Neochordae implantation made easy with an adjustable device: Early report of acute and chronic animal experiments

Francesco Maisano, Paolo Denti, Hugo K. Vanermen, Joerg Seeburger, Volkmar Falk, Friedrich Wilhelm Mohr, Michael J. MacK, Ottavio R. Alfieri

Research output: Contribution to journalArticle

Abstract

Objective: Neochordae implantation is a well-established surgical solution for the treatment of mitral valve prolapse. The main limitation to wide usage of the technique has been the difficulty associated with accurate determination of neochordal length. We describe a system specifically designed to facilitate rapid, uncomplicated implantation and off-pump, beating heart adjustment of neochordae. Methods: Five swine underwent implantation of the adjustable neochordal system (V-Chordal; Valtech Cardio LTD, Israel) while on cardiopulmonary bypass after cutting native chordae to create a significant lesion. Neochordae length was adjusted with millimeter-level resolution, off-pump after discontinuation from bypass. Results: In all animals, the implant was successful. Under echocardiographic monitoring, flail lesions were corrected in all cases, using the anatomic landmarks or the degree of mitral regurgitation for real-time guidance. At postmortem gross examination, the implant and the neochordae were completely healed with evidence of tissue ingrowth. Conclusions: Preliminary animal experience suggests that the V-Chordal-adjustable neochordae system can be safely and effectively implanted, with accurate and precise adjustment of chordal length. The design of the device is suitable for a minimally invasive environment because of the long, flexible shafted design of the delivery system.

Original languageEnglish
Pages (from-to)287-290
Number of pages4
JournalInnovations: Technology and Techniques in Cardiothoracic and Vascular Surgery
Volume5
Issue number4
DOIs
Publication statusPublished - Jul 2010

Fingerprint

Anatomic Landmarks
Equipment Design
Equipment and Supplies
Mitral Valve Prolapse
Mitral Valve Insufficiency
Israel
Cardiopulmonary Bypass
Autopsy
Swine

Keywords

  • Dynamic devices
  • Mitral repair
  • Neochordae implant

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Surgery

Cite this

Neochordae implantation made easy with an adjustable device : Early report of acute and chronic animal experiments. / Maisano, Francesco; Denti, Paolo; Vanermen, Hugo K.; Seeburger, Joerg; Falk, Volkmar; Mohr, Friedrich Wilhelm; MacK, Michael J.; Alfieri, Ottavio R.

In: Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, Vol. 5, No. 4, 07.2010, p. 287-290.

Research output: Contribution to journalArticle

Maisano, Francesco ; Denti, Paolo ; Vanermen, Hugo K. ; Seeburger, Joerg ; Falk, Volkmar ; Mohr, Friedrich Wilhelm ; MacK, Michael J. ; Alfieri, Ottavio R. / Neochordae implantation made easy with an adjustable device : Early report of acute and chronic animal experiments. In: Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery. 2010 ; Vol. 5, No. 4. pp. 287-290.
@article{bad33954b313467687737ea88d0bb19a,
title = "Neochordae implantation made easy with an adjustable device: Early report of acute and chronic animal experiments",
abstract = "Objective: Neochordae implantation is a well-established surgical solution for the treatment of mitral valve prolapse. The main limitation to wide usage of the technique has been the difficulty associated with accurate determination of neochordal length. We describe a system specifically designed to facilitate rapid, uncomplicated implantation and off-pump, beating heart adjustment of neochordae. Methods: Five swine underwent implantation of the adjustable neochordal system (V-Chordal; Valtech Cardio LTD, Israel) while on cardiopulmonary bypass after cutting native chordae to create a significant lesion. Neochordae length was adjusted with millimeter-level resolution, off-pump after discontinuation from bypass. Results: In all animals, the implant was successful. Under echocardiographic monitoring, flail lesions were corrected in all cases, using the anatomic landmarks or the degree of mitral regurgitation for real-time guidance. At postmortem gross examination, the implant and the neochordae were completely healed with evidence of tissue ingrowth. Conclusions: Preliminary animal experience suggests that the V-Chordal-adjustable neochordae system can be safely and effectively implanted, with accurate and precise adjustment of chordal length. The design of the device is suitable for a minimally invasive environment because of the long, flexible shafted design of the delivery system.",
keywords = "Dynamic devices, Mitral repair, Neochordae implant",
author = "Francesco Maisano and Paolo Denti and Vanermen, {Hugo K.} and Joerg Seeburger and Volkmar Falk and Mohr, {Friedrich Wilhelm} and MacK, {Michael J.} and Alfieri, {Ottavio R.}",
year = "2010",
month = "7",
doi = "10.1097/IMI.0b013e3181ee865d",
language = "English",
volume = "5",
pages = "287--290",
journal = "Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery",
issn = "1556-9845",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - Neochordae implantation made easy with an adjustable device

T2 - Early report of acute and chronic animal experiments

AU - Maisano, Francesco

AU - Denti, Paolo

AU - Vanermen, Hugo K.

AU - Seeburger, Joerg

AU - Falk, Volkmar

AU - Mohr, Friedrich Wilhelm

AU - MacK, Michael J.

AU - Alfieri, Ottavio R.

PY - 2010/7

Y1 - 2010/7

N2 - Objective: Neochordae implantation is a well-established surgical solution for the treatment of mitral valve prolapse. The main limitation to wide usage of the technique has been the difficulty associated with accurate determination of neochordal length. We describe a system specifically designed to facilitate rapid, uncomplicated implantation and off-pump, beating heart adjustment of neochordae. Methods: Five swine underwent implantation of the adjustable neochordal system (V-Chordal; Valtech Cardio LTD, Israel) while on cardiopulmonary bypass after cutting native chordae to create a significant lesion. Neochordae length was adjusted with millimeter-level resolution, off-pump after discontinuation from bypass. Results: In all animals, the implant was successful. Under echocardiographic monitoring, flail lesions were corrected in all cases, using the anatomic landmarks or the degree of mitral regurgitation for real-time guidance. At postmortem gross examination, the implant and the neochordae were completely healed with evidence of tissue ingrowth. Conclusions: Preliminary animal experience suggests that the V-Chordal-adjustable neochordae system can be safely and effectively implanted, with accurate and precise adjustment of chordal length. The design of the device is suitable for a minimally invasive environment because of the long, flexible shafted design of the delivery system.

AB - Objective: Neochordae implantation is a well-established surgical solution for the treatment of mitral valve prolapse. The main limitation to wide usage of the technique has been the difficulty associated with accurate determination of neochordal length. We describe a system specifically designed to facilitate rapid, uncomplicated implantation and off-pump, beating heart adjustment of neochordae. Methods: Five swine underwent implantation of the adjustable neochordal system (V-Chordal; Valtech Cardio LTD, Israel) while on cardiopulmonary bypass after cutting native chordae to create a significant lesion. Neochordae length was adjusted with millimeter-level resolution, off-pump after discontinuation from bypass. Results: In all animals, the implant was successful. Under echocardiographic monitoring, flail lesions were corrected in all cases, using the anatomic landmarks or the degree of mitral regurgitation for real-time guidance. At postmortem gross examination, the implant and the neochordae were completely healed with evidence of tissue ingrowth. Conclusions: Preliminary animal experience suggests that the V-Chordal-adjustable neochordae system can be safely and effectively implanted, with accurate and precise adjustment of chordal length. The design of the device is suitable for a minimally invasive environment because of the long, flexible shafted design of the delivery system.

KW - Dynamic devices

KW - Mitral repair

KW - Neochordae implant

UR - http://www.scopus.com/inward/record.url?scp=77956244593&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77956244593&partnerID=8YFLogxK

U2 - 10.1097/IMI.0b013e3181ee865d

DO - 10.1097/IMI.0b013e3181ee865d

M3 - Article

C2 - 22437459

AN - SCOPUS:77956244593

VL - 5

SP - 287

EP - 290

JO - Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery

JF - Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery

SN - 1556-9845

IS - 4

ER -