisoli, M, Gualtieri, G & Squarzoni, S 1993, 'Morphometric and microanalytical analyses of alumina wear particles in hip prostheses', Biomaterials, vol. 14, no. 8, pp. 583-587. https://doi.org/10.1016/0142-9612(93)90175-2
Pizzoferrato, A. ; Stea, S. ; Sudanese, A. ; Toni, A. ; Nigrisoli, M. ; Gualtieri, G. ; Squarzoni, S. / Morphometric and microanalytical analyses of alumina wear particles in hip prostheses. In: Biomaterials. 1993 ; Vol. 14, No. 8. pp. 583-587.
@article{3c12cb05b32f4830876790c61d3dd0fd,
title = "Morphometric and microanalytical analyses of alumina wear particles in hip prostheses",
abstract = "We analysed tissues harvested during 24 retrievals of hip joint prostheses, with one or both articular components made of alumina. We describe the morphology of wear particles, measure their size and analyse their chemical composition. We relate histopathological aspects to the parameters that characterize ceramic wear particles, and notice that tissue reaction relates to the physical aspect and amount of wear debris and does not necessarily depend on their chemical composition.",
keywords = "alumina, Hip joint prosthesis, tissue reaction, wear",
author = "A. Pizzoferrato and S. Stea and A. Sudanese and A. Toni and M. Nigrisoli and G. Gualtieri and S. Squarzoni",
year = "1993",
doi = "10.1016/0142-9612(93)90175-2",
language = "English",
volume = "14",
pages = "583--587",
journal = "Biomaterials",
issn = "0142-9612",
publisher = "Elsevier BV",
number = "8",

}

TY - JOUR

T1 - Morphometric and microanalytical analyses of alumina wear particles in hip prostheses

AU - Pizzoferrato, A.

AU - Stea, S.

AU - Sudanese, A.

AU - Toni, A.

AU - Nigrisoli, M.

AU - Gualtieri, G.

AU - Squarzoni, S.

PY - 1993

Y1 - 1993

N2 - We analysed tissues harvested during 24 retrievals of hip joint prostheses, with one or both articular components made of alumina. We describe the morphology of wear particles, measure their size and analyse their chemical composition. We relate histopathological aspects to the parameters that characterize ceramic wear particles, and notice that tissue reaction relates to the physical aspect and amount of wear debris and does not necessarily depend on their chemical composition.

AB - We analysed tissues harvested during 24 retrievals of hip joint prostheses, with one or both articular components made of alumina. We describe the morphology of wear particles, measure their size and analyse their chemical composition. We relate histopathological aspects to the parameters that characterize ceramic wear particles, and notice that tissue reaction relates to the physical aspect and amount of wear debris and does not necessarily depend on their chemical composition.

KW - alumina

KW - Hip joint prosthesis

KW - tissue reaction

KW - wear

UR - http://www.scopus.com/inward/record.url?scp=0027626613&partnerID=8YFLogxK