TY - JOUR
T1 - New adaptive statistical iterative reconstruction ASiR-V
T2 - Assessment of noise performance in comparison to ASiR
AU - De Marco, Paolo
AU - Origgi, Daniela
PY - 2018/1/1
Y1 - 2018/1/1
N2 - Purpose: To assess the noise characteristics of the new adaptive statistical iterative reconstruction (ASiR-V) in comparison to ASiR. Methods: A water phantom was acquired with common clinical scanning parameters, at five different levels of CTDIvol. Images were reconstructed with different kernels (STD, SOFT, and BONE), different IR levels (40%, 60%, and 100%) and different slice thickness (ST) (0.625 and 2.5 mm), both for ASiR-V and ASiR. Noise properties were investigated and noise power spectrum (NPS) was evaluated. Results: ASiR-V significantly reduced noise relative to FBP: noise reduction was in the range 23%-60% for a 0.625 mm ST and 12%-64% for the 2.5 mm ST. Above 2 mGy, noise reduction for ASiR-V had no dependence on dose. Noise reduction for ASIR-V has dependence on ST, being greater for STD and SOFT kernels at 2.5 mm. For the STD kernel ASiR-V has greater noise reduction for both ST, if compared to ASiR. For the SOFT kernel, results varies according to dose and ST, while for BONE kernel ASIR-V shows less noise reduction. NPS for CT Revolution has dose dependent behavior at lower doses. NPS for ASIR-V and ASiR is similar, showing a shift toward lower frequencies as the IR level increases for STD and SOFT kernels. The NPS is different between ASiR-V and ASIR with BONE kernel. NPS for ASiR-V appears to be ST dependent, having a shift toward lower frequencies for 2.5 mm ST. Conclusions: ASiR-V showed greater noise reduction than ASiR for STD and SOFT kernels, while keeping the same NPS. For the BONE kernel, ASiR-V presents a completely different behavior, with less noise reduction and modified NPS. Noise properties of the ASiR-V are dependent on reconstruction slice thickness. The noise properties of ASiR-V suggest the need for further measurements and efforts to establish new CT protocols to optimize clinical imaging.
AB - Purpose: To assess the noise characteristics of the new adaptive statistical iterative reconstruction (ASiR-V) in comparison to ASiR. Methods: A water phantom was acquired with common clinical scanning parameters, at five different levels of CTDIvol. Images were reconstructed with different kernels (STD, SOFT, and BONE), different IR levels (40%, 60%, and 100%) and different slice thickness (ST) (0.625 and 2.5 mm), both for ASiR-V and ASiR. Noise properties were investigated and noise power spectrum (NPS) was evaluated. Results: ASiR-V significantly reduced noise relative to FBP: noise reduction was in the range 23%-60% for a 0.625 mm ST and 12%-64% for the 2.5 mm ST. Above 2 mGy, noise reduction for ASiR-V had no dependence on dose. Noise reduction for ASIR-V has dependence on ST, being greater for STD and SOFT kernels at 2.5 mm. For the STD kernel ASiR-V has greater noise reduction for both ST, if compared to ASiR. For the SOFT kernel, results varies according to dose and ST, while for BONE kernel ASIR-V shows less noise reduction. NPS for CT Revolution has dose dependent behavior at lower doses. NPS for ASIR-V and ASiR is similar, showing a shift toward lower frequencies as the IR level increases for STD and SOFT kernels. The NPS is different between ASiR-V and ASIR with BONE kernel. NPS for ASiR-V appears to be ST dependent, having a shift toward lower frequencies for 2.5 mm ST. Conclusions: ASiR-V showed greater noise reduction than ASiR for STD and SOFT kernels, while keeping the same NPS. For the BONE kernel, ASiR-V presents a completely different behavior, with less noise reduction and modified NPS. Noise properties of the ASiR-V are dependent on reconstruction slice thickness. The noise properties of ASiR-V suggest the need for further measurements and efforts to establish new CT protocols to optimize clinical imaging.
KW - Computed tomography
KW - Iterative reconstruction
KW - Noise power spectrum
UR - http://www.scopus.com/inward/record.url?scp=85040863668&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85040863668&partnerID=8YFLogxK
U2 - 10.1002/acm2.12253
DO - 10.1002/acm2.12253
M3 - Article
AN - SCOPUS:85040863668
JO - Journal of Applied Clinical Medical Physics
JF - Journal of Applied Clinical Medical Physics
SN - 1526-9914
ER -